Place and Temporal Cues in Cochlear Implant Pitch and Melody Perception

被引:18
|
作者
Swanson, Brett A. [1 ]
Marimuthu, Vijay [2 ]
Mannell, Robert H. [2 ]
机构
[1] Cochlear Ltd, Sydney, NSW, Australia
[2] Macquarie Univ, Fac Human Sci, Dept Linguist, Sydney, NSW, Australia
关键词
cochlear implant; pitch; melody; discrimination; sound coding; ELECTRICAL-STIMULATION; COMPLEX TONES; FREQUENCY-DISCRIMINATION; MUSICAL INTERVALS; PHASE SENSITIVITY; RELATIVE PITCH; PULSE TRAINS; RECOGNITION; STRATEGY; RANKING;
D O I
10.3389/fnins.2019.01266
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
The present study compared pitch and melody perception using cochlear place of excitation and temporal cues in six adult nucleus cochlear implant (CI) recipients. The stimuli were synthesized tones presented through a loudspeaker, and recipients used the Advanced Combinational Encoder (ACE) sound coding strategy on their own sound processors. Three types of tones were used, denoted H3, H4, and P5. H3 tones were harmonic tones with fundamental frequencies in the range C3-C4 (131-262 Hz), providing temporal pitch cues alone. H4 tones were harmonic tones with fundamental frequencies in the range C4-C5 (262-523 Hz), providing a mixture of temporal and place cues. P5 tones were pure tones with fundamental frequencies in the range C5-C6 (523-1046 Hz), providing place pitch cues alone. Four experimental procedures were used: pitch discrimination, pitch ranking, backward modified melodies, and warped modified melodies. In each trial of the modified melodies tests, subjects heard a familiar melody and a version with modified pitch (in randomized order), and had to select the unmodified melody. In all four procedures, many scores were much lower than would be expected for normal hearing listeners, implying that the strength of the perceived pitch was weak. Discrimination and ranking with H3 and P5 tones was poor for two-semitone intervals, but near perfect for intervals of five semitones and larger. H4 tones provided the lowest group mean scores in all four procedures, with some pitch reversals observed in pitch ranking. Group mean scores for P5 tones (place cues alone) were at least as high as those for H3 tones (temporal cues alone). The relatively good scores on the melody tasks with P5 tones were surprising, given the lack of temporal cues, raising the possibility of musical pitch using place cues alone. However, the alternative possibility that the CI recipients perceived the place cues as brightness, rather than musical pitch per se, cannot be excluded. These findings show that pitch perception models need to incorporate neural place representations alongside temporal cues if they are to predict pitch and melody perception in the absence of temporal cues.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] Cochlear Implant Rate Pitch and Melody Perception as a Function of Place and Number of Electrodes
    Marimuthu, Vijay
    Swanson, Brett A.
    Mannell, Robert
    [J]. TRENDS IN HEARING, 2016, 20
  • [2] Extending the Limits of Place and Temporal Pitch Perception in Cochlear Implant Users
    Olivier Macherey
    John M. Deeks
    Robert P. Carlyon
    [J]. Journal of the Association for Research in Otolaryngology, 2011, 12 : 233 - 251
  • [3] Extending the Limits of Place and Temporal Pitch Perception in Cochlear Implant Users
    Macherey, Olivier
    Deeks, John M.
    Carlyon, Robert P.
    [J]. JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2011, 12 (02): : 233 - 251
  • [4] Training of Cochlear Implant Users to Improve Pitch Perception in the Presence of Competing Place Cues
    Vandali, Andrew
    Sly, David
    Cowan, Robert
    van Hoesel, Richard
    [J]. EAR AND HEARING, 2015, 36 (02): : E1 - E13
  • [5] Cochlear Implant Users can Effectively Combine Place and Timing Cues for Pitch Perception
    Goldsworthy, Raymond L.
    Bissmeyer, Susan R. S.
    [J]. EAR AND HEARING, 2023, 44 (06): : 1410 - 1422
  • [6] Place-Pitch Interval Perception With a Cochlear Implant
    Stupak, Natalia
    Todd, Ann E.
    Landsberger, David M.
    [J]. EAR AND HEARING, 2021, 42 (02): : 301 - 312
  • [7] An integrated model of pitch perception incorporating place and temporal pitch codes with application to cochlear implant research
    Saeedi, Nafise Erfanian
    Blamey, Peter J.
    Burkitt, Anthony N.
    Grayden, David B.
    [J]. HEARING RESEARCH, 2017, 344 : 135 - 147
  • [8] Effect of stimulus level and place of stimulation on temporal pitch perception by cochlear implant users
    Carlyon, Robert P.
    Lynch, Catherine
    Deeks, John M.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2010, 127 (05): : 2997 - 3008
  • [9] Spectral and temporal cues in cochlear implant speech perception
    Nie, K
    Barco, A
    Zeng, FG
    [J]. EAR AND HEARING, 2006, 27 (02): : 208 - 217
  • [10] Contributions of temporal and place cues in pitch perception in absolute pitch possessors
    Fujisaki, W
    Kashino, M