Interfacing of a closed-cycle helium refrigerator to a vibrating sample magnetometer for low-temperature (12-300 K) magnetization studies

被引:2
|
作者
Shah, VR [1 ]
Markandeyulu, G [1 ]
Rao, KVSR [1 ]
机构
[1] Indian Inst Technol, Dept Phys, Magnetisme & Mat Magnet Lab, Madras 600036, Tamil Nadu, India
关键词
closed-cycle helium refrigerator; vibrating sample magnetometer; eddy currents;
D O I
10.1088/0957-0233/9/9/036
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A top-loading sample chamber tube, radiation shield and the vacuum shroud for interfacing a closed-cyde helium refrigerator for use with a vibrating sample magnetometer have been designed and fabricated. A lowest temperature of 12 K is achieved at the sample. The eddy current effects in the copper sample chamber tube are minimized by the use of an electrically insulating epoxy resin medium. The sample temperature is measured by a GaAlAs diode sensor attached on a sensor mount which is inserted into the sample zone and kept in helium atmosphere. The vibrations of the cryocooler's cold head are isolated from the vibrating sample magnetometer's head by incorporating a bellows. Using this facility the magnetic transitions of some rare earth elements have been measured and compared with the literature data.
引用
收藏
页码:1627 / 1631
页数:5
相关论文