AN AUTOMATIC SYSTEM FOR MEASURING THE FERROELECTRIC HYSTERESIS LOOPS USING THE MODIFIED SAWYER - TOWER CIRCUIT

被引:0
|
作者
Vakulenko, A. F. [1 ]
Vakhrushev, S. B. [2 ]
Filimonov, A. V. [1 ]
Zhang, N. [3 ]
机构
[1] Peter Great St Petersburg Polytech Univ, 29 Politech Skaya St, St Petersburg 195251, Russia
[2] Russian Acad Sci, Ioffe Inst, 26 Politech Skaya St, St Petersburg 194021, Russia
[3] Xi An Jiao Tong Univ, 28 Xianning W Rd,Jiaoda Commerce Block, Xian, Shaanxi, Peoples R China
关键词
polarization; ferroelectric; hysteresis; Sawyer - Tower circuit; MODEL; POLARIZATION;
D O I
10.18721/JPM.12409
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
In the article, an equipment and practical application of an automatic system (created by the authors) for measuring the ferroelectric hysteresis loops using the Atmega 328 microcontroller have been considered. The modern approaches to the classical Sawyer - Tower circuit's application was analyzed, and practical need for such development was proven. The schematic diagram and description of the main device's components were given. Test results on measuring the hysteresis loops in a barium titanate single crystal were presented, and they were compared with the data published earlier. Moreover, the results on measuring the ferroelectric hysteresis loops of an [110]-oriented 0.8Pb(Mb(1/3)Nb(2/3))O-3-0.2PbTiO(3) single crystal in a temperature range of 120 -300 K at frequencies from 2 to 50 Hz were presented.
引用
收藏
页码:89 / 96
页数:8
相关论文
共 50 条
  • [1] Polarization hysteresis loops of ferroelectric gate capacitors measured by Sawyer-Tower circuit
    Yoshimura, T. (tyoshi@ams.osakafu-u.ac.jp), 1600, Japan Society of Applied Physics (42):
  • [2] Polarization hysteresis loops of ferroelectric gate capacitors measured by Sawyer-tower circuit
    Yoshimura, T
    Fujimura, N
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2003, 42 (9B): : 6011 - 6014
  • [3] PC BASED FERROELECTRIC ANALYZER USING MODIFIED SAWYER AND TOWER CIRCUIT
    Singh, K.
    Limaye, S. S.
    Tiwari, R. U.
    Nath, S.
    Bhoga, S. S.
    FERROELECTRICS, 1996, 189 : 9 - 15
  • [4] MODIFIED SAWYER AND TOWER CIRCUIT FOR INVESTIGATION OF FERROELECTRIC SAMPLES
    SINHA, JK
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1965, 42 (09): : 696 - &
  • [5] AUTOMATIC SAMPLING CIRCUIT FOR FERROELECTRIC HYSTERESIS LOOPS
    GLAZER, AM
    GROVES, P
    SMITH, DT
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (02): : 95 - 97
  • [6] Numerical extraction of the true ferroelectric polarization due to switching domains from hysteresis loops measured using a Sawyer-Tower circuit
    Bouregba, R
    Poullain, G
    FERROELECTRICS, 2002, 274 : 165 - 181
  • [7] Direct hysteresis measurements on ferroelectret films by means of a modified Sawyer-Tower circuit
    Qiu, Xunlin
    Hollaender, Lars
    Wirges, Werner
    Gerhard, Reimund
    Basso, Heitor Cury
    JOURNAL OF APPLIED PHYSICS, 2013, 113 (22)
  • [8] A new modification to the Sawyer-Tower ferroelectric hysteresis loop tracer
    Ornelas-Arciniega, G
    Reyes-Gomez, J
    Castellanos-Guzman, AG
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 1998, 32 : S380 - S381
  • [9] New technique for measuring ferroelectric and antiferroelectric hysteresis loops
    Fukunaga, Mamoru
    Noda, Yukio
    JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 2008, 77 (06)
  • [10] Triangular current: Method for measuring hysteresis loops of ferroelectric capacitors
    Wang, Ding-Yeong
    Chang, Chun-Yen
    Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 2004, 43 (9 A): : 6225 - 6228