Multidimensional B-spline parameterization of the detection probability of the PET scanner Biograph 16 using GATE

被引:1
|
作者
Rehfeld, Niklas S. [1 ]
Vauclin, Sebastien [2 ]
Stute, Simon [1 ]
Buvat, Irene [1 ]
机构
[1] Paris 11 Univ, Paris 7 Univ, CNRS, IMNC,UMR 8165, 15 Rue Georges Clemenceau, F-91406 Orsay, France
[2] Univ Rouen, LITIS EA Lab, Rouen, France
关键词
RECONSTRUCTION;
D O I
10.1109/NSSMIC.2009.5402443
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
GATE (Geant4 Application for Emission Tomography) [I] is a widespread, well validated and very versatile application for Monte Carlo simulations in emission tomography. It allows very detailed simulations of positron emission tomography (PET) or single photon emission computed tomography (SPECT) scanners using the general purpose Monte Carlo code Geant4 [2], [3]. We present a method to accelerate GATE simulations when on the one hand the detection probability in the different crystals needs to be computed accurately taking into account the complex geometry of the detection system and on the other hand information on photon statistics is not required. The simulation is split into two parts. Inside the scanner (phantom or voxelized patient) the particles are tracked conventionally. When a particle leaves the phantom, the particle is transported to a virtual boundary that replaces the detection system in the simulation and the detection proability is calculated analytically by evaluating a multi-dimensional B-spline function taking into account the state of the photon on this virual border (position, incident angles, energy). The B-spline function itself is calculated by performing a conventional Monte Carlo simulation. This "pre"simulation needs to be performed only once for each scanner type. The presented method was tested by simulating the PET scanner Biograph HiREZ and using a realistic voxelized attenuation phantom. The results show good agreement between conventional GATE simulations and simulations using the B-spline. The simulations could be accerelated by a factor of around 23 yielding the same statistical accuracy.
引用
收藏
页码:4197 / +
页数:2
相关论文
共 50 条
  • [1] Multidimensional B-spline parameterization of the detection probability of PET systems to improve the efficiency of Monte Carlo simulations
    Rehfeld, Niklas S.
    Vauclin, Sebastien
    Stute, Simon
    Buvat, Irene
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : 3339 - 3361
  • [3] Learning mixtures of polynomials of multidimensional probability densities from data using B-spline interpolation
    Lopez-Cruz, Pedro L.
    Bielza, Concha
    Larranaga, Pedro
    INTERNATIONAL JOURNAL OF APPROXIMATE REASONING, 2014, 55 (04) : 989 - 1010
  • [4] Volumetric parameterization and trivariate B-spline fitting using harmonic functions
    Martin, T.
    Cohen, E.
    Kirby, R. M.
    COMPUTER AIDED GEOMETRIC DESIGN, 2009, 26 (06) : 648 - 664
  • [5] Surface reconstruction from stereo images using B-spline parameterization
    Yu, WR
    Xu, BG
    APPLICATIONS OF DIGITAL IMAGE PROCESSING XXVI, 2003, 5203 : 109 - 120
  • [6] Fast Multidimensional B-spline Interpolation Using Template Metaprogramming
    Huizinga, Wyke
    Klein, Stefan
    Poot, Dirk H. J.
    BIOMEDICAL IMAGE REGISTRATION (WBIR 2014), 2014, 8545 : 11 - 20
  • [7] Topology optimization of elastic contact problems using B-spline parameterization
    Li, Jiajia
    Zhang, Weihong
    Niu, Cao
    Gao, Tong
    STRUCTURAL AND MULTIDISCIPLINARY OPTIMIZATION, 2021, 63 (04) : 1669 - 1686
  • [8] Volumetric Parameterization and Trivariate B-spline Fitting using Harmonic Functions
    Martin, Tobias
    Cohen, Elaine
    Kirby, Robert M.
    SPM 2008: PROCEEDINGS OF THE ACM SOLID AND PHYSICAL MODELING SYMPOSIUM, 2008, : 269 - 280
  • [9] Topology optimization of elastic contact problems using B-spline parameterization
    Jiajia Li
    Weihong Zhang
    Cao Niu
    Tong Gao
    Structural and Multidisciplinary Optimization, 2021, 63 : 1669 - 1686
  • [10] Validation of a GATE simulation model for the siemens PET/CT biograph™ 6 scanner, using NEMA 2001 standards
    Gonias, P.