Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals

被引:27
|
作者
Wu, Hui [1 ]
Peng, Cheng [1 ,2 ]
Bai, Yulong [1 ]
Hu, Xin [1 ]
Wang, Lei [1 ]
Li, Changyi [1 ]
机构
[1] Tianjin Med Univ, Sch Stomatol, Tianjin 300070, Peoples R China
[2] Tianjin Med Univ, Dept Stomatol, Hosp 2, Tianjin 300070, Peoples R China
来源
BMC ORAL HEALTH | 2015年 / 15卷
关键词
Central axis transportation; Curvature straightening; ProTaper Universal; WaveOne; ProTaper Next; NICKEL-TITANIUM INSTRUMENTS; CLEANING EFFECTIVENESS; FILE-SYSTEMS; ROTARY; CURVATURES; RACE;
D O I
10.1186/s12903-015-0012-z
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: The purpose of this study was to compare the shaping ability of the ProTaper Universal (PTU; Dentsply Maillefer, Ballaigues, Switzerland), WaveOne (WO; Dentsply Maillefer) and ProTaper Next (PTN; Dentsply Maillefer) in simulated L-shaped and S-shaped root canals respectively. Methods: 30 simulated L-shaped and 30 simulated S-shaped root canals in resin blocks were employed and randomly divided into 3 groups (n = 10), respectively. The canals were prepared to a tip size 25 using PTU, WO or PTN: PTU F2 (taper 0.08 over the first 3 mm from apical tip), WO Primary (taper 0.08 over the first 3 mm from apical tip), and PTN X2 (taper 0.06 over the first 3 mm from apical tip). Photos of the simulated root canals were taken pre-and postinstrumentation. The 2 layers were superimposed after a series of image processing and 10 points were selected from apical constriction with 1 mm interval. And then the central axis transportation and straightened curvature were measured with software of image analysis. Results: In simulated L-shaped root canals, PTU and PTN caused less transportation than WO at curved section (P < 0.05), and PTN caused the least transportation at apical constriction (P < 0.05). Moreover, PTN maintained the canal curvature best among the 3 groups (P < 0.05). But PTN produced more transportation at straight section compared with PTU and WO (P < 0.05). In simulated S-shaped root canals, PTN preserved the coronal curvature best (P < 0.05), but there was no significant difference in apical curvature since all the files straightened the curvature obviously. Conclusions: PTN showed a better shaping ability than PTU and WO at the curved section of root canals, and PTN maintained the best apical constriction. But all the files had a tendency to straighten the apical curvature in multi-curved canals.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals
    Hui Wu
    Cheng Peng
    Yulong Bai
    Xin Hu
    Lei Wang
    Changyi Li
    [J]. BMC Oral Health, 15
  • [2] Shaping ability of ProTaper NEXT, ProTaper Universal and iRace files in simulated S-shaped canals
    Hiran-us, Sirawut
    Pimkhaokham, Somsinee
    Sawasdichai, Jirapat
    Ebihara, Arata
    Suda, Hideaki
    [J]. AUSTRALIAN ENDODONTIC JOURNAL, 2016, 42 (01) : 32 - 36
  • [3] Shaping ability of ProTaper Gold and WaveOne Gold nickel-titanium rotary instruments in simulated S-shaped root canals
    Shi, Lu
    Zhou, Junling
    Wan, Jie
    Yang, Yunfei
    [J]. JOURNAL OF DENTAL SCIENCES, 2022, 17 (01) : 430 - 437
  • [4] Influence of Operator's Experience on the Shaping Ability of Protaper Universal and Waveone Systems: A Comparative Study on Simulated Root Canals
    Troiano, Giuseppe
    Dioguardi, Mario
    Cocco, Armando
    Giannatempo, Giovanni
    Laino, Luigi
    Ciavarella, Domenico
    Berutti, Elio
    Lo Muzio, Lorenzo
    [J]. OPEN DENTISTRY JOURNAL, 2016, 10 : 546 - 552
  • [5] Comparative Assessment of the Shaping Ability of Reciproc Blue, WaveOne Gold, and ProTaper Gold in Simulated Root Canals
    Orel, Laura
    Velea-Barta, Oana-Alexandra
    Sinescu, Cosmin
    Duma, Virgil-Florin
    Nica, Luminita-Maria
    Horhat, Razvan Mihai
    Chirila, Raul Dorin
    Tudor, Anca
    Vulcanescu, Dan-Dumitru
    Negrutiu, Meda Lavinia
    [J]. MATERIALS, 2022, 15 (09)
  • [6] Comparative analysis of simulated root canals shaping, using ProTaper Universal, Next and Gold
    Neto, Filipa
    Ginjeira, Antonio
    [J]. REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2016, 57 (02): : 82 - 86
  • [7] Shaping ability of nickel-titanium rotary instruments in simulated S-shaped root canals
    Madureira, Rui Goncalves
    Navarro, Leopoldo Forner
    Llena, Mari Carmen
    Costa, Marcia
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2010, 109 (02): : E136 - E144
  • [8] Shaping Ability of Different Rotary and Reciprocating File Systems in Simulated S-Shaped Root Canals
    Mahmoud, Okba
    Alhimairi, Sara
    Sultan, Diana
    Ali, Ayah
    [J]. EUROPEAN ENDODONTIC JOURNAL, 2024, 9 (02): : 114 - 123
  • [9] Torque and Force Induced by ProTaper Universal and ProTaper Next during Shaping of Large and Small Root Canals in Extracted Teeth
    Arias, Ana
    Singh, Rupinderpal
    Peters, Ove A.
    [J]. JOURNAL OF ENDODONTICS, 2014, 40 (07) : 973 - 976
  • [10] Shaping ability of WaveOne Gold reciprocating file with and without glidepath in artificial S-shaped canals
    Karam, Mariana
    Zogheib, Carla
    Khalil, Issam
    [J]. GIORNALE ITALIANO DI ENDODONZIA, 2021, 35 (01): : 84 - 93