The case of the missing glutamine

被引:28
|
作者
Hancu, Ileana [1 ]
Port, John [2 ]
机构
[1] GE Global Res Ctr, Niskayuna, NY 12309 USA
[2] Mayo Clin, Dept Radiol, Rochester, MN USA
基金
美国国家卫生研究院;
关键词
glutamine; spectroscopy; MRS; simulations; 3; T; MAGNETIC-RESONANCE-SPECTROSCOPY; H-1 MR SPECTROSCOPY; HUMAN BRAIN; PROTON SPECTROSCOPY; METABOLIC COMPARTMENTATION; MAMMALIAN BRAIN; SIMULATIONS; SPECTRA; NMR; QUANTIFICATION;
D O I
10.1002/nbm.1620
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A theoretical study was performed to determine the accuracy and repeatability of multiple one-dimensional pulse sequences in the quantification of glutamine concentration at 3 T. Variable repeatability (12% to > 50%) and significant absolute error (-50% to +70%) were noted for the eight pulse sequences considered. Data acquired in vivo using three of the pulse sequences used for simulation matched the predicted repeatability well; among the pulse sequences considered, point-resolved spectroscopy (TE = 80 ms) offered minimal error and acceptable repeatability (12%) for brain glutamine measurements. Following correction for the expected bias of each pulse sequence, consistent glutamine measurements, in the 1-mM range, were reported with the three sequences. An explanation for the mismatch between in vivo (1)H MRS and in vitro (13)C/(1)H MRS at high field was attempted. Copyright (C) 2011 John Wiley & Sons, Ltd.