Evaluation of different stenting techniques in coronary bifurcations

被引:0
|
作者
Garcia, J. [1 ]
Doce, Y. [2 ]
Manuel, F. [1 ]
Castro, F. [3 ]
Goicolea, J. [4 ]
Fernandez, A. [4 ]
机构
[1] Univ Politecn Madrid, Dept Ingn Energet, C Jose Gutierrez Abascal 2, E-28006 Madrid, Spain
[2] Univ Politecn Madrid, Dept Ingn Mecan & Construcc, Rda Valencia 3, Madrid 28005, Spain
[3] Univ Valladolid, Dept Ingn Energet & Fluidomeccn, Pl Santa Cruz 8, Valladolid 47002, Spain
[4] Hosp Univ Puerta Hierro, C San Martin Porres 4, Madrid 28035, Spain
关键词
Stent; Coronary bifurcation; Computational fluid dynamics; Wall shear stress; WALL SHEAR-STRESS; CAROTID BIFURCATION; IN-VITRO; LESIONS; FLOW; ATHEROSCLEROSIS; IMPLANTATION; STRATEGY;
D O I
10.1016/j.rimni.2015.04.003
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The objective of this work is to analyze the local hemodynamic changes caused in a coronary bifurcation by three different stenting techniques: simple stenting of the main vessel, simple stenting of the main vessel with kissing balloon in the side branch and culotte. To carry out this study an idealized geometry of a coronary bifurcation is used, and two bifurcation angles, 45 degrees and 90 degrees, are chosen as representative of the wide variety of real configurations. Both numerical simulations and experimental measurements are performed. First, steady simulations are carried out with the commercial code ANSYS Fluent, then, experimental measurements with PIV (Particle Image Velocimetry), obtained in the laboratory, are used to validate the numerical simulations. The steady computational simulations show a good overall agreement with the experimental data. Second, pulsatile flow is considered to take into account the transient effects. The time averaged wall shear stress, oscillatory shear index ( OSI) and pressure drop obtained numerically are used to compare the behavior of the stenting techniques. As conclusions obtained, simple stenting technique shows the worst results in terms of pressure drop, and kissing the best ones. This criterion could be applied for DES (Drug Eluting Stents) where the restenosis is not as critical as the reestablishment of the blood flow. But considering the regions of low wall shear stress (<0.4 Pa) and elevated OSI (> 0.1), the culotte procedure shows the worst performance and the simple stent, the best. This could be applied to BMS (Bare Metal Stents). (C) 2015 CIMNE (Universitat Politecnica de Catalunya). Published by Elsevier Espana, S.L.U.
引用
收藏
页码:181 / 187
页数:7
相关论文
共 50 条
  • [1] Drug delivery patterns for different stenting techniques in coronary bifurcations: a comparative computational study
    Cutri, Elena
    Zunino, Paolo
    Morlacchi, Stefano
    Chiastra, Claudio
    Migliavacca, Francesco
    [J]. BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2013, 12 (04) : 657 - 669
  • [2] Drug delivery patterns for different stenting techniques in coronary bifurcations: a comparative computational study
    Elena Cutrì
    Paolo Zunino
    Stefano Morlacchi
    Claudio Chiastra
    Francesco Migliavacca
    [J]. Biomechanics and Modeling in Mechanobiology, 2013, 12 : 657 - 669
  • [3] Atheroma and coronary bifurcations: before and after stenting
    Yazdani, Saami K.
    Nakano, Masataka
    Otsuka, Fumiyuki
    Kolodgie, Frank D.
    Virmani, Renu
    [J]. EUROINTERVENTION, 2010, 6 : J24 - J30
  • [4] Kissing, Snugging, or "Potting"? The Evolution of Stenting Techniques in Bifurcations
    Di Mario, Carlo
    Foin, Nicolas
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2016, 9 (13) : 1407 - 1409
  • [5] Stenting bifurcations: The last frontier for fantasy in coronary interventions
    Colombo, A
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 67 (03) : 410 - 411
  • [6] The crush and culotte: Two different stent techniques but same results in coronary bifurcations
    Balta, Sevket
    Demirkol, Sait
    Cakar, Mustafa
    Kurt, Omer
    Sarlak, Hakan
    Celik, Turgay
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (03) : 2894 - 2895
  • [7] Coronary stenting with different indications
    Kobayashi, Y
    Ferraro, M
    Reimers, B
    Akiyama, T
    de Gregorio, J
    Di Mario, C
    Martini, G
    Di Francesco, L
    Finci, L
    Colombo, A
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 99A - 99A
  • [8] Tools & Techniques: PCI in coronary bifurcations lesions
    Louvard, Yves
    Lefevre, Thierry
    [J]. EUROINTERVENTION, 2011, 7 (01) : 160 - 163
  • [9] CORONARY STENTING - RESULTS WITH DIFFERENT INDICATIONS
    COLOMBO, A
    ALMAGOR, Y
    MAIELLO, L
    KHLAT, B
    GIANROSSI, R
    FINCI, L
    [J]. CIRCULATION, 1993, 88 (04) : 122 - 122
  • [10] Hemodynamic performance of different stent strategies for coronary bifurcations. Evaluation with a mathematical model
    Fernandez Diaz, J. A.
    Goicolea Ruigomez, J.
    Garcia Garcia, J.
    Manuel Martin, F.
    Doce Carrasco, Y.
    Garcia Touchard, A.
    Oteo Dominguez, J. F.
    Castro Ruiz, F.
    Crespo Martinez, A.
    Pulpon Rivera, L. A.
    [J]. EUROPEAN HEART JOURNAL, 2011, 32 : 1047 - 1047