Material Analysis of Spontaneously Subluxated Iris-Fixated Phakic Intraocular Lenses

被引:10
|
作者
Tandogan, Tamer [1 ,2 ]
Holzer, Mike P. [1 ,2 ]
Choi, Chul Young [1 ,2 ]
Auffarth, Gerd U. [1 ,2 ]
Gerten, Georg [3 ]
Khoramnia, Ramin [1 ,2 ]
机构
[1] Heidelberg Univ, David J Apple Int Lab Ophthalm Pathol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
[2] Heidelberg Univ, Int Vis Correct Res Ctr, Dept Ophthalmol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
[3] Augenklin Neumarkt, Cologne, Germany
关键词
FOLLOW-UP; IMPLANTATION; OUTCOMES; MYOPIA;
D O I
10.3928/1081597X-20160601-01
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To conduct a material analysis of spontaneously subluxated iris-fixated phakic intraocular lenses (IOLs) using light and scanning electron microscopy. METHODS: Six explanted Artisan/Verisyse and Artiflex/Veriflex IOLs (Ophtec, Groningen, Netherlands/AMO, Santa Ana, CA) were analyzed in a laboratory using light and scanning electron microscopy. Four of the IOLs had been explanted after spontaneous subluxation leading to a decrease in visual acuity, whereas the remaining two IOLs did not demonstrate signs of disenclavation but had been explanted in the course of planned cataract surgery and thus served as the control in this study. RESULTS: Light microscopy enabled the detection of clear deformations of the fixation arms on one or both haptics of the lens that had subluxated, and scanning electron microscopy revealed micro-cracks in the material. The deformations, which appeared to have been caused by some form of manipulation of the fixation arms, meant that adequate closure of the fixation arms was no longer possible and a reliable fixation in the iris stroma was unattainable. At least three of the lenses had been implanted with the use of an IOL haptic expander for enclavation, which may have contributed to the malformation of the haptics. The two control lenses exhibited no signs of material deformation or subsequent disenclavation. CONCLUSIONS: Deformations of the haptics of iris-fixated phakic IOLs can cause irreversible damage leading to disenclavation and inadequate re-enclavation. In such cases, explantation appears to be the only therapeutic option. It is important to adhere to an appropriate implantation technique to reduce the risk of spontaneous subluxation.
引用
收藏
页码:618 / +
页数:8
相关论文
共 50 条
  • [2] Iris-Fixated Phakic Intraocular Lenses: New Results
    Kohnen, Thomas
    [J]. OPHTHALMOLOGY, 2018, 125 (04) : 495 - 495
  • [3] Outcomes of reenclavation of subluxated iris-fixated phakic intraocular lenses: Comparison with primary surgery outcomes
    Titiyal, Jeewan S.
    Sharma, Namrata
    Mannan, Rashim
    Pruthi, Archna
    Vajpayee, Rasik B.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (04): : 577 - 581
  • [4] Risk factors for explantation of iris-fixated phakic intraocular lenses
    Jonker, Soraya M. R.
    Van Averbeke, Annelies A. C.
    Berendschot, Tos T. J. M.
    Saelens, Isabelle E. Y.
    Nuijts, Rudy M. M. A.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (08): : 1092 - 1098
  • [5] The Effect of Iris-Fixated Foldable Phakic Intraocular Lenses on Retinal Straylight
    Paarlberg, Jos C.
    Doors, Muriel
    Webers, Carroll A. B.
    Berendschot, Tos T. J. M.
    van den Berg, Thomas J. T. P.
    Nuijts, Rudy M. M. A.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2011, 152 (06) : 969 - 975
  • [6] Iris-fixated phakic intraocular lens explantation
    Duignan, Emma S.
    Quigley, Clare
    Treacy, Maxwell P.
    Eldouri, Azher
    O'Keefe, Michael
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (03) : 988 - 993
  • [7] Laser in situ Keratomileusis following the implantation of iris-fixated phakic Intraocular lenses
    Meltendorf, Christian
    Cichocki, Magdalena
    Kohnen, Thomas
    [J]. OPHTHALMOLOGICA, 2008, 222 (02) : 69 - 73
  • [8] Traumatic dislocation of an iris-fixated phakic intraocular lens
    A C Day
    Y Ramkissoon
    S George
    M C Corbett
    [J]. Eye, 2007, 21 : 121 - 122
  • [9] Late-Onset Decentration of Iris-Fixated Phakic Intraocular Lenses: A Case Series
    Doors, Muriel
    Eggink, Fred A.
    Webers, Carroll A. B.
    Nuijts, Rudy M. M. A.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (06) : 997 - 1003
  • [10] Iris-fixated posterior chamber intraocular lenses
    Condon, Garry P.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (09): : 1409 - 1409