Isoelectric focusing in a poly(dimethylsiloxane) microfluidic chip

被引:126
|
作者
Cui, HC
Horiuchi, K
Dutta, P
Ivory, CF [1 ]
机构
[1] Washington State Univ, Dept Chem Engn, Pullman, WA 99164 USA
[2] Washington State Univ, Sch Mech & Mat Engn, Pullman, WA 99164 USA
关键词
D O I
10.1021/ac048915+
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper reports the application of ampholyte-based isoelectric focusing in poly(dimethylsiloxane) (PDMS) using methylcellulose (MC) to reduce electroosmosis and peak drift. Although the characteristics of PDMS make it possible to fabricate microfluidic chips using soft lithography, unstable electroosmotic flow (EOF) and cathodic drift are significant problems when this medium is used. Ibis paper demonstrates that EOF is greatly reduced in PDMS by applying a dynamic coat of MC to the channel walls and that higher concentrations of MC can be used to increase the viscosity of the electrode solutions in order to suppress pH gradient drift and reduce "compression" of the pH gradient. To illustrate the effect of MC on performance, several fluorescent proteins were focused in microchip channels 5 mum deep by 300 mum wide by 2 cm long in 3-10 min using broad-range ampholytes at electric field strengths ranging from 25 to 100 V/cm.
引用
收藏
页码:1303 / 1309
页数:7
相关论文
共 50 条
  • [1] Integration of dialysis membranes into a poly(dimethylsiloxane) microfluidic chip for isoelectric focusing of proteins using whole-channel imaging detection
    Ou, Junjie
    Glawdel, Tornasz
    Samy, Razim
    Wang, Shuwen
    Liu, Zhen
    Ren, Carolyn L.
    Pawliszyn, Janusz
    ANALYTICAL CHEMISTRY, 2008, 80 (19) : 7401 - 7407
  • [2] Multistage isoelectric focusing in a polymeric microfluidic chip
    Cui, HC
    Horiuchi, K
    Dutta, P
    Ivory, CF
    ANALYTICAL CHEMISTRY, 2005, 77 (24) : 7878 - 7886
  • [3] Coaxial flow focusing in poly(dimethylsiloxane) microfluidic devices
    Tran, Tuan M.
    Cater, Sean
    Abate, Adam R.
    BIOMICROFLUIDICS, 2014, 8 (01):
  • [4] PREPARATION OF POLY(DIMETHYLSILOXANE) CHIP-BASED CARTRIDGE FOR ISOELECTRIC FOCUSING AND WHOLE-CHANNEL IMAGING DETECTION
    Ou, Junjie
    Wang, Shuwen
    Ren, Carolyn L.
    Pawliszyn, Janusz
    MicroNano2008-2nd International Conference on Integration and Commercialization of Micro and Nanosystems, Proceedings, 2008, : 221 - 222
  • [5] Multi-stage isoelectric focusing in a polymeric microfluidic chip
    Cui, HC
    Ivory, CF
    Horiuchi, K
    Dutta, P
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2005, 229 : U245 - U245
  • [6] Surface modification of poly (dimethylsiloxane)-fabricated microfluidic chip by UV radiation
    Meng, Fei
    Chen, Heng-Wu
    Fang, Qun
    Zhu, Hai-Lin
    Fang, Zhao-Lun
    Kao Teng Hsueh Hsiao Hua Heush Hsueh Pao/ Chemical Journal of Chinese Universities, 2002, 23 (07):
  • [7] Surface modification of poly(dimethylsiloxane)-fabricated microfluidic chip by UV radiation
    Meng, F
    Chen, HW
    Fang, Q
    Zhu, HL
    Fang, ZL
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2002, 23 (07): : 1264 - 1268
  • [8] INTEGRATION OF A PDMS/SU-8/QUARTZ MICROFLUIDIC CHIP WITH A NOVEL MACROPOROUS POLY DIMETHYLSILOXANE (PDMS) MEMBRANE FOR ISOELECTRIC FOCUSING OF PROTEINS USING WHOLE-CHANNEL IMAGING DETECTION
    Shameli, Seyed Mostafa
    Elbuken, Caglar
    Ren, Carolyn L.
    Pawliszyn, Janusz
    PROCEEDINGS OF THE 8TH INTERNATIONAL CONFERENCE ON NANOCHANNELS, MICROCHANNELS AND MINICHANNELS, 2010, PTS A AND B, 2011, : 905 - 906
  • [9] Parallel isoelectric focusing chip
    Zilberstein, G
    Korol, L
    Bukshpan, S
    Baskin, E
    PROTEOMICS, 2004, 4 (09) : 2533 - 2540
  • [10] Fully integrated PDMS/SU-8/quartz microfluidic chip with a novel macroporous poly dimethylsiloxane (PDMS) membrane for isoelectric focusing of proteins using whole-channel imaging detection
    Shameli, Seyed Mostafa
    Elbuken, Caglar
    Ou, Junjie
    Ren, Carolyn L.
    Pawliszyn, Janusz
    ELECTROPHORESIS, 2011, 32 (3-4) : 333 - 339