Quantitative temperature distribution measurements by non-contact scanning thermal microscopy using Wollaston probes under ambient conditions

被引:13
|
作者
Zhang, Yun [1 ]
Zhu, Wenkai [1 ]
Han, Liang [1 ]
Borca-Tasciuc, Theodorian [1 ]
机构
[1] Rensselaer Polytech Inst, Mech Aerosp & Nucl Engn Dept, Troy, NY 12180 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2020年 / 91卷 / 01期
关键词
HEAT-TRANSFER; CONDUCTIVITY;
D O I
10.1063/1.5099981
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Temperature measurement using Scanning Thermal Microscopy (SThM) usually involves heat transfer across the mechanical contact and liquid meniscus between the thermometer probe and the sample. Variations in contact conditions due to capillary effects at sample-probe contact and wear and tear of the probe and sample interfere with the accurate determination of the sample surface temperature. This paper presents a method for quantitative temperature sensing using SThM in noncontact mode. In this technique, the thermal probe is scanned above the sample at a distance comparable with the mean free path of ambient gas molecules. A Three-Dimensional Finite Element Model (3DFEM) that includes the details of the heat transfer between the sample and the probe in the diffusive and transition heat conduction regimes was found to accurately simulate the temperature profiles measured using a Wollaston thermal probe setup. In order to simplify the data reduction for the local sample temperature, analytical models were developed for noncontact measurements using Wollaston probes. Two calibration strategies (active calibration and passive calibration) for the sample-probe thermal exchange parameters are presented. Both calibration methods use sample-probe thermal exchange resistance correlations developed using the 3DFEM to accurately capture effects due to sample-probe gap geometry and the thermal exchange radii in the diffusive and transition regimes. The analytical data reduction methods were validated by experiments and 3DFEM simulations using microscale heaters deposited on glass and on dielectric films on silicon substrates. Experimental and predicted temperature profiles were independent of the probe-sample clearance in the range of 100-200 nm, where the sample-probe thermal exchange resistance is practically constant. The difference between the SThM determined and actual average microheater temperature rise was between 0.1% and 0.5% when using active calibration on samples with known thermal properties and between similar to 1.6% and 3.5% when using passive calibration, which yields robust sample-probe thermal exchange parameters that can be used also on samples with unknown thermal properties. Published under license by AIP Publishing
引用
收藏
页数:12
相关论文
共 48 条
  • [1] Thermal conductivity measurements of thin films by non-contact scanning thermal microscopy under ambient conditions
    Zhang, Yun
    Zhu, Wenkai
    Borca-Tasciuc, Theodorian
    [J]. NANOSCALE ADVANCES, 2021, 3 (03): : 692 - 702
  • [2] Sensitivity and spatial resolution for thermal conductivity measurements using noncontact scanning thermal microscopy with thermoresistive probes under ambient conditions
    Zhang, Yun
    Zhu, Wenkai
    Borca-Tasciuc, Theodorian
    [J]. OXFORD OPEN MATERIALS SCIENCE, 2021, 1 (01):
  • [3] Tip-sample Heat Transfer in Non-contact mode Scanning Thermal Microscopy with Wollaston Microprobes
    Zhang, Yanliang
    Han, Liang
    Borca-Tasciuc, Theodorian
    [J]. PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2011, VOL 10, PTS A AND B, 2012, : 405 - 406
  • [4] The effect of the contact point asymmetry on the accuracy of thin films thermal conductivity measurement by scanning thermal microscopy using Wollaston probes
    Hapenciuc, C. L.
    Negut, I.
    Visan, A.
    Borca-Tasciuc, T.
    Mihailescu, I. N.
    [J]. JOURNAL OF APPLIED PHYSICS, 2022, 131 (09)
  • [5] Quantitative sub-surface and non-contact imaging using scanning microwave microscopy
    Gramse, Georg
    Brinciotti, Enrico
    Lucibello, Andrea
    Patil, Samadhan B.
    Kasper, Manuel
    Rankl, Christian
    Giridharagopal, Rajiv
    Hinterdorfer, Peter
    Marcelli, Romolo
    Kienberger, Ferry
    [J]. NANOTECHNOLOGY, 2015, 26 (13)
  • [6] Accurate Non-Contact Body Temperature Measurement with Thermal Camera under Varying Environment Conditions
    Song, Changhoon
    Lee, Sukhan
    [J]. PROCEEDINGS OF THE 2022 16TH INTERNATIONAL CONFERENCE ON UBIQUITOUS INFORMATION MANAGEMENT AND COMMUNICATION (IMCOM 2022), 2022,
  • [7] Thermal Modelling Comparing High Temperature Fixed Point Measurements by Contact and Non-Contact Thermometry
    Castro, P.
    Machin, G.
    Pearce, J. V.
    [J]. TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 : 358 - 362
  • [8] A Full-Field Non-Contact Thermal Modal Testing Technique Under Ambient Excitation
    Y. J. Hu
    Z. S. Huang
    W. D. Zhu
    H. L. Li
    [J]. Experimental Techniques, 2021, 45 : 257 - 271
  • [9] A Full-Field Non-Contact Thermal Modal Testing Technique Under Ambient Excitation
    Hu, Y. J.
    Huang, Z. S.
    Zhu, W. D.
    Li, H. L.
    [J]. EXPERIMENTAL TECHNIQUES, 2021, 45 (03) : 257 - 271
  • [10] Non-contact temperature measurements using an infrared camera in aerospace welding applications
    Henrikson, P
    Ericsson, M
    [J]. TRENDS IN WELDING RESEARCH, PROCEEDINGS, 2003, : 930 - 935