Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness

被引:82
|
作者
Doyle, A [1 ]
Lachkar, Y [1 ]
机构
[1] Hop St Joseph, Inst Glaucome Fdn, F-75674 Paris, France
关键词
central corneal thickness; dynamic contour tonometer; Goldmann applanation tonometer; structurally normal cornea;
D O I
10.1097/01.ijg.0000169393.40298.05
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare dynamic contour tonometry with Goldmann applanation tonometry in structurally normal corneas over a wide range of central corneal thickness (CCT). Patients and Methods: Twenty-five patients each with normal CCT (group A), thin corneas (group B), and thick corneas (group C) had IOP measured with the Goldmann (GAT) and dynamic contour tonometer (DCT). Results: In group A (mean CCT = 5 52 +/- 16 mu m) the mean GAT was 15.9 +/- 3.1 mm Hg and mean DCT was 16 +/- 3.3 mm Hg (P = 0.91). In group B (mean CCT = 491 +/- 19 mu m) the mean GAT was 13.2 +/- 3.5 mm Hg and the mean DCT was 15.9 +/- 3.5 mm Hg (P = 0.009). For group C (mean CCT = 615 +/- 22 mu m), the mean GAT was 17.4 +/- 3.8 mm Hg and the mean DCT was 17.4 +/- 3.5 mm Hg (P = 0.95). The 95% agreement limits for DCT were -3.1 mm Hg to 2.9 mm Hg. The mean GAT-DCT difference was -2.6 mm Hg in thin corneas and -0.06 mm Hg in thick corneas. Below 520 mu m reduction of 10 mu m in CCT appears to result in a significant underestimation of the GAT IOP by 0.7 mm Hg (P < .001) and above 580 mu m a nonsignificant overestimation of 0.2 mm Hg per 10 mu m increase in CCT (P = 0.27). Conclusion: Dynamic contour tonometer agrees well on average with GAT but the agreement limits are wide. In structurally normal thin corneas DCT may give a more accurate assessment of the true IOP but it does not appear to have any benefit over GAT in thick corneas.
引用
收藏
页码:288 / 292
页数:5
相关论文
共 50 条
  • [1] Comparison of Goldmann applanation tonometry and Pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature
    Andreanos, Konstantinos
    Koutsandrea, Chryssanthi
    Papaconstantinou, Dimitris
    Diagourtas, Andreas
    Kotoulas, Andreas
    Dimitrakas, Panagiotis
    Moschos, Marilita M.
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 2477 - 2484
  • [2] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in relation to central corneal thickness in primary congenital glaucoma
    Marcelo Lopes Silva Jordão
    Vital Paulino Costa
    Maria de Lourdes Veronese Rodrigues
    Jayter Silva Paula
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2013, 251 : 117 - 121
  • [3] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in relation to central corneal thickness in primary congenital glaucoma
    Silva Jordao, Marcelo Lopes
    Costa, Vital Paulino
    Veronese Rodrigues, Maria de Lourdes
    Paula, Jayter Silva
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 251 (01) : 117 - 121
  • [5] Corneal thickness and Intraocular pressure After DSAEK: comparison of Dynamic Contour tonometry with Goldmann Applanation tonometry
    Borges, Barbara
    Vicente, Andre
    Anjos, Rita
    Vieira, Luisa
    Santos, Arnaldo Dias
    Ferreira, Joana
    Maduro, Vitor
    Alves, Nuno
    Feijao, Joao
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [6] The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    Kotecha, A
    White, ET
    Shewry, JM
    Garway-Heath, DF
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (12) : 1572 - 1575
  • [7] The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    Kotecha, A
    White, ET
    Shewry, JM
    Garway-Heath, DF
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [8] Central corneal thickness and applanation tonometry
    Huang, Yuqiang
    Tham, Clement C. Y.
    Zhang, Mingzhi
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (03): : 347 - 347
  • [9] Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness
    Iliev, M. E.
    Goldblum, D.
    Katsoulis, K.
    Amstutz, C.
    Frueh, B.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (07) : 833 - 835
  • [10] Comparison of dynamic contour tonometry with Goldmann applanation tonometry
    Kaufmann, C
    Bachmann, LM
    Thiel, MA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (09) : 3118 - 3121