2-Methylisoborneol;
Geosmin;
Solid-phase microextraction;
Gas chromatography-mass spectrometry;
D O I:
10.3724/SP.J.1096.2011.01932
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A method for the determination of 2-methylisoborneol (MIB) and geosmin (GSM) in water was developed by solid phase microextraction (SPME) coupled with gas chromatography-mass spectrometry(GC-MS). The best extraction conditions were as follows: water bath heating (The level of water surface in bath was set at the same level of sample), extraction temperature of 65 degrees C and extraction time of 30 min. The method was highly sensitive, better precisive and accuracy based on GC-MS results. The correlative coefficients for MIB and GSM were both over 0. 999; detection limits and RSD were 0. 2 ng/L, 0. 3 ng/L and 5. 8%, 3. 2% respectively; recoveries of spiked respectively were 102. 0% and 101. 7%. The method can be used to detect MIB and GSM in water.