Determination of Bromate at Trace Level in Sudanese Bottled Drinking Water Using Ion Chromatography

被引:5
|
作者
Musa, Mei [1 ]
Ahmed, Ibrahim M.
Atakruni, Ismat [2 ]
机构
[1] Minist Sci & Technol, Cent Lab, Khartoum, Sudan
[2] Scope Sci Consultancy, Khartoum, Sudan
关键词
Bromate; Ion chromatography; Sudanese bottled water; Determination; SPECTROPHOTOMETRIC DETERMINATION; POSTCOLUMN REACTION; COLUMN;
D O I
10.1155/2010/895306
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Bromate ion was determined by ion chromatography with spectro-photometric detection after post column reaction with o-dianisidine reagent in acidic medium. IonPac AS19 anion-exchange column was used with 20 mM sodium hydroxide eluent. The eluent from the column was then allowed to react with o-dianisidine at 60 degrees C. The developed colour of the final product was measured spectrophotometrically at 450 nm. Linearity of the method was checked up to 200 mu g/L by using 200 mu L sample injection volume and the linear correlation coefficient was 0.9998. The method detection limit was 0.56 mu g/L and the precision was 6.2% at 100 mu g/L. The spiked recovery for bromate in water samples was 93%. Interferences from some anions eg. nitrate, chloride and nitrite were studied. The method was applied for the determination of bromate ion in ten sudanese bottled drinking water. Bromate was detected in 60% of the samples at concentrations ranged between 5-169 mu g/L.
引用
收藏
页码:S283 / S293
页数:11
相关论文
共 50 条
  • [1] Simultaneous determination of trace level bromate and chlorinated haloacetic acids in bottled drinking water by ion chromatography
    Liu, YJ
    Mou, SF
    [J]. MICROCHEMICAL JOURNAL, 2003, 75 (02) : 79 - 86
  • [2] Determination of Trace Bromate in Drinking Water with Ion Chromatography
    Chen Rui
    Fu Xue-qi
    Shi Ying-lin
    [J]. 2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,
  • [3] Determination of trace level bromate in drinking water by direct injection ion chromatography
    Jackson, LK
    Joyce, RJ
    Laikhtman, M
    Jackson, PE
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1998, 829 (1-2) : 187 - 192
  • [4] TRACE-LEVEL DETERMINATION OF BROMATE IN OZONATED DRINKING-WATER USING ION CHROMATOGRAPHY
    JOYCE, RJ
    DHILLON, HS
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) : 165 - 171
  • [5] Determination of trace level bromate and perchlorate in drinking water by ion chromatography with an evaporative preconcentration technique
    Liu, YJ
    Mou, S
    Heberling, S
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 956 (1-2) : 85 - 91
  • [6] Determination of trace concentrations of bromate in municipal and bottled drinking waters using a hydroxide-selective column with ion chromatography
    De Borba, BM
    Rohrer, JS
    Pohl, CA
    Saini, C
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1085 (01) : 23 - 32
  • [7] Determination of Trace Bromate in Drinking Water by Ion Chromatography with Suppressed Conductivity Detection
    Ying Bo
    Li Shumin
    Yue Yinling
    E Xueli
    [J]. CHINESE JOURNAL OF CHROMATOGRAPHY, 2006, 24 (03) : 302 - 304
  • [8] Determination of bromate in bottled water marketed in Iran by ion chromatography
    Djam, Sima
    Najafi, Mostafa
    Ahmadi, Seyyed Hamid
    Shoeibi, Shahram
    [J]. JOURNAL OF CHEMICAL METROLOGY, 2019, 13 (02): : 47 - 52
  • [9] Determination of Trace Bromate in Drinking Water with High Chloride Matrix by Cyclic Ion Chromatography
    Ye, Mingli
    Nesterenko, Pavel N.
    Lu, Ying
    Huang, Xueting
    Jin, Xiaoling
    Chen, Meilan
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2021, 59 (03) : 217 - 222
  • [10] Determination of trace level perchlorate in drinking water and ground water by ion chromatography
    Jackson, PE
    Laikhtman, M
    Rohrer, JS
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1999, 850 (1-2) : 131 - 135