Sources of error in intra-arterial pressure measurements across a stenosis

被引:11
|
作者
McWilliams, RG [1 ]
Robertson, I [1 ]
Smye, SW [1 ]
Wijesinghe, L [1 ]
Kessel, D [1 ]
机构
[1] St James Univ Hosp, Dept Vasc Radiol, Leeds LS9 7TF, W Yorkshire, England
关键词
angiography; haemodynamics;
D O I
10.1016/S1078-5884(98)80116-0
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: To investigate potential errors associated with different techniques of intra-arterial pressure measurement at angiography. Materials and methods: An experimental model of an arterial stenosis was developed. Experiments were performed to assess the relevance of catheter position, catheter direction and catheter type on the recorded intraluminal pressure. Transstenotic pressure gradients were recorded with and without angiographic catheters crossing the stenosis. Results: At physiological flow rates angiographic catheter type does not influence the recorded pressure. At high flow rates through tight stenoses there is a significant catheter-related difference in recorded pressure adjacent to a stenosis. Downstream pressures may be altered by up to 85 mmHg when standard angiographic catheters are placed across a stenosis. Conclusion: The different techniques employed to measure pressure differences across a model stenosis may introduce significant errors tip to 85 mmHg. Care must be taken when pressure measurements alone are used to interpret the clinical significance of a stenosis. In low flow conditions there may not be a detectable pressure gradient across a 95% stenosis.
引用
收藏
页码:535 / 540
页数:6
相关论文
共 50 条
  • [1] Intra-arterial blood pressure measurement: sources of error and solutions
    Devasahayam, Suresh
    Gangadharan, Naveen
    Surekha, C.
    Baskaran, Bowya
    Mukadam, Farhan Adam
    Subramani, Sathya
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2022, 60 (04) : 1123 - 1138
  • [2] Intra-arterial blood pressure measurement: sources of error and solutions
    Suresh Devasahayam
    Naveen Gangadharan
    C. Surekha
    Bowya Baskaran
    Farhan Adam Mukadam
    Sathya Subramani
    [J]. Medical & Biological Engineering & Computing, 2022, 60 : 1123 - 1138
  • [3] Intra-arterial pressure measurements
    Tomlinson, M
    Mitchell, F
    Belli, AM
    Buckenham, TM
    Loosemore, TM
    Dormandy, JA
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1999, 17 (05) : 458 - 459
  • [4] INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS
    ZSIGMOND, EK
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1981, 245 (07): : 703 - 703
  • [5] Intra-arterial pressure measurements - Authors' reply
    Robertson, I
    McWilliams, RG
    Kessel, D
    Smye, S
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1999, 17 (05) : 459 - 459
  • [6] ERRORS IN INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS
    ZINGG, W
    BAINES, WD
    MATHUR, AN
    [J]. CANADIAN JOURNAL OF SURGERY, 1968, 11 (01) : 93 - &
  • [7] INTRA-ARTERIAL PRESSURE MONITORING
    GILSTON, A
    [J]. ANAESTHESIA, 1976, 31 (04) : 566 - 567
  • [8] AUSCULTATORY AND INTRA-ARTERIAL PRESSURE
    LONDE, S
    MOSS, AJ
    ADAMS, FH
    [J]. JOURNAL OF PEDIATRICS, 1967, 70 (04): : 663 - +
  • [9] BLOOD PRESSURE MEASUREMENTS IN OBESE PERSONS - COMPARISON OF INTRA-ARTERIAL AND AUSCULTATORY MEASUREMENTS
    BERLINER, K
    YILDIZ, M
    FUJIY, H
    GARNIER, B
    LEEDH
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1961, 7 (01): : 10 - &
  • [10] Estimating respiratory parameters using intra-arterial partial pressure measurements
    Jeremic, Aleksandar
    Tan, Kenneth
    [J]. 2006 FORTIETH ASILOMAR CONFERENCE ON SIGNALS, SYSTEMS AND COMPUTERS, VOLS 1-5, 2006, : 851 - +