Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor

被引:75
|
作者
Jones, CR [1 ]
Taylor, K [1 ]
Poston, L [1 ]
Shennan, AH [1 ]
机构
[1] St Thomas Hosp, Guys Kings & St Thomas Sch Med, Fetal Hlth Res Grp, Dept Obstet & Gynaecol, London SE1 7EH, England
关键词
blood pressure; measurement; validation; oscillometric;
D O I
10.1038/sj.jhh.1001141
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study has determined the accuracy of the Welch Allyn 'Vital Signs' blood pressure (BP) monitoring device in an adult population, according to the criteria of the British Hypertension Society (British Hypertension Society, 1993). Validation was also classified by the American 'Association for the Advancement of Medical Instrumentation' (AAMI). The Welch Allyn 'Vital Signs' monitor achieved a BHS grade A for both systolic and diastolic BP. It also met the criteria for the AAMI protocol. The mean differences between trained observers using a mercury sphygmomanometer and the device were -1 +/- 5 mm Hg for systolic BP and -3 +/- 4 mm Hg for diastolic BP. The Welch Allyn 'Vital Signs' monitor can be recommended for clinical use in the adult population. This is one of the first oscillometric monitors designed for routine use in a hospital environment that has been found to provide an accurate assessment of BP by accepted validation procedures.
引用
收藏
页码:191 / 195
页数:5
相关论文
共 50 条
  • [1] Validation of the Welch Allyn ‘Vital Signs’ oscillometric blood pressure monitor
    CR Jones
    K Taylor
    L Poston
    AH Shennan
    [J]. Journal of Human Hypertension, 2001, 15 : 191 - 195
  • [2] Validation of the Welch Allyn 'vital sign' oscillometric blood pressure monitor
    Jones, C
    Taylor, K
    Poston, L
    Shennan, A
    [J]. JOURNAL OF HYPERTENSION, 2000, 18 : S121 - S121
  • [3] Validation of the Welch Allyn 'Vital Signs' blood pressure measurement device in pregnancy and pre-eclampsia
    Reinders, A
    Cuckson, AC
    Jones, CR
    Poet, R
    O'Sullivan, G
    Shennan, AH
    [J]. BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2003, 110 (02) : 134 - 138
  • [4] Validation of the Welch Allyn QuietTrak ambulatory blood pressure monitor in pregnancy
    Penny, JA
    Shennan, AH
    Rushbrook, J
    Halligan, AW
    Taylor, DJ
    deSwiet, M
    [J]. HYPERTENSION IN PREGNANCY, 1996, 15 (03) : 313 - 321
  • [5] Accuracy of the Welch Allyn Vital Signs Monitor 52000 automatic blood pressure measuring device according to a modified British Hypertension Society protocol
    Braam, RL
    de Maat, C
    Thien, T
    [J]. BLOOD PRESSURE MONITORING, 2002, 7 (03) : 185 - 189
  • [6] Comparison of the Omron HeartGuide to the Welch Allyn ProBP 3400 blood pressure monitor
    Harfmann, Brianna D.
    Neph, Sylvia E.
    Gardner, Mason M.
    Plouffe, Audrey A.
    Vranish, Jennifer R.
    Montoye, Alexander H. K.
    [J]. BLOOD PRESSURE MONITORING, 2024, 29 (01) : 45 - 54
  • [7] Validation of the Welch Allyn Connex® Spot BP monitor on the accuracy of blood pressure measurements in pregnant and preeclamptic women
    Geddes-Barton, D.
    Palaniappan, V
    [J]. BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2018, 125 : 5 - 5
  • [8] Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator
    P D Davis
    J L Dennis
    R Railton
    [J]. Journal of Human Hypertension, 2005, 19 : 197 - 203
  • [9] Evaluation of the A&D UA-767 and Welch Allyn Spot Vital Signs noninvasive blood pressure monitors using a blood pressure simulator
    Davis, PD
    Dennis, JL
    Railton, R
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2005, 19 (03) : 197 - 203
  • [10] Validation of the Welch Allyn Spot Vital Signs blood pressure device according to the ANSI/AAMI SP10:2002.: Accuracy and cost-efficiency successfully combined
    Alpert, Bruce S.
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (05) : 345 - 347