Anaesthetic breathing circuit obstruction due to blockage of tracheal tube connector by a foreign body - Two cases

被引:14
|
作者
Foreman, MJ [1 ]
Moyes, DG [1 ]
机构
[1] Queen Elizabeth Hosp, Dept Anaesthesia, Adelaide, SA, Australia
关键词
techniques; equipment : breathing circuit; tracheal tube; complications; obstruction; foreign body; tracheal tube connector; crisis management;
D O I
10.1177/0310057X9902700116
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Two cases are presented which illustrate the disastrous consequences possible when an anaesthetic breathing circuit is obstructed by a foreign body. Despite reports of previous similar cases, work practices and equipment manufacture or design continue to allow for such events to occur The importance of both pre-anaesthetic testing of the entire circuit including attachments such as the tracheal tube connector and filters, and the removal of these parts should obstruction occur; is emphasised. Use of "clear" transparent breathing circuit components and opaque or brightly coloured packaging and caps which could potentially cause obstruction should decrease the incidence and facilitate the diagnosis of this problem.
引用
收藏
页码:73 / 75
页数:3
相关论文
共 11 条
  • [1] Breathing circuit obstruction by a foreign body
    Sethi, AK
    Mohta, M
    Sharma, P
    ANAESTHESIA AND INTENSIVE CARE, 2004, 32 (01) : 139 - 141
  • [2] TRACHEAL OBSTRUCTION DUE TO A FOREIGN BODY
    JACKSON, H
    BRITISH JOURNAL OF RADIOLOGY, 1950, 23 (273): : 574 - 575
  • [3] Hypoxemia due to the obstruction of a right tracheal bronchus by an endotracheal tube in a child with left bronchial foreign body aspiration
    Fourier, C
    Cremer, R
    Leteurtre, S
    Martinot, A
    Boussard, L
    Marquette, CH
    Declerck, N
    Pruvot, FR
    Leclerc, F
    PEDIATRIC PULMONOLOGY, 1998, 25 (05) : 343 - 344
  • [4] FOREIGN-BODY IN AIR PASSAGES DUE TO RUPTURE OF TRACHEAL TUBE CUFF
    LANDRY, YC
    BEDARD, O
    CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1981, 28 (01) : 89 - 89
  • [5] TRACHEAL OBSTRUCTION DUE TO GLOSSAL ENTRAPMENT BY A STRING FOREIGN BODY IN A PEAHEN (PAVO CRISTATUS)
    Rettenmund, Christy L.
    Chen, Sue
    JOURNAL OF EXOTIC PET MEDICINE, 2013, 22 (02) : 200 - 205
  • [6] Near-Complete tracheal obstruction due to mucormycosis: A report of two cases
    Fallahi, Mohammad Javad
    Nikandish, Reza
    Ziaian, Bizhan
    Shahriarirad, Reza
    CLINICAL CASE REPORTS, 2022, 10 (08):
  • [7] Two Cases of Severe Complications Due to an Esophageal Fish Bone Foreign Body
    Han, Ji-Hee
    Cha, Ra-Ri
    Kwak, Ji-Yoon
    Jeon, Hankyu
    Lee, Sang-Soo
    Jung, Jae Jun
    Cho, Jin Kyu
    Kim, Hyun Jin
    MEDICINA-LITHUANIA, 2023, 59 (09):
  • [8] Gluteal Abscess Secondary to Rectal Perforation Due to Ingested Foreign Body: A Report of Two Cases
    Amirhom, Emad
    Braganza, Joshua S.
    Gowda, Vinayak
    Shah, Ajay
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (10)
  • [9] Retained nasal tube bridle system insertion stylet presenting as nasal foreign body: A report of two cases
    Jackson, Ryan S.
    Sharma, Saurabh
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2015, 36 (02) : 296 - 298
  • [10] Foreign body sensation in the throat due to displacement of the superior cornu of the thyroid cartilage: two cases and a literature review
    Nadig, S. K.
    Uppal, S.
    Back, G. W.
    Coatesworth, A. P.
    Grace, A. R. H.
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2006, 120 (07): : 608 - 609