Validation of electronic visual acuity (EVA) measurement against standardised ETDRS charts in patients with visual field loss from inherited retinal degenerations

被引:11
|
作者
Jolly, Jasleen K. [1 ,2 ,3 ]
Junemann, Kristin [3 ]
Boagey, Heather [1 ]
Nadsady, Marie [2 ]
Bridge, Holly [3 ]
Maclaren, Robert E. [1 ,2 ]
机构
[1] Univ Oxford, Nuffield Dept Clin Neurosci, Nuffield Lab Ophthalmol, Oxford, England
[2] Oxford Univ Hosp NHS Fdn Trust, Oxford Eye Hosp, Oxford, England
[3] Univ Oxford, Oxford Ctr Funct MRI Brain FMRIB, Wellcome Ctr Integrat Neuroimaging, Oxford, England
关键词
field of vision; diagnostic tests; investigation; degeneration; psychophysics; VISual perception; RELIABILITY; EYE;
D O I
10.1136/bjophthalmol-2019-315124
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background With the increase in clinical trials testing therapy for retinal disease, there is a need to ensure that outcome measures are both accurate and standardised. The US Food and Drug Administration favours the use of visual acuity measured using ETDRS logMAR charts. The loss of visual field can interfere with visual tracking across the charts, leading to increased variability of measurements. Electronic visual acuity (EVA) presents the optotype on the centre of a screen, thereby removing the tracking element of the task, and may provide a more precise measurement. Methods Visual acuity was measured twice using ETDRS charts, EVA automated single letter (E-ETDRS) and EVA single line (EVA-SL) presentation (EMMES). Patients underwent microperimetry (MAIA; Centervue) to determine visual field. We tested 65 patients with rod-cone dystrophies and 41 healthy volunteers. Results Both participant groups read 2-3 letters more on average on the electronic charts compared with ETDRS. Limits of agreement using a modified Bland-Altman analysis account for replicates were wider in eyes with foveal defects (-9 to 18) compared with eyes without foveal defects (-11 to 15). Electronic charts in the presence of foveal defects reduced the range (-11 to 13). Conclusion EVA may provide more accurate measures of visual acuity than traditional ETDRS charts in patients when the visual field loss encroached on the central vision. Electronic presentation with a single line of letters was the favoured style reported by patients and should be considered in future interventional clinical trials.
引用
收藏
页码:924 / 931
页数:8
相关论文
共 5 条
  • [1] Validation of Electronic Visual Acuity Measurement Against Standardised ETDRS Charts In Retinal Disease
    Jolly, Jasleen Kaur
    Juenemann, Kristin
    Boagey, Heather
    Moschandreas, Joanna
    Nadsady, Marie
    Bridge, Holly
    MacLaren, Robert E.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [2] Effect of reduced visual acuity on microperimetry performance in patients with inherited retinal degenerations
    Davies, Alexandra
    Nanda, Anika
    MacLaren, Robert E.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [3] Validation of a computerised logMAR visual acuity measurement system (COMPlog): comparison with ETDRS and the electronic ETDRS testing algorithm in adults and amblyopic children
    Laidlaw, D. A. H.
    Tailor, V.
    Shah, N.
    Atamian, S.
    Harcourt, C.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2008, 92 (02) : 241 - 244
  • [4] Reliability of self-measurement of visual acuity in AMD patients with two electronic devices based on the ETDRS chart: A randomized study
    Queguiner, F.
    Bezirganyan, K.
    Courjaret, J. C.
    Curel, L.
    Penaranda, G.
    Bonomini, J.
    Chossegros, M. Righini
    [J]. JOURNAL FRANCAIS D OPHTALMOLOGIE, 2024, 47 (01):