Single event upsets in implantable cardioverter defibrillators

被引:60
|
作者
Bradley, PD [1 ]
Normand, E
机构
[1] Univ Wollongong, Dept Engn Phys, Wollongong, NSW 2522, Australia
[2] Boeing Def & Space Grp, Seattle, WA 98124 USA
关键词
D O I
10.1109/23.736549
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Single event upsets (SEU) have been observed in implantable cardiac defibrillators. The incidence of SEUs is well modeled by upset rate calculations attributable to the secondary cosmic ray neutron flux. The effect of recent interpretations of the shape of the heavy ion cross-section curve on neutron burst generation rate calculations is discussed. The model correlates well with clinical experience and is consistent with the expected geographical variation of the secondary cosmic ray neutron flux. The observed SER was 9.3 x 10(-12) upsets/bit-hr from 22 upsets collected over a total of 284672 device days. This is the first clinical data set obtained indicating the effects of cosmic radiation on implantable devices. Importantly, it may be used to predict the susceptibility of future implantable device designs to cosmic radiation. The significance of cosmic radiation effects relative to other radiation sources applicable to implantable devices is discussed.
引用
收藏
页码:2929 / 2940
页数:12
相关论文
共 50 条
  • [1] Implantable cardioverter defibrillators
    Shah, CP
    Thakur, RK
    Xie, BY
    Hoon, VK
    EMERGENCY MEDICINE CLINICS OF NORTH AMERICA, 1998, 16 (02) : 463 - +
  • [2] Implantable cardioverter defibrillators
    Cardiac Pacemakers, Inc, St. Paul, United States
    Proc IEEE, 3 (468-479):
  • [3] Implantable Cardioverter Defibrillators
    不详
    MEDICAL LETTER ON DRUGS AND THERAPEUTICS, 2009, 51 (1304): : 6 - 7
  • [4] Implantable cardioverter defibrillators
    Harris, Stuart
    Dhinoja, Mehul
    CLINICAL MEDICINE, 2007, 7 (04) : 397 - 400
  • [5] Implantable cardioverter defibrillators
    Warren, JA
    Dreher, RD
    Jaworski, RV
    Putzke, JJ
    Russie, RJ
    PROCEEDINGS OF THE IEEE, 1996, 84 (03) : 468 - 479
  • [6] Implantable cardioverter-defibrillators
    DiMarco, JP
    NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (19): : 1836 - 1847
  • [7] Deactivating implantable cardioverter defibrillators
    Lynn, J
    ANNALS OF INTERNAL MEDICINE, 2005, 143 (09) : 691 - 691
  • [8] Athletes with Implantable Cardioverter Defibrillators
    Ponamgi, Shiva P.
    DeSimone, Christopher V.
    Ackerman, Michael J.
    CLINICS IN SPORTS MEDICINE, 2015, 34 (03) : 473 - +
  • [9] IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    KUTALEK, SP
    DREIFUS, LS
    ADVANCES IN INTERNAL MEDICINE, VOLUME 38, 1993, 38 : 421 - 438
  • [10] Implantable cardioverter-defibrillators
    Pinski, SL
    Fahy, GJ
    AMERICAN JOURNAL OF MEDICINE, 1999, 106 (04): : 446 - 458