Dosimetric comparison of two treatment planning systems for spine SBRT

被引:3
|
作者
Lee, Young Kyung [1 ,2 ]
Munawar, Iram [3 ]
Mashouf, Shahram [4 ]
Sahgal, Arjun [2 ,5 ]
Ruschin, Mark [1 ,2 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Dept Med Phys, Odette Canc Ctr, T Wing,Rm TG 271, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M4N 3M5, Canada
[3] Credit Valley Hosp, Trillium Hlth Partners, Dept Med Phys, Mississauga, ON, Canada
[4] Marshfield Clin Hlth Syst, Dept Med Phys, Marshfield, WI 54449 USA
[5] Sunnybrook Hlth Sci Ctr, Dept Radiat Oncol, Odette Canc Ctr, Toronto, ON, Canada
关键词
Automatic segmentation; Adaptive planning; Head and neck; Cancer; Radiotherapy; STEREOTACTIC BODY RADIOTHERAPY; CARLO DOSE CALCULATION; ARC; OPTIMIZATION;
D O I
10.1016/j.meddos.2019.07.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Two commonly used treatment planning systems (TPS) are compared for the planning of spine stereotactic body radiotherapy (SBRT). The main purpose is to highlight relative advantages and disadvantages of each system and propose a methodologic approach for comparisons. Twenty clinical plans were inversely planned with step-and-shoot intensity-modulated radiotherapy (IMRT) each using 9 to 11 beams, referred to as IMRT_P. The prescription dose was 24 Gy in 2 fractions, and the plans were generated following our institutional protocol using the Pinnacle(3) v9.2. Each case was replanned using a 2-arc volumetric modulated arc therapy (VMAT) approach, referred as VMAT_P. CT and structure sets were DICOM exported to Monaco v5.10 and planned in 2 different ways: IMRT (IMRT_M) and VMAT (VMAT_M) using the same prescription dose. Dose volume histograms (DVH) and other dose statistics of planning target volumes (PTV) and organ-at-risk (OAR) were analyzed and compared between plans. The gradient index (GI= ratio of 50% isodose volume to prescribed isodose volume) was used to measure dose fall-off outside of the PTV. Another metric - Gradient Index Inner (GI(inner) =the rate (in Gy/mm) - at which the dose changes from the level of the spinal cord/thecal sac toward the prescription dose) was developed and compared. All plans were considered clinically acceptable by institutional guidelines and achieved all of the OAR dose constraints. VMAT_M and IMRT_M showed comparable dose statistics for the PTV when compared to VMAT_P and IMRT_P, respectively. For IMRT plans, the median GI(inner) was 1.88 Gy/mm vs 1.52 Gy/mm for IMRT_M and IMRT_P respectively (p< 0.001). All other IMRT metrics were statistically similar except for the PTV maximum dose (D-max), which was higher for IMRT_M than IMRT_P (median 30.7 Gy vs 29.0 Gy, p< 0.001). For VMAT plans, only PTV D-min showed a statistical different between VMAT_M and VMAT_P of median 12.7 Gy vs 9.7 Gy (p< 0.001). In terms of beam sequencing parameters, the number of monitor units was statistically higher for VMAT_P compared to VMAT_M (median = 6764 vs 5376) whereas the number of segments for IMRT_M was statistically greater than IMRT_P (median =155 vs 73). We were able to generate clinically acceptable plans for different types of spine SBRT using 2 different TPS. We used an evaluation strategy involving coverage, conformity, and dose gradient that can compared between TPS. (C) 2019 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:77 / 84
页数:8
相关论文
共 50 条
  • [1] A dosimetric analysis of a spine SBRT specific treatment planning system
    Saenz, Daniel L.
    Crownover, Richard
    Stathakis, Sotirios
    Papanikolaou, Niko
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01): : 154 - 159
  • [2] Evaluating Treatment Couch Dosimetric Effects in Spine SBRT Treatment Planning
    Winey, B.
    Wolfgang, J.
    Speier, C.
    Crawford, B.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [3] A dosimetric comparative analysis of Brainlab elements and Eclipse RapidArc for spine SBRT treatment planning
    Deshazer, Garron
    Narayanasamy, Ganesh
    Bimali, Milan
    Galhardo, Edvaldo
    Kalantari, Faraz
    Xia, Fen
    Penagaricano, Jose
    Morrill, Steven
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (02):
  • [4] DOSIMETRIC COMPARISON OF SPINE SBRT TECHNIQUES: A STUDY BASED ON RPC SPINE PHANTOM
    Shi, Wenyin
    Liu, Haisong
    Studenski, Matthew
    Fu, Lei
    Peng, Chang
    Gunn, Vickie
    Rudoler, Shari
    Farrell, Christopher
    Andrews, David
    Chu, James
    Turian, Julius
    [J]. NEURO-ONCOLOGY, 2012, 14 : 140 - 141
  • [5] Single Fraction Stereotactic Body Radiation Therapy (SBRT) for Spine Metastasis: A Dosimetric Comparison of Multiple Treatment Delivery Systems
    Kaufman, I.
    Pandya, S.
    Powell, R.
    Fredrick, J.
    Bossenberger, T.
    Ramirez, E.
    Yan, Y.
    Solberg, T.
    Burmeister, J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S736 - S736
  • [6] Evaluation of Intrafractional Motion and Dosimetric Impact in Spine SBRT Treatment
    Xu, Yiwen
    Hu, Yida David
    Bredfeldt, Jeremy
    Hacker, Fred
    [J]. MEDICAL PHYSICS, 2019, 46 (11) : 5380 - 5380
  • [7] Pre treatment quality assurance in SBRT: comparison of different dosimetric systems for VMAT
    Bruschi, A.
    Russo, S.
    Esposito, M.
    Pini, S.
    Ghirelli, A.
    Bastiani, P.
    Zatelli, G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S751 - S752
  • [8] Dosimetric comparison of two commercial IMRT planning and delivery systems
    Duan, J
    Fiveash, JB
    Shen, S
    Murshed, H
    Ove, P
    Liengswangwong, V
    [J]. RADIOLOGY, 2001, 221 : 602 - 602
  • [9] Dosimetric variability in long cancer IMRT/SBRT among institutions with identical treatment planning systems
    Das, I. J.
    Desrosiers, C. M.
    Srivastava, S. P.
    Chopra, K. L.
    Khadivi, K. O.
    Taylor, M. S.
    Zhao, Q.
    Johnstone, P. A. S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S604 - S605
  • [10] A Dosimetric Study Comparing Two Beam Arrangement Strategies in Fractionated Thoracic Spine Stereotactic Body Radiotherapy (SBRT) Planning
    Wang, C.
    Xie, Y.
    Hu, Z.
    Yin, F. F.
    Reitman, Z.
    Cui, Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E557 - E557