An Optimized Colorimetric Readout Method for Lateral Flow Immunoassays

被引:19
|
作者
Park, Jongwon [1 ]
机构
[1] Kyungil Univ, Dept Biomed Engn, 50 Gamasil Gil, Gyongsan 38428, Gyeongbuk Do, South Korea
基金
新加坡国家研究基金会;
关键词
colorimetry; photometry; assay reader; quantification;
D O I
10.3390/s18124084
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Despite its broad penetration of various markets, the quantitative lateral flow immunoassay (LFIA) suffers from sensitivity issues in some cases. To solve this problem, an optimized colorimetric readout method for LFIA quantification is proposed in this study. An assay reader device utilizing a color camera and an analysis method using a Bayer filtered image were developed. Spectrometric measurements of the assay test line were performed to determine the color channel that contains the test line information and effectively minimizes noise. The change in the intensity ratio with increasing concentration of the target substance in the sample was largest in the green channel. The linear range of the output curve ranged from 0 to 10 ng/mL, and the detection limit was 2 ng/mL. The suggested instrumentation and analysis methods are expected to effectively resolve the low-sensitivity problems of the former LFIA systems and to offer other prospective functionalities for LFIA quantification.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Direct comparison of colorimetric signal amplification techniques in lateral flow immunoassays
    Sathishkumar, N.
    Toley, Bhushan J.
    ANALYTICAL METHODS, 2024, 16 (42) : 7200 - 7209
  • [2] Magnetic Lateral Flow Immunoassays
    Moyano, Amanda
    Serrano-Pertierra, Esther
    Salvador, Maria
    Martinez-Garcia, Jose Carlos
    Rivas, Montserrat
    Blanco-Lopez, M. Carmen
    DIAGNOSTICS, 2020, 10 (05)
  • [3] Plasmonic-Enhanced Colorimetric Lateral Flow Immunoassays Using Bimetallic Silver-Coated Gold Nanostars
    Atta, Supriya
    Zhao, Yuanhao
    Sanchez, Sebastian
    Seedial, Deven
    Devadhasan, Jasmine Pramila
    Summers, Alexander Jarrett
    Gates-Hollingsworth, Marcellene A.
    Pflughoeft, Kathryn J.
    Gu, Jian
    Montgomery, Douglas C.
    AuCoin, David P.
    Zenhausern, Frederic
    Vo-Dinh, Tuan
    ACS APPLIED MATERIALS & INTERFACES, 2024, 16 (40) : 54907 - 54918
  • [4] Analytical sensitivity limits for lateral flow immunoassays
    Gordon, Julian
    Michel, Gerd
    CLINICAL CHEMISTRY, 2008, 54 (07) : 1250 - 1251
  • [5] Nanolabels for biosensors based on lateral flow immunoassays
    Saweres-Arguelles, C.
    Sanchez-Calvo, A.
    Serrano-Pertierra, E.
    Matos, M.
    Blanco-Lopez, M. C.
    ANALYTICA CHIMICA ACTA, 2025, 1340
  • [6] A SIMPLEX OPTIMIZED COLORIMETRIC METHOD FOR FORMALDEHYDE
    HOULE, MJ
    LONG, DE
    SMETTE
    ANALYTICAL LETTERS, 1970, 3 (08) : 401 - &
  • [7] Automated Readout of a SERS Lateral Flow Assay
    Baritaux, Jean-Charles
    Hoang, Antoine
    Morel, Nathalie
    CLINICAL AND PRECLINICAL OPTICAL DIAGNOSTICS II, 2019, 11073
  • [8] Silver and gold enhancement methods for lateral flow immunoassays
    Oliveira Rodriguez, Myriam
    Blanco Covian, Lucia
    Costa Garcia, Agustin
    Carmen Blanco-Lopez, Maria
    TALANTA, 2016, 148 : 272 - 278
  • [9] Lateral flow immunoassays for antigens, antibodies and haptens detection
    Li, Ge
    Li, Qingmei
    Wang, Xun
    Liu, Xiao
    Zhang, Yuhang
    Li, Rui
    Guo, Junqing
    Zhang, Gaiping
    INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2023, 242
  • [10] Design of SERS nanotags for multiplexed lateral flow immunoassays
    Sanchez-Purra, Maria
    Roig-Solvas, Biel
    Versiani, Alice
    Rodriguez-Quijada, Cristina
    de Puig, Helena
    Bosch, Irene
    Gehrke, Lee
    Hamad-Schifferli, Kimberly
    MOLECULAR SYSTEMS DESIGN & ENGINEERING, 2017, 2 (04): : 401 - 409