MAIA microperimeter for short-duration fixation stability measurements in central vision loss: Repeatability and comparison with the Nidek MP1

被引:3
|
作者
Pyatova, Yulia [1 ]
Markowitz, Samuel N. [1 ]
Devenyi, Robert G. [1 ,2 ]
Tarita-Nistor, Luminita [2 ]
机构
[1] Univ Toronto, Dept Ophthalmol & Vis Sci, Toronto, ON, Canada
[2] Toronto Western Hosp, Krembil Res Inst, Donald K Johnson Eye Inst, Toronto, ON, Canada
关键词
age-related macular degeneration; fixation stability; MAIA microperimeter; MP1; microperimeter; test-retest; PREFERRED RETINAL LOCUS; BIOFEEDBACK STIMULATION; SENSITIVITY; LOCATION;
D O I
10.1111/opo.12960
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose This study reports the repeatability of 20 s-duration fixation stability measurements recorded with the Macular Integrity Assessment (MAIA) microperimeter in patients with central vision loss, in contrast to the Nidek MP1 microperimeter. Methods Fixation stability was recorded in 39 eyes of 25 patients with macular disease using MAIA and the MP1 for 20 s intervals, twice for each eye, with each instrument. Twenty eyes were identified as the better eye (BE) and 19 eyes as the worse eye (WE). Fixation stability was quantified with the 95% bivariate contour ellipse area (BCEA), logarithmically transformed. Bland-Altman plots were used to determine the 95% limits of agreement. Results For MAIA, the 95% limits of agreement were +/- 0.84 log deg(2) for the BE and +/- 0.66 log deg(2) for the WE. Similarly, for the MP1 these limits were +/- 0.48 log deg(2) for the BE and +/- 0.72 log deg(2) for the WE. Inter-device repeatability was modest, +/- 1.09 log deg(2) for the BE and +/- 1.01 log deg(2) for the WE, and a proportional bias was detected. Occasionally, MAIA did not register all the expected number of data points, and included far outliers in the BCEA calculation; the inter-device repeatability did not improve when these outliers were removed. Conclusions Repeatability of 20 s-duration fixation stability examination in patients with central vision loss is specific to the instrument used. We recommend that only data from same type of microperimeter with the same fixation duration should be compared when using fixation stability as an outcome measure to monitor disease progression, effect of treatment or in clinical trials.
引用
收藏
页码:633 / 643
页数:11
相关论文
共 9 条
  • [1] MAIA microperimeter for central vision loss: Repeatability of short-duration fixation stability measurements
    Pyatova, Yulia
    Markowitz, Samuel
    Devenyi, Robert
    Tarita-Nistor, Luminita
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [2] FIXATION STABILITY MEASUREMENT USING THE MP1 MICROPERIMETER
    Crossland, Michael D.
    Dunbar, Hannah M. P.
    Rubin, Gary S.
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (05): : 651 - 656
  • [3] Repeatability of Nidek MP-1 Fixation Measurements in Patients With Bilateral Central Field Loss
    Bedell, Harold E.
    Pratt, Joshua D.
    Krishnan, Arunkumar
    Kisilevsky, Eli
    Brin, Taylor A.
    Gonzalez, Esther G.
    Steinbach, Martin J.
    Tarita-Nistor, Luminita
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (04) : 2624 - 2630
  • [4] Evaluation of fixation stability using different targets with the MP1 microperimeter
    M. Cesareo
    D. Manca
    E. Ciuffoletti
    V. De Giovanni
    F. Ricci
    C. Nucci
    L. Cerulli
    International Ophthalmology, 2015, 35 : 11 - 17
  • [5] Evaluation of fixation stability using different targets with the MP1 microperimeter
    Cesareo, M.
    Manca, D.
    Ciuffoletti, E.
    De Giovanni, V.
    Ricci, F.
    Nucci, C.
    Cerulli, L.
    INTERNATIONAL OPHTHALMOLOGY, 2015, 35 (01) : 11 - 17
  • [6] Intrasession Repeatability of Fixation Stability Assessment with the Nidek MP-1
    Chen, Fred K.
    Patel, Praveen J.
    Xing, Wen
    Crossland, Michael D.
    Bunce, Catey
    Rubin, Gary S.
    Da Cruz, Lyndon
    OPTOMETRY AND VISION SCIENCE, 2011, 88 (06) : 742 - 750
  • [7] Quantification of Fixation Stability by Bivariate Contour Ellipse Area (bcea) Using the Mp1 Microperimeter
    Salvatore, S.
    De Marco, U.
    Marenco, M.
    Vingolo, E. M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [8] Inter-device comparison of retinal sensitivity measurements: the CenterVue MAIA and the Nidek MP-1
    Wong, Evan N.
    Mackey, David A.
    Morgan, William H.
    Chen, Fred K.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 44 (01): : 15 - 23
  • [9] Fixation Stability: A Comparison between the Nidek MP-1 and the Rodenstock Scanning Laser Ophthalmoscope in Persons with and without Diabetic Maculopathy
    Dunbar, Hannah M. P.
    Crossland, Michael D.
    Rubin, Gary S.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (08) : 4346 - 4350