Newborn Incubators Do Not Protect from High Noise Levels in the Neonatal Intensive Care Unit and Are Relevant Noise Sources by Themselves

被引:13
|
作者
Restin, Tanja [1 ,2 ]
Gaspar, Mikael [2 ]
Bassler, Dirk [1 ]
Kurtcuoglu, Vartan [2 ]
Scholkmann, Felix [1 ]
Haslbeck, Friederike Barbara [1 ]
机构
[1] Univ Hosp Zurich, Dept Neonatol, Newborn Res Zurich, CH-8091 Zurich, Switzerland
[2] Univ Zurich, Inst Physiol, CH-8057 Zurich, Switzerland
来源
CHILDREN-BASEL | 2021年 / 8卷 / 08期
关键词
newborn incubators; neonatal intensive care unit; noise; sound; BIRTH-WEIGHT NEWBORNS; SINGLE-FAMILY-ROOM; PRETERM INFANTS; SOUND; SPEECH; DESIGN; EXPOSURE; IMPACT; MUSIC; NICU;
D O I
10.3390/children8080704
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: While meaningful sound exposure has been shown to be important for newborn development, an excess of noise can delay the proper development of the auditory cortex. Aim: The aim of this study was to assess the acoustic environment of a preterm baby in an incubator on a newborn intensive care unit (NICU). Methods: An empty but running incubator (Giraffe Omnibed, GE Healthcare) was used to evaluate the incubator frequency response with 60 measurements. In addition, a full day and night period outside and inside the incubator at the NICU of the University Hospital Zurich was acoustically analyzed. Results: The fan construction inside the incubator generates noise in the frequency range of 1.3-1.5 kHz with a weighted sound pressure level (SPL) of 40.5 dB(A). The construction of the incubator narrows the transmitted frequency spectrum of sound entering the incubator to lower frequencies, but it does not attenuate transient noises such as alarms or opening and closing of cabinet doors substantially. Alarms, as generated by the monitors, the incubator, and additional devices, still pass to the newborn. Conclusions: The incubator does protect only insufficiently from noise coming from the NICU The transmitted frequency spectrum is changed, limiting the impact of NICU noise on the neonate, but also limiting the neonate's perception of voices. The incubator, in particular its fan, as well as alarms from patient monitors are major sources of noise. Further optimizations with regard to the sound exposure in the NICU, as well as studies on the role of the incubator as a source and modulator, are needed to meet the preterm infants' multi-sensory needs.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Internal noise levels in neonatal intensive care unit incubators
    Peixoto, Priscila Vendramini
    Balbino, Flavia Simphronio
    Eliana Moreira Pinheiro, Veridiana Chimirri
    Kakehashi, Tereza Yoshiko
    [J]. ACTA PAULISTA DE ENFERMAGEM, 2011, 24 (03) : 359 - 364
  • [2] Evaluation of noise levels and noise sources in an Irish neonatal intensive care unit
    McCallig, Margaret
    Pakrashi, Vikram
    Durkin, Carmel
    [J]. ANNALS OF WORK EXPOSURES AND HEALTH, 2024, 68 (05) : 550 - 555
  • [3] Neonatal intensive care unit incubators reduce language and noise levels more than the womb
    Brian B. Monson
    Jenna Rock
    Molly Cull
    Vitaliy Soloveychik
    [J]. Journal of Perinatology, 2020, 40 : 600 - 606
  • [4] Neonatal intensive care unit incubators reduce language and noise levels more than the womb
    Monson, Brian B.
    Rock, Jenna
    Cull, Molly
    Soloveychik, Vitaliy
    [J]. JOURNAL OF PERINATOLOGY, 2020, 40 (04) : 600 - 606
  • [5] Noise exposure of the newborn in the incubator at the neonatal intensive care unit
    Bieber, C.
    Thoma, R.
    [J]. KLINISCHE PADIATRIE, 2007, 219 (02): : 105 - 105
  • [6] Innovative instrumentation to strong reduction of the noise levels inside newborn incubators used in the neonatal intensive care units
    Ramos, Paulo Roberto Medina
    Zebende, Gilney Figueira
    Cruz, Juan Alberto Leyva
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (04)
  • [7] The characterization of noise levels in a neonatal intensive care unit and the implications for noise management
    Juan Carlos Fortes-Garrido
    Andres Mauricio Velez-Pereira
    Manuel Gázquez
    Montserrat Hidalgo-Hidalgo
    Juan Pedro Bolívar
    [J]. Journal of Environmental Health Science and Engineering, 12
  • [8] The characterization of noise levels in a neonatal intensive care unit and the implications for noise management
    Carlos Fortes-Garrido, Juan
    Mauricio Velez-Pereira, Andres
    Gazquez, Manuel
    Hidalgo-Hidalgo, Montserrat
    Pedro Bolivar, Juan
    [J]. JOURNAL OF ENVIRONMENTAL HEALTH SCIENCE AND ENGINEERING, 2014, 12
  • [9] Noise levels in a neonatal intensive care unit in the Cape metropole
    Nathan, Lisa M.
    Tuomi, Seppo K.
    Muller, Alida M. U.
    Kirsten, Gert F.
    [J]. SOUTH AFRICAN JOURNAL OF CHILD HEALTH, 2008, 2 (02) : 50 - 54
  • [10] NOISE-LEVELS IN NEONATAL INTENSIVE-CARE UNIT
    VIDYASAGAR, D
    JOSEPH, ME
    HAMILTON, LR
    [J]. JOURNAL OF PEDIATRICS, 1976, 88 (01): : 115 - 115