Clinical validation of a smartphone-based handheld fundus camera for the evaluation of optic nerve head

被引:3
|
作者
Titoneli, Carolina C. [1 ]
Filho, Marcio S. [2 ]
Lencione, Diego [3 ]
Vieira, Flavio Pascoal [3 ]
Stuchi, Jose Augusto [3 ]
Paula, Jayter S. [1 ]
机构
[1] Univ Sao Paulo, Fac Med Ribeirao Preto, Ribeirao Preto, SP, Brazil
[2] Fundacao PIO XII, Barretos, SP, Brazil
[3] Phelcom Technol, Sao Carlos, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Photography/instrumentation; Smartphone; Optic nerve; Telemedicine; GLOBAL PREVALENCE; VISION IMPAIRMENT; INTRAOBSERVER; AGREEMENT; BLINDNESS; DISTANCE; DISK;
D O I
10.5935/0004-2749.20210080
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the quality of retinal images captured with a smartphone-based, handheld fundus camera with that of retinal images captured with a commercial fundus camera and to analyze their agreement in determining the cup-to-disc ratio for a cohort of ophthalmological patients. Methods: A total of 50 patients from a secondary ophthalmic outpatient service center underwent a bilateral fundus examination under mydriasis with a smartphone-based, handheld fundus camera and with a commercial fundus camera (4 images/patient by each). Two experienced ophthalmologists evaluated all the fundus images and graded them on the Likert 1-5 scale for quality. Multivariate regression analyses was then performed to evaluate the factors associated with the image quality. Two masked ophthalmologists determined the vertical cup-to-disc ratio of each fundus image, and both the intraobserver (between devices) and interobserver agreement between them was calculated. Results: Ninety-eight images from 49 patients were processed in this study for their quality analysis. Ten images from five patients (four from commercial fundus camera and one from smartphone-based, handheld fundus camera) were not included in the analyses due to their extremely poor quality. The medians [interquartile interval] of the image quality were not significantly different between those from the smartphone-based, handheld fundus camera and from the commercial fundus camera (4 [4-5] versus 4 [3-4] respectively, p = 0.06); however, both the images captured with the commercial fundus camera and the presence of media opacity presented a significant negative correlation with the image quality. Both the intraobserver [intraclass correlation coefficient (ICC)=0.82, p<0.001 and 0.83, p<0.001, for examiners 1 and 2, respectively] and interobserver (ICC=0.70, p=0.001 and 0.81; p<0.001, for smartphone-based handheld fundus camera and commercial fundus camera, respectively) agreements were excellent and statistically significant. Conclusions: Our results thus indicate that the smartphone-based, handheld fundus camera yields an image quality similar to that from a commercial fundus camera, with significant agreement in the cup-to-disc ratios between them. In addition to the good outcomes recorded, the smartphone-based, handheld fundus camera offers the advantages of portability and low-cost to serve as an alternative for fundus documentation for future telemedicine approaches in medical interventions.
引用
收藏
页码:531 / 537
页数:7
相关论文
共 50 条
  • [1] Smartphone-based evaluation of the optic nerve head
    Pujari, Amar
    Mukhija, Ritika
    Chawla, Rohan
    Phuljhele, Swati
    Saxena, Rohit
    Sharma, Pradeep
    [J]. INDIAN JOURNAL OF OPHTHALMOLOGY, 2018, 66 (11) : 1617 - 1618
  • [2] Undilated versus dilated monoscopic smartphone-based fundus photography for optic nerve head evaluation
    Maximilian W. M. Wintergerst
    Christian K. Brinkmann
    Frank G. Holz
    Robert P. Finger
    [J]. Scientific Reports, 8
  • [3] Undilated versus dilated monoscopic smartphone-based fundus photography for optic nerve head evaluation
    Wintergerst, Maximilian W. M.
    Brinkmann, Christian K.
    Holz, Frank G.
    Finger, Robert P.
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [4] Clinical utility of handheld fundus and smartphone-based camera for monitoring diabetic retinal diseases: a review study
    Naz, Huma
    Nijhawan, Rahul
    Ahuja, Neelu Jyothi
    [J]. INTERNATIONAL OPHTHALMOLOGY, 2024, 44 (01)
  • [5] Clinical utility of handheld fundus and smartphone-based camera for monitoring diabetic retinal diseases: a review study
    Huma Naz
    Rahul Nijhawan
    Neelu Jyothi Ahuja
    [J]. International Ophthalmology, 44
  • [6] Smartphone-based optic nerve head evaluation: mydriasis is what makes the difference
    Wintergerst, Maximilian W. M.
    Brinkmann, Christian Karl
    Finger, Robert P.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [7] Using a smartphone-based digital fundus camera for screening of retinal and optic nerve diseases in veterinary medicine: A preliminary investigation
    Sebbag, Lionel
    Ofri, Ron
    Arad, Dikla
    Handel, Karin W.
    Pe'er, Oren
    [J]. VETERINARY RECORD, 2024, 194 (09)
  • [8] Assessment of a smartphone-based camera for fundus imaging in animals
    Balland, Olivier
    Russo, Andrea
    Isard, Pierre-Francois
    Mathieson, Iona
    Semeraro, Francesco
    Dulaurent, Thomas
    [J]. VETERINARY OPHTHALMOLOGY, 2017, 20 (01) : 89 - 94
  • [9] Remote Screening for Optic Nerve Cupping Using Smartphone-based Nonmydriatic Fundus Photography
    LaMonica, Lauren C.
    Bhardwaj, Mahesh K.
    Hawley, Nicola L.
    Naseri, Take
    Reupena, Muagatutia S.
    Cooper, Michael L.
    Cotran, Paul R.
    Roh, Shiyoung
    Ramsey, David J.
    [J]. JOURNAL OF GLAUCOMA, 2021, 30 (01) : 58 - 60
  • [10] Evaluation of the SVOne: A Handheld, Smartphone-Based Autorefractor
    Ciuffreda, Kenneth J.
    Rosenfield, Mark
    [J]. OPTOMETRY AND VISION SCIENCE, 2015, 92 (12) : 1133 - 1139