Minimizing the effects of magnetization transfer asymmetry on inhomogeneous magnetization transfer (ihMT) at ultra-high magnetic field (11.75 T)

被引:21
|
作者
Prevost, Valentin H. [1 ]
Girard, Olivier M. [1 ]
Varma, Gopal [2 ]
Alsop, David C. [2 ]
Duhamel, Guillaume [1 ]
机构
[1] Aix Marseille Univ, CRMBM, CEMEREM, UMR 7339,CNRS,Fac Med, 27 Blvd Jean Moulin, F-13005 Marseille, France
[2] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Radiol, Boston, MA USA
关键词
Myelin; Inhomogeneous magnetization transfer; ihMT; MT-asymmetry; Dipolar order; Dipolar relaxation time; BROADENED LINES IHMT; CROSS-RELAXATION; HUMAN BRAIN; MULTIPLE-SCLEROSIS; CHEMICAL-EXCHANGE; POTENTIAL MARKER; NMR-SPECTROSCOPY; WATER RESONANCE; MYELIN CONTENT; WHITE-MATTER;
D O I
10.1007/s10334-015-0523-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The recently reported inhomogeneous magnetization transfer technique (ihMT) has been proposed for specific imaging of inhomogeneously broadened lines, and has shown great promise for characterizing myelinated tissues. The ihMT contrast is obtained by subtracting magnetization transfer images obtained with simultaneous saturation at positive and negative frequency offsets (dual frequency saturation experiment, MT (+/-)) from those obtained with single frequency saturation (MT (+)) at the same total power. Hence, ihMT may be biased by MT-asymmetry, especially at ultra-high magnetic field. Use of the average of single positive and negative frequency offset saturation MT images, i.e., (MT (+)+MT (-)) has been proposed to correct the ihMT signal from MT-asymmetry signal. The efficiency of this correction method was experimentally assessed in this study, performed at 11.75 T on mice. Quantitative corrected ihMT and MT-asymmetry ratios (ihMTR and MTRasym) were measured in mouse brain structures for several MT-asymmetry magnitudes and different saturation parameter sets. Our results indicated a "safe" range of magnitudes (/MTRasym/< 4 %) for which MT-asymmetry signal did not bias the corrected ihMT signal. Moreover, experimental evidence of the different natures of both MT-asymmetry and inhomogeneous MT contrasts were provided. In particular, non-zero ihMT ratios were obtained at zero MTRasym values. MTRasym is not a confounding factor for ihMT quantification, even at ultra-high field, as long as MTRasym is restricted to +/- 4 %.
引用
收藏
页码:699 / 709
页数:11
相关论文
共 50 条
  • [1] Minimizing the effects of magnetization transfer asymmetry on inhomogeneous magnetization transfer (ihMT) at ultra-high magnetic field (11.75 T)
    Valentin H. Prevost
    Olivier M. Girard
    Gopal Varma
    David C. Alsop
    Guillaume Duhamel
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2016, 29 : 699 - 709
  • [2] Inhomogeneous magnetization transfer (ihMT): application for multiple sclerosis (MS)
    Van Obberghen, E.
    Girard, O.
    Le Troter, A.
    Soulier, E.
    Viout, P.
    Guye, M.
    Varma, G.
    Alsop, D. C.
    Ranjeva, J. P.
    Pelletier, J.
    Duhamel, G.
    MULTIPLE SCLEROSIS JOURNAL, 2016, 22 : 516 - 517
  • [3] Characterization of the cortical myeloarchitecture with inhomogeneous magnetization transfer imaging (ihMT)
    Munsch, Fanny
    Varma, Gopal
    Taso, Manuel
    Girard, Olivier
    Guidon, Arnaud
    Duhamel, Guillaume
    Alsop, David C.
    NEUROIMAGE, 2021, 225
  • [4] A comparison study of inhomogeneous magnetization transfer (ihMT) and magnetization transfer (MT) in multiple sclerosis based on whole brain acquisition at 3.0 T
    Zhang, Lei
    Wen, Baohong
    Chen, Tao
    Tian, Hongzhe
    Xue, Hongqiang
    Ren, Huipeng
    Li, Li
    Fan, Qing
    Ren, Zhuanqin
    MAGNETIC RESONANCE IMAGING, 2020, 70 : 43 - 49
  • [5] A strategy to reduce the sensitivity of inhomogeneous magnetization transfer (ihMT) imaging to radiofrequency transmit field variations at 3 T
    Soustelle, Lucas
    Troalen, Thomas
    Hertanu, Andreea
    Mchinda, Samira
    Ranjeva, Jean-Philippe
    Guye, Maxime
    Varma, Gopal
    Alsop, David C.
    Duhamel, Guillaume
    Girard, Olivier M.
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (03) : 1346 - 1359
  • [6] Evaluation of the Sensitivity of Inhomogeneous Magnetization Transfer (ihMT) MRI for Multiple Sclerosis
    Van Obberghen, E.
    Mchinda, S.
    le Troter, A.
    Prevost, V. H.
    Viout, P.
    Guye, M.
    Varma, G.
    Alsop, D. C.
    Ranjeva, J. -P.
    Pelletier, J.
    Girard, O.
    Duhamel, G.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (04) : 634 - 641
  • [7] Validating the sensitivity of inhomogeneous magnetization transfer (ihMT) MRI to myelin with fluorescence microscopy
    Duhamel, G.
    Prevost, V. H.
    Cayre, M.
    Hertanu, A.
    Mchinda, S.
    Carvalho, V. N.
    Varma, G.
    Durbec, P.
    Alsop, D. C.
    Girard, O. M.
    NEUROIMAGE, 2019, 199 : 289 - 303
  • [8] Inhomogeneous Magnetization Transfer (ihMT) imaging in the acute cuprizone mouse model of demyelination/remyelination
    Hertanu, Andreea
    Soustelle, Lucas
    Buron, Julie
    Le Priellec, Julie
    Cayre, Myriam
    Le Troter, Arnaud
    Prevost, Valentin H.
    Ranjeva, Jean-Philippe
    Varma, Gopal
    Alsop, David C.
    Durbec, Pascale
    Girard, Olivier M.
    Duhamel, Guillaume
    NEUROIMAGE, 2023, 265
  • [9] Reproducibility of inhomogeneous magnetization transfer (ihMT): A test-retest, multi-site study
    Zhang, Lei
    Chen, Tao
    Tian, Hongzhe
    Xue, Hongqiang
    Ren, Huipeng
    Li, Li
    Fan, Qing
    Wen, Baohong
    Ren, Zhuanqin
    MAGNETIC RESONANCE IMAGING, 2019, 57 : 243 - 249
  • [10] Inhomogeneous magnetization transfer (ihMT) as a new sensitive marker of the microstructural spinal cord changes in multiple sclerosis
    Demortiere, S.
    Rasoanandrianina, H.
    Trabelsi, A.
    Girard, O.
    Duhamel, G.
    Guye, M.
    Ranjeva, J. -P.
    Pelletier, J.
    Audoin, B.
    Callot, V.
    MULTIPLE SCLEROSIS JOURNAL, 2018, 24 : 423 - 423