Jugular bulb temperature compared with noninvasive temperatures and cerebral arteriovenous oxygen saturation differences during open heart surgery

被引:9
|
作者
Sandström, K [1 ]
Nilsson, K [1 ]
Andréasson, S [1 ]
Larsson, LE [1 ]
机构
[1] Sahlgrens Univ Hosp, Dept Paediat Anaesthesia & Intens Care, SE-41685 Gothenburg, Sweden
来源
PAEDIATRIC ANAESTHESIA | 1999年 / 9卷 / 02期
关键词
cardiac surgery; cardiopulmonary bypass; hypothermia; temperature; cerebral arteriovenous oxygen saturation difference;
D O I
10.1046/j.1460-9592.1999.9220326.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Limited information is available about the correlation between cerebral temperature and routine temperature measurements during cardiopulmonary bypass in infants. Nasopharyngeal, tympanic membrane and rectal temperatures were compared with jugular bulb temperature in ten infants operated on with moderate or deep hypothermia. The cerebral arteriovenous saturation differences were correlated with the temperatures at the four measurement sites. The jugular bulb and nasopharyngeal temperatures showed the most rapid response during cooling and rewarming. The tympanic temperature response varied in an unpredictable way. Rectal temperature, which was the target for rewarming, lagged behind during both cooling and rewarming. Overwarming at the end of cardiopulmonary bypass, seen as jugular bulb and nasopharyngeal temperatures exceeding 38 degrees C, was common after deep hypothermia. A high correlation was found between the cerebral arteriovenous oxygen saturation differences and the jugular bulb temperature (r= 0.81) and the nasopharyngeal and the tympanic temperature (r = 0.79), whereas the correlation with rectal temperature was weaker (0.66).
引用
收藏
页码:123 / 128
页数:6
相关论文
共 34 条
  • [1] Jugular venous bulb oxygen saturation monitoring in arteriovenous malformation surgery
    WilderSmith, OHG
    Fransen, P
    deTribolet, N
    Tassonyi, E
    [J]. JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 1997, 9 (02) : 162 - 165
  • [2] Cerebral Oxygenation during Laparoscopic Surgery: Jugular Bulb versus Regional Cerebral Oxygen Saturation
    Choi, Seung Ho
    Kim, Soo Hwan
    Lee, Sung Jin
    Soh, Sa Rah
    Oh, Young Jun
    [J]. YONSEI MEDICAL JOURNAL, 2013, 54 (01) : 225 - 230
  • [3] Jugular venous bulb oxygen saturation compared with cerebral saturation measured by near infrared spectroscopy.
    OwenReece, H
    Smith, M
    Elwell, CE
    Lovell, AT
    Goldstone, J
    [J]. ANESTHESIA AND ANALGESIA, 1997, 84 : S370 - S370
  • [4] The Correlation of Two Cerebral Saturation Monitors With Jugular Bulb Oxygen Saturation in Children Undergoing Cardiopulmonary Bypass for Congenital Heart Surgery
    Naguib, Aymen N.
    Winch, Peter D.
    Sebastian, Roby
    Gomez, Daniel
    Guzman, Luisa
    Rice, Julie
    Tumin, Dmitry
    Galantowicz, Mark
    Tobias, Joseph D.
    [J]. JOURNAL OF INTENSIVE CARE MEDICINE, 2017, 32 (10) : 603 - 608
  • [5] Effect of slow versus rapid rewarming on jugular bulb oxygen saturation in adult patients undergoing open heart surgery
    Ali, Mohmaed Shaaban
    Sayed, Sameh Abd Al Rhman
    Mohamoud, Maged Salah
    Abd-Elshafy, Sayed Kaoud
    Almaz, Mohamed Gomaa
    [J]. SAUDI JOURNAL OF ANAESTHESIA, 2014, 8 (02) : 178 - 182
  • [6] Cerebral oxygen saturation monitoring during laparoscopic surgery under sevoflurane anesthesia: jugular bulb oxygen saturation versus near-infrared spectroscopy
    Lee, Woo Kyung
    Oh, Young Jun
    Shin, Yang-Sik
    Lee, Jeong Rim
    Choi, Seung Ho
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 : 125 - 125
  • [7] Reliability of continuous jugular venous bulb hemoglobin oxygen saturation during cardiac surgery
    Anastasiou, E
    Gerolioliou, K
    Karakoulas, K
    Peftoulidou, M
    Giala, M
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1999, 13 (03) : 276 - 279
  • [8] CONTINUOUS MONITORING OF JUGULAR BULB OXYGEN-SATURATION AS A MEASURE OF THE SHUNT FLOW OF CEREBRAL ARTERIOVENOUS-MALFORMATIONS
    KATAYAMA, Y
    TSUBOKAWA, T
    HIRAYAMA, T
    HIMI, K
    [J]. JOURNAL OF NEUROSURGERY, 1994, 80 (05) : 826 - 833
  • [9] Jugular bulb oxygen saturation and middle cerebral blood flow velocity during cardiopulmonary bypass
    vonKnobelsdorff, G
    Hanel, F
    Werner, C
    Esch, JSA
    [J]. JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 1997, 9 (02) : 128 - 133
  • [10] Jugular bulb saturation monitoring during emergence from anesthesia for elective cerebral tumor surgery
    Geelen, J.
    De Deyne, C.
    Wuyts, J.
    Vandermeersch, E.
    Heylen, R.
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 : 99 - 99