Microcomputed Tomography Evaluation of the Marginal and Internal Fit of Crown and Inlay Restorations Fabricated via Different Digital Scanners Belonging to the Same CAD/CAM System

被引:11
|
作者
Ekici, Zeynep [1 ]
Kilicarslan, Mehmet Ali [1 ]
Bilecenoglu, Burak [2 ]
Ocak, Mert [3 ]
机构
[1] Ankara Univ, Fac Dent, Dept Prosthodont, Ankara, Turkey
[2] Ankara Medipol Univ, Fac Med, Dept Anat, Ankara, Turkey
[3] Ankara Univ, Fac Dent, Dept Anat, Ankara, Turkey
关键词
CERAMIC INLAYS; CLINICAL-PERFORMANCE; CEREC; ADAPTATION; IMPRESSIONS; THICKNESS; ACCURACY;
D O I
10.11607/ijp.6822
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To evaluate the marginal and internal adaptation of all-ceramic crowns and inlays fabricated using different scanners of the same CAD/CAM system. Materials and Methods: All-ceramic crown preparations were performed on typodont maxillary first premolars, and mandibular first molars were prepared for ceramic Class II mesio-occlusal inlays. Two intraoral scanners (CEREC Bluecam and Omnicam, Dentsply Sirona) and one model scanner (CEREC inEos X5, Dentsply Sirona) were used to scan the preparations. All restorations were fabricated by milling single-feldspathic ceramic blocks (CEREC Blocs). The marginal and internal discrepancies of restorations were evaluated via microcomputed tomography (micro-CT) analyses. Results: For linear crown measurements, the marginal gaps were 63.75 mu m, 88.24 mu m, and 90.89 mu m for Bluecam, Omnicam, and inEos X5, respectively. For crowns at central groove areas, the maximum values for Bluecam, Omnicam, and inEos X5 were found to be 144.78 mu m, 165.19 mu m, and 129.49 mu m, respectively. For inlays, the highest range at the midpoint of the axio-pulpal line angle for Bluecam, Omnicam, and inEos X5 were determined as 138.57 mu m, 184.33 pm, and 179.71 pm, respectively. In volumetric measurements, inEos X5 showed lower gaps for both crowns (11.47 mm(3)) and inlays (5.65 mm(3)) compared to both intraoral scanners. These results are within a clinically acceptable range. Conclusion: When all-ceramic crowns were evaluated, there were generally no significant differences found among scanners for the regional linear measurements, but more volumetric gaps occurred in restorations obtained with intraoral scanners. On the other hand, when the inlay restorations were evaluated, significant differences were found between groups except for the midpoint of the axio-pulpal line angle and the midpoint of the mesio-gingival margin. However, the marginal and internal gaps of both crowns and inlays presented mean values < 150 mu m in many surfaces, which could be considered clinically acceptable.
引用
收藏
页码:381 / 389
页数:9
相关论文
共 50 条
  • [1] Effect of different preparation finishing procedures on the marginal and internal fit of CAD-CAM-produced restorations: A microcomputed tomography evaluation
    Tamam, Evsen
    Gungor, Merve Bankoglu
    Nemli, Secil Karakoca
    Ocak, Mert
    Bilecenog, Burak
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2023, 130 (04): : 605 - 610
  • [2] Effect of different intraoral scanning strategies on the marginal and internal fit of CAD-CAM inlay restorations: An in vitro study
    Salem, Mohamed M.
    Elmahy, Waleed A.
    Nasr, Dina M.
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (03):
  • [3] EVALUATION OF THE MARGINAL FIT OF CAD/CAM CROWNS USING CBCT AND DIGITAL SCANNERS
    Shenoy, Amrutha
    Nallaswamy, Deepak
    Maiti, Subhabrata
    [J]. ANNALS OF DENTAL SPECIALTY, 2023, 11 (03): : 37 - 44
  • [4] Marginal and internal fit of CAD-CAM inlay/onlay restorations: A systematic review of in vitro studies
    Goujat, Alexis
    Abouelleil, Hazem
    Colon, Pierre
    Jeannin, Christophe
    Pradelle, Nelly
    Seux, Dominique
    Grosgogeat, Brigitte
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2019, 121 (04): : 590 - +
  • [5] Effect of Chairside CAD/CAM Restoration Type on Marginal Fit Accuracy: A Comparison of Crown, Inlay and Onlay Restorations
    Merrill, Tyson C.
    Mackey, Thomas
    Luc, Raymond
    Lung, Dominique
    Naseem, Aneela
    Abduo, Jaafar
    [J]. EUROPEAN JOURNAL OF PROSTHODONTICS AND RESTORATIVE DENTISTRY, 2021, 29 (02): : 119 - 127
  • [6] Evaluation of marginal fit of CAD/CAM restorations fabricated through cone beam computerized tomography and laboratory scanner data
    Seker, Emre
    Ozcelik, Tuncer Burak
    Rathi, Nakul
    Yilmaz, Burak
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2016, 115 (01): : 47 - 51
  • [7] Clinical Marginal and Internal Fit of Crowns Fabricated Using Different CAD/CAM Technologies
    Huang, Zhuoli
    Zhang, Lu
    Zhu, Jingwei
    Zhao, Yiwei
    Zhang, Xiuyin
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2015, 24 (04): : 291 - 295
  • [8] Influence of preparation angle on marginal and internal fit of CAD/CAM-fabricated zirconia crown copings
    Beuer, Florian
    Aggstaller, Hans
    Richter, Juliane
    Edelhoff, Daniel
    Gernet, Wolfgang
    [J]. QUINTESSENCE INTERNATIONAL, 2009, 40 (03): : 243 - 250
  • [9] Evaluation of Marginal and Internal Fit of CAD/CAM Endocrowns with Different Cavity Tapers
    Hajimahmoodi, Mohammedreza
    Raseipour, Sasan
    Mroue, Mohammad
    Ghodsi, Safoura
    [J]. INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2023, 36 (02) : 189 - 193
  • [10] MARGINAL AND INTERNAL ADAPTATION OF CERAMIC CROWN RESTORATIONS FABRICATED WITH CAD/CAM TECHNOLOGY AND THE HEAT-PRESS TECHNIQUE
    Mously, Hisham A.
    Finkelman, Matthew
    Zandparsa, Roya
    Hirayama, Hiroshi
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2014, 112 (02): : 249 - 256