Comparison of a reusable with a disposable vessel-sealing device in a sheep model: efficacy and costs

被引:15
|
作者
Klar, Maximilian [1 ]
Haberstroh, Joerg [2 ]
Timme, Sylvia [3 ]
Fritzsch, Gernod [4 ]
Gitsch, Gerald [1 ]
Denschlag, Dominik [1 ]
机构
[1] Univ Freiburg, Sch Med, Dept Obstet & Gynecol, D-79106 Freiburg, Germany
[2] Univ Freiburg, Sch Med, BioMed Ctr, Div Expt Surg, D-79106 Freiburg, Germany
[3] Univ Freiburg, Sch Med, Dept Pathol, D-79106 Freiburg, Germany
[4] KLS Martin Grp, Tuttlingen, Germany
关键词
Vessel sealing; reusable device; efficacy; cost; animal model; in vivo;
D O I
10.1016/j.fertnstert.2010.09.014
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective: To compare the efficacy of the new reusable vessel-sealing device MarSeal (KLS Martin, Tuttlingen, Germany) with the conventional standard disposable LigaSure device (Covidien-Valleylab, Boulder, CO) in an in vivo animal model. Design: Prospective animal study. Setting: Center of Experimental Surgery at a university hospital. Animals: Thirteen sheep. Intervention(s): The carotid and femoral arteries were dissected bilaterally and sealed randomly with MarSeal unilaterally and consecutively with LigaSure contralaterally in vivo. Afterward the burst pressure was determined for each artery. Main Outcome Measure(s): Sealing time, failure rate, and burst pressure. Result(s): The mean diameter of all arteries sealed did not differ significantly between the two groups (MarSeal 5.40 mm vs. LigaSure 5.35 mm). The mean sealing time was significantly shorter with use of the reusable device (MarSeal 5.2 seconds vs. LigaSure 9.1 seconds). We did not find a significant difference in sealing failure rates between the groups (MarSeal 7.0% vs. LigaSure 9.1%). In addition, analysis of mean burst pressure did not reveal a significant difference between the different devices (MarSeal 429 mm Hg vs. LigaSure 484 mm Hg). There was no significant difference with respect to lateral thermal damage (MarSeal 0.91 cm vs. LigaSure 0.93 cm). Conclusion(s): In our in vivo animal study, the new reusable MarSeal device appears to be equivalently effective for vessel sealing when compared with the conventional disposable LigaSure device. (Fertil Steril (R) 2011;95:795-8. (C)2011 by American Society for Reproductive Medicine.)
引用
收藏
页码:795 / 798
页数:4
相关论文
共 50 条
  • [1] Efficacy and quality of vessel sealing - Comparison of a reusable with a disposable device and effects of clamp surface geometry and structure
    Richter, S
    Kollmar, O
    Schilling, MK
    Pistorius, GA
    Menger, MD
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (06): : 890 - 894
  • [2] Bipolar Electrosurgical Vessel-Sealing Device With Compressive Force Monitoring
    Chen, Roland K.
    Chastagner, Matthew W.
    Geiger, James D.
    Shih, Albert J.
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (06):
  • [3] Use of a bipolar vessel-sealing device in resection of canine insulinoma
    Wouters, E. G. H.
    Buishand, F. O.
    Kik, M.
    Kirpensteijn, J.
    JOURNAL OF SMALL ANIMAL PRACTICE, 2011, 52 (03) : 139 - 145
  • [4] Evaluation of a vessel-sealing device for use in laparoscopic ovariectomy in mares
    Hand, R
    Rakestraw, P
    Taylor, T
    VETERINARY SURGERY, 2002, 31 (03) : 240 - 244
  • [5] Retroperitoneal Lumboaortic Lymphadenectomy Using a Vessel-Sealing Device in 10 Steps
    Schaub, Marie
    Lecointre, Lise
    Faller, Emilie
    Boisrame, Thomas
    Wattiez, Arnaud
    Baldauf, Jean-Jacques
    Akladios, Cherif
    JOURNAL OF MINIMALLY INVASIVE GYNECOLOGY, 2018, 25 (05) : 765 - 766
  • [6] Comparison of Two Radiofrequency Vessel-Sealing Device for Laparoscopic Ovariectomy in African Lionesses (Panthera Leo)
    Lacitignola, Luca
    Laricchiuta, Pietro
    Guadalupi, Marta
    Stabile, Marzia
    Scardia, Annalaura
    Cinone, Mario
    Staffieri, Francesco
    ANIMALS, 2022, 12 (18):
  • [7] Use of a novel vessel-sealing device for peripheral lung biopsy and lung lobectomy in a cadaveric model
    Bruckner, M.
    Heblinski, N.
    Henrich, M.
    JOURNAL OF SMALL ANIMAL PRACTICE, 2019, 60 (07) : 411 - 416
  • [8] Laparoscopic liver resection using a bipolar vessel-sealing device: LigaSure®
    Slakey, Douglas P.
    HPB, 2008, 10 (04) : 253 - 255
  • [9] Evaluation of a vessel-sealing device to assist eye enucleation and exenteration in the horse
    Launois, M. T.
    Vandeweerd, J. M. E. F.
    Perrin, R. A. R.
    Brogniez, L.
    Gabriel, A.
    Coomer, R.
    Clegg, P. D.
    Desbrosse, A. M.
    Desbrosse, F. G.
    EQUINE VETERINARY EDUCATION, 2009, 21 (11) : 596 - 601
  • [10] Clinical significance of Vessel-sealing Device Usage for Pancreatectomy: A Retrospective Cohort Study
    Nanashima, Atsushi
    Abo, Takafumi
    Takagi, Katsunori
    Wada, Hideo
    Arai, Junichi
    Kunizaki, Masaki
    To, Kazuo
    Takeshita, Hiroaki
    Hidaka, Shigekazu
    Nagayasu, Takeshi
    HEPATO-GASTROENTEROLOGY, 2014, 61 (134) : 1767 - 1774