Dynamic insertion study on cochlear implant electrodes with x-ray microscopy visualization.

被引:2
|
作者
Hüttenbrink, KB [1 ]
Zahnert, T [1 ]
Hofmann, G [1 ]
Kronholz, G [1 ]
Jolly, C [1 ]
机构
[1] Tech Univ Dresden, Univ Klinikum Carl Gustav Carus, Klin & Poliklin HNO Heilkunde Kopf & Halschirurg, Dresden, Germany
来源
LARYNGO-RHINO-OTOLOGIE | 2001年 / 80卷 / 02期
关键词
X-ray fluoroscopy; X-ray microscopy; cochlear implant electrodes; perimodiolar electrodes;
D O I
10.1055/s-2001-11897
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Background: The movements of the electrode cable of a cochlear implant inside the cochlea during the insertion procedure generally are invisible, even in temporal bone experiments. Yet, the development of new designs of electrodes, their positioning near to the modiolus etc, requires an exact knowledge on the dynamic behaviour of the electrodes inside the cochlea. The exclusive method for looking through the undamaged cochlea walls is by x-ray technique. Methods: A specific x-ray tube with a focal spot size of barely 5 mum allows useful direct magnifications of more than 20-30 times, We performed temporal bone experiments with this tube and studied the behaviour of conventional and perimodiolar electrode cables of a MedEl-CI during the insertion procedure under fluoroscopic viewing. Results: The high resolution imaging revealed the influence of stiffness of the electrode cable on the insertion depth of a conventional electrode. if the angle of the direction of insertion was too steep, the resulting kinking of the cable at the medial wail of the basal turn was instantly visible. The direct visualisation of the movements of the cable was useful in optimizing the design and stiffness of various prototypes of perimodiolar electrodes for a deep insertion. Conclusions: The dynamic examination technique disclosed that the stiffness of actual electrode cables is not optimally adapted to the form of the cochlear cavity for deep insertion. The non-destructive visualisation technique will facilitate the development of advanced electrode design, especially for various concepts of perimodiolar electrodes.
引用
收藏
页码:66 / 70
页数:5
相关论文
共 50 条
  • [1] Movements of cochlear implant electrodes inside the cochlea during insertion:: An X-ray Microscopy study
    Hüttenbrink, KB
    Zahnert, T
    Jolly, C
    Hofmann, G
    [J]. OTOLOGY & NEUROTOLOGY, 2002, 23 (02) : 187 - 191
  • [2] Effectiveness of skull X-RAY to determine cochlear implant insertion depth
    Vinay Fernandes
    Yiqiao Wang
    Robert Yeung
    Sean Symons
    Vincent Lin
    [J]. Journal of Otolaryngology - Head & Neck Surgery, 47
  • [3] Effectiveness of skull X-RAY to determine cochlear implant insertion depth
    Fernandes, Vinay
    Wang, Yiqiao
    Yeung, Robert
    Symons, Sean
    Lin, Vincent
    [J]. JOURNAL OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2018, 47
  • [4] Visualization of Rapid Dynamic Interactions by Flash Soft X-Ray Microscopy
    Menoni, C. S.
    Carbajo, S.
    Howlett, I. D.
    Chao, W.
    Anderson, E. H.
    Vinogradov, A. V.
    Artyukov, I. A.
    Buchanan, K.
    Marconi, M. C.
    Rocca, J. J.
    [J]. X-RAY LASERS 2012, 2014, 147 : 185 - 192
  • [5] Potential insertion complications with cochlear implant electrodes
    Ishiyama, Akira
    Risi, Frank
    Boyd, Paul
    [J]. COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (04) : 206 - 219
  • [6] Impact of the Insertion Speed of Cochlear Implant Electrodes on the Insertion Forces
    Kontorinis, Georgios
    Lenarz, Thomas
    Stoever, Timo
    Paasche, Gerrit
    [J]. OTOLOGY & NEUROTOLOGY, 2011, 32 (04) : 565 - 570
  • [7] Hearing results with deep insertion of cochlear implant electrodes
    Hodges, AV
    Villasuso, E
    Balkany, T
    Bird, PA
    Butts, S
    Lee, D
    Gomez, O
    [J]. AMERICAN JOURNAL OF OTOLOGY, 1999, 20 (01): : 53 - 55
  • [8] Force measurement at the insertion process of cochlear implant electrodes
    Majdani, Omid
    Kobler, Jan-Phillipp
    Beckmann, Daniel
    Ortmaier, Tobias
    Lenarz, Thomas
    Rau, Thomas
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 : 438 - 438
  • [9] Factors Associated with Incomplete Insertion of Electrodes in Cochlear Implant Surgery: A Histopathologic Study
    Lee, Joonhan
    Nadol, Joseph B., Jr.
    Eddington, Donald K.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2011, 16 (02) : 69 - 81
  • [10] Round Window Membrane Insertion With Perimodiolar Cochlear Implant Electrodes
    Coordes, Annekatrin
    Ernst, Arne
    Brademann, Goetz
    Todt, Ingo
    [J]. OTOLOGY & NEUROTOLOGY, 2013, 34 (06) : 1027 - 1032