Validation of a new piezoelectric device for noninvasive measurement of central aortic systolic blood pressure

被引:3
|
作者
Cuffaro, Paula E. [1 ]
Morales, Margarita S. [1 ]
Barochiner, Jessica [1 ]
Rada, Marcelo A. [1 ]
Alfie, Jose [1 ]
Aparicio, Lucas S. [1 ]
Galarza, Carlos R. [1 ]
Micali, Ruben G. [1 ]
Marin, Marcos J. [1 ]
Waisman, Gabriel D. [1 ]
机构
[1] Hosp Italiano Buenos Aires, Dept Internal Med, Hypertens Sect, Juan D Peron 4190,C1181ACH, Buenos Aires, DF, Argentina
关键词
aortic pressure; aortic; blood pressure; pulse wave analysis; SphygmoCor; transfer function;
D O I
10.1097/MBP.0000000000000298
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The objective of this study was to compare the aortic piezoelectric device for noninvasive measurement of central aortic systolic blood pressure (cSBP) with the SphygmoCor. Participants and methods A total of 85 participants from both sexes, aged 18-80 years, were stratified into three age groups (< 30, 30-60, > 60 years), with an equal number of healthy volunteers and hypertensive patients. We performed three cSBP measurements with each device, in an alternate manner, using the Bland-Altman method to determine the level of agreement. The standard of the Association for the Advancement of Medical Instrumentation for brachial blood pressure evaluation was used for the comparison. Results The mean cSBPs were 109.3 +/- 12.05 and 109.0 +/- 12.2 mmHg with the SphygmoCor and the Aortic device, respectively, showing a strong correlation (r = 0.98, P < 0.001). A mean difference of 0.35 +/- 2.43 mmHg (95% confidence interval: 0.17-0.87, P=NS) was obtained with the Bland-Altman method. The 95% limits of agreement was -4.4 to + 5.1 mmHg. Conclusion Complying with the Association for the Advancement of Medical Instrumentation criteria, cSBP measurements obtained with the Aortic and the SphygmoCor devices are equivalent. Copyright (C) 2018 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:49 / 51
页数:3
相关论文
共 50 条
  • [1] METHOD FOR NONINVASIVE MEASUREMENT OF CENTRAL AORTIC SYSTOLIC PRESSURE
    MARMOR, AT
    BLONDHEIM, DS
    GOZLAN, E
    NAVO, E
    FRONT, D
    CLINICAL CARDIOLOGY, 1987, 10 (04) : 215 - 221
  • [2] Is Noninvasive Brachial Systolic Blood Pressure an Accurate Estimate of Central Aortic Systolic Blood Pressure?
    Shih, Yuan-Ta
    Cheng, Hao-Min
    Sung, Shih-Hsien
    Chuang, Shao-Yuan
    Hu, Wei-Chih
    Chen, Chen-Huan
    AMERICAN JOURNAL OF HYPERTENSION, 2016, 29 (11) : 1283 - 1291
  • [3] Measurement of Central Systolic Blood Pressure by Pulse Volume Plethysmography With a Noninvasive Blood Pressure Monitor
    Sung, Shih-Hsien
    Cheng, Hao-Min
    Chuang, Shao-Yuan
    Shih, Yuan-Ta
    Wang, Kang-Ling
    Chen, Ying-Hwa
    Lin, Shing-Jong
    Yu, Wen-Chung
    Chen, Chen-Huan
    AMERICAN JOURNAL OF HYPERTENSION, 2012, 25 (05) : 542 - 548
  • [4] Validation of Devices and Methods for Noninvasive Estimation of Central Aortic Blood Pressure in Children
    Papaioannou, Theodore G.
    Karageorgopoulou, Theofani
    Stefanadis, Christodoulos
    Tousoulis, Dimitrios
    HYPERTENSION, 2015, 66 (02) : E7 - E7
  • [5] Validity of noninvasive central aortic pressure measurement
    O'Rourke, Michael F.
    Takazawa, Kenji
    Tanaka, Nobuhiro
    JOURNAL OF HYPERTENSION, 2019, 37 (11) : 2300 - 2301
  • [6] Calibration of noninvasive central blood pressure devices and negative aortic-to-brachial systolic pressure amplification
    Papaioannou, Theodore G.
    Vavuranakis, Manolis
    Tousoulis, Dimitrios
    KIDNEY INTERNATIONAL, 2017, 91 (01) : 253 - 254
  • [7] Calibration of noninvasive central blood pressure devices and negative aortic-to-brachial systolic pressure amplification Reply
    Carlsen, Rasmus K.
    Peters, Christian D.
    Khatir, Dinah S.
    Laugesen, Esben
    Botker, Hans Erik
    Winther, Simon
    Buus, Niels H.
    KIDNEY INTERNATIONAL, 2017, 91 (01) : 254 - 254
  • [8] VALIDATION OF A NEW CENTRAL AORTIC PRESSURE MEASUREMENT EQUIPMENT AORTIC COMPARED WITH SPHYGMOCOR
    Cuffaro, P.
    Morales, M.
    Rada, M.
    Galarza, C.
    Alfie, J.
    Marin, M.
    Micali, R.
    Barochiner, J.
    Aparicio, L.
    Waisman, G.
    JOURNAL OF HYPERTENSION, 2017, 35 : E246 - E246
  • [9] Response to Validation of Devices and Methods for Noninvasive Estimation of Central Aortic Blood Pressure in Children
    Milne, Laura
    Keehn, Louise
    Guilcher, Antoine
    Reidy, John F.
    Karunanithy, Narayan
    Rosenthal, Eric
    Qureshi, Shakeel
    Chowienczyk, Phil J.
    Sinha, Manish D.
    HYPERTENSION, 2015, 66 (02) : E8 - E8
  • [10] Validity of noninvasive central aortic pressure measurement Reply
    Sharman, James E.
    Picone, Dean S.
    Peng, Xiaoqing
    Black, J. Andrew
    Dwyer, Nathan
    Roberts-Thomson, Philip
    Schultz, Martin G.
    JOURNAL OF HYPERTENSION, 2019, 37 (11) : 2301 - 2301