High-heat-flux sensor calibration using black-body radiation

被引:26
|
作者
Murthy, AV [1 ]
Tsai, BK
Saunders, RD
机构
[1] AeroTech Inc, Hampton, VA 23666 USA
[2] Natl Inst Stand & Technol, Opt Technol Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/35/4/50
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper deals with the radiative calibration aspects of high-heat-flux sensors using black-body radiation. In the last two years, several heat-flux sensors were calibrated up to 50 kW/m(2) using a 25 mm diameter aperture variable-temperature black body and a reference room-temperature electrical-substitution radiometer. Tests on a typical Schmidt-Boelter heat-flux sensor showed long-term repeatability of calibration is within +/-0.6%. Plans for extending the present calibration capability to 100 kW/m(2) are discussed.
引用
收藏
页码:501 / 504
页数:4