Verification of an articulated arm coordinate measuring machine using a laser tracker as reference equipment and an indexed metrology platform

被引:23
|
作者
Acero, R. [1 ]
Brau, A. [2 ]
Santolaria, J. [3 ]
Pueo, M. [1 ]
机构
[1] Ctr Univ Defensa, Acad Gen Militar, Zaragoza 50090, Spain
[2] Univ Sonora, Dept Ind Engn, Hermosillo 83000, Sonora, Mexico
[3] Univ Zaragoza, Dept Design & Mfg Engn, Zaragoza 50018, Spain
关键词
Laser tracker; Articulated arm coordinate measuring machine; Verification; Calibration; Gauge; MEASUREMENT UNCERTAINTY; CALIBRATION; ARRANGEMENT; ERROR; MULTILATERATION; PERFORMANCE; DESIGN; SENSOR; MODEL;
D O I
10.1016/j.measurement.2015.03.023
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In this work the analysis of the use of a laser tracker as a reference instrument in calibration and verification procedures for articulated arm coordinate measuring machines (AACMM) with an indexed metrology platform is presented. In the case of AACMM verification procedures, where it is necessary to evaluate its maximum working volume, this technique represents an alternative to conventional one-dimensional gauges and avoids the need of materializing the lengths required for a conventional gauge by increasing the flexibility for defining test positions and broadening the definition of reference test lengths depending on equipment calibration or verification requirements. First a procedure for the AACMM verification with the indexed metrology platform (IMP) and the laser tracker (LT) used as a reference instrument was defined and the required accuracy of the equipment was assessed. The experimental testing to validate its proper application was carried out by means of laser tracker measurements of six retroreflector targets located in a mesh that simulated the AACMM working volume. Finally, the results obtained with both equipment, laser tracker and AACMM with indexed metrology platform, showed that the procedure presented is suitable and that the laser tracker could be considered as a reference instrument for AACMM verification processes when the nominal accuracy of the laser tracker is assured. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:52 / 63
页数:12
相关论文
共 50 条
  • [1] Evaluation of the use of a laser tracker and an indexed metrology platform as gauge equipment in articulated arm coordinate measuring machine verification procedures
    Acero, R.
    Brau, A.
    Santolaria, J.
    Pueo, M.
    Cajal, C.
    [J]. MESIC MANUFACTURING ENGINEERING SOCIETY INTERNATIONAL CONFERENCE 2015, 2015, 132 : 740 - 747
  • [2] Verification of a laser tracker with an indexed metrology platform
    Acero, R.
    Santolaria, J.
    Pueo, M.
    Brau, A.
    [J]. INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2016, 84 (1-4): : 595 - 606
  • [3] Verification of a laser tracker with an indexed metrology platform
    R. Acero
    J. Santolaria
    M. Pueo
    A. Brau
    [J]. The International Journal of Advanced Manufacturing Technology, 2016, 84 : 595 - 606
  • [4] Articulated Arm Coordinate Measuring Machine Calibration by Laser Tracker Multilateration
    Santolaria, Jorge
    Majarena, Ana C.
    Samper, David
    Brau, Agustin
    Velazquez, Jesus
    [J]. SCIENTIFIC WORLD JOURNAL, 2014,
  • [5] Kinematic parameter identification procedure of an articulated arm coordinate measuring machine based on a metrology platform
    Brau-Avila, A.
    Acero, R.
    Santolaria, J.
    Valenzuela-Galvan, M.
    Icasio-Hernandez, Octavio
    [J]. INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2019, 104 (1-4): : 1027 - 1040
  • [6] Kinematic parameter identification procedure of an articulated arm coordinate measuring machine based on a metrology platform
    A. Brau-Avila
    R. Acero
    J. Santolaria
    M. Valenzuela-Galvan
    Octavio Icasio-Hernández
    [J]. The International Journal of Advanced Manufacturing Technology, 2019, 104 : 1027 - 1040
  • [7] Kinematic Modeling and Verification of an Articulated Arm Coordinate Measuring Machine
    Zhang, Huaishan
    Gao, Guanbin
    Wang, Wen
    Na, Jing
    Wu, Xing
    [J]. SEVENTH INTERNATIONAL SYMPOSIUM ON PRECISION MECHANICAL MEASUREMENTS, 2016, 9903
  • [8] Application of virtual distances methodology to laser tracker verification with an indexed metrology platform
    Acero, R.
    Santolaria, J.
    Pueo, M.
    Aguilar, J. J.
    Brau, A.
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2015, 26 (11)
  • [9] Uncertainty estimation of an indexed metrology platform for the verification of portable coordinate measuring instruments
    Acero, R.
    Santolaria, J.
    Pueo, M.
    Abad, J.
    [J]. MEASUREMENT, 2016, 82 : 202 - 220
  • [10] Standard conforming involute gear metrology using an articulated arm coordinate measuring system
    Haertig, Frank
    Lin, Hu
    Kniel, Karin
    Shi, Zhaoyao
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (10)