PET imaging for treatment verification of ion therapy: Implementation and experience at GSI Darmstadt and MGH Boston

被引:57
|
作者
Parodi, Katia [1 ]
Bortfeld, Thomas [2 ]
Enghardt, Wolfgang [3 ,4 ]
Fiedler, Fine [3 ]
Knopf, Antje [2 ]
Paganetti, Harald [2 ]
Pawelke, Joerg [3 ]
Shakirin, Georgy [3 ]
Shih, Helen [2 ]
机构
[1] Heidelberg Ion Beam Therapy Ctr, D-69120 Heidelberg, Germany
[2] Massachusetts Gen Hosp, Boston, MA 02114 USA
[3] Forschungszentrum Dresden Rossendorf, D-01314 Dresden, Germany
[4] Tech Univ Dresden, D-01307 Dresden, Germany
关键词
positron emission tomography; proton therapy; carbon ion therapy;
D O I
10.1016/j.nima.2008.03.075
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Ion beams offer the possibility of improved conformation of the dose delivered to the tumor with better sparing of surrounding tissue and critical structures in comparison to conventional photon and electron external radiation treatment modalities. Full clinical exploitation of this advantage can benefit from in vivo confirmation of the actual beam delivery and, in particular, of the ion range in the patient. During irradiation, positron emitters like (15)O (half-life T(1/2) approximate to 2 min) and (11)C (T(1/2) approximate to 20 min) are formed in nuclear interactions between the ions and the tissue. Detection of this transient radioactivity via positron emission tomography (PET) and comparison with the expectation based on the prescribed beam application may serve as an in vivo, non-invasive range validation method of the whole treatment planning and delivery chain. For technical implementation, PET imaging during irradiation (in-beam) requires the development of customized, limited angle detectors with data acquisition synchronized with the beam delivery. Alternatively, commercial PET or PET/CT scanners in close proximity to the treatment site enable detection of the residual activation from long-lived emitters shortly after treatment (offline). This paper reviews two clinical examples using a dedicated in-beam PET scanner for verification of carbon ion therapy at GSI Darmstadt, Germany, as well as a commercial offline PET/CT tomograph for post-radiation imaging of proton treatments at Massachusetts General Hospital, Boston, USA. Challenges as well as pros and cons of the two imaging approaches in dependence of the different ion type and beam delivery system are discussed. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:282 / 286
页数:5
相关论文
共 50 条
  • [1] Implementation and initial clinical experience of offline PET/CT-based verification of scanned carbon ion treatment
    Bauer, Julia
    Unholtz, Daniel
    Sommerer, Florian
    Kurz, Christopher
    Haberer, Thomas
    Herfarth, Klaus
    Welzel, Thomas
    Combs, Stephanie E.
    Debus, Juergen
    Parodi, Katia
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 107 (02) : 218 - 226
  • [2] Experience and new prospects of PET imaging for ion beam therapy monitoring
    Parodi, Katia
    Yamaya, Taiga
    Moskal, Pawel
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (01): : 22 - 34
  • [3] Compton-PET Imaging of 10C for Range Verification of Carbon Ion Therapy
    Mohammadi, Akram
    Yoshida, Eiji
    Okumura, Yusuke
    Nitta, Munetaka
    Nishikido, Fumihiko
    Kitagawa, Atsushi
    Kamada, Kei
    Parodi, Katia
    Yamaya, Taiga
    [J]. 2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [4] A dedicated software application for treatment verification with off-line PET/CT imaging at the Heidelberg Ion Beam Therapy Center
    Chen, W.
    Bauer, J.
    Kurz, C.
    Tessonnier, T.
    Handrack, J.
    Haberer, T.
    Debus, J.
    Parodi, K.
    [J]. MICRO-MINI & NANO-DOSIMETRY & INNOVATIVE TECHNOLOGIES IN RADIATION THERAPY (MMND&ITRO2016), 2017, 777
  • [5] Simulation of proton therapy treatment verification via PET imaging of induced positron-emitters
    Beebe-Wang, J
    Vaska, P
    Dilmanian, FA
    Peggs, SG
    Schlyer, DJ
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2496 - 2500
  • [6] In-Beam PET Imaging in Carbon Therapy for Dose Verification
    Li, Zhongxing
    Fan, Yaming
    Dong, Ming
    Tong, Linge
    Zhao, Liang
    Yin, Yongzhi
    Chen, Ximeng
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (01) : 61 - 67
  • [7] Implementation of hypoxia PET imaging in radiation therapy planning
    Geets, X.
    Gregoire, V.
    Lee, J. A.
    [J]. QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 57 (03): : 271 - 282
  • [8] PET/CT imaging for treatment verification after proton therapy: A study with plastic phantoms and metallic implants
    Parodi, Katia
    Paganetti, Harald
    Cascio, Ethan
    Flanz, Jacob B.
    Bonab, Ali A.
    Alpert, Nathaniel M.
    Lohmann, Kevin
    Bortfeld, Thomas
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 419 - 435
  • [9] Comparison of Prompt Gamma and PET Imaging for Range Verification in Proton Therapy
    Moteabbed, M.
    Espana, S.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Clinical evaluation of 4D PET motion compensation strategies for treatment verification in ion beam therapy
    Gianoli, Chiara
    Kurz, Christopher
    Riboldi, Marco
    Bauer, Julia
    Fontana, Giulia
    Baroni, Guido
    Debus, Juergen
    Parodi, Katia
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (11): : 4141 - 4155