Determination of the weighted CT dose index in modern multi-detector CT scanners

被引:34
|
作者
Perisinakis, K.
Damilakis, J.
Tzedakis, A.
Papadakis, A.
Theocharopoulos, N.
Gourtsoyiannis, N.
机构
[1] Univ Crete, Fac Med, Dept Med Phys, Iraklion 71003, Crete, Greece
[2] Educ Technol Inst Crete, Dept Nat Sci, Iraklion 71004, Crete, Greece
[3] Univ Crete, Fac Med, Dept Radiol, Iraklion 71003, Crete, Greece
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 21期
关键词
D O I
10.1088/0031-9155/52/21/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the present study was to (a) evaluate the underestimation in the value of the free-in-air (CTDIair) and the weighted CT dose index (CTDIw) determined with the standard 100 mm pencil chamber, i.e. the CTDI100 concept, for the whole range of nominal radiation beam collimations selectable in a modern multi-slice CT scanner, (b) estimate the optimum length of the pencil-chamber and phantoms for accurate CTDIw measurements and (c) provide CTDIw values normalized to free-in-air CTDI for different tube-voltage, nominal radiation beam collimations and beam filtration values. The underestimation in the determination of CTDIair and CTDIw using the CTDI100 concept was determined from measurements obtained with standard polymethyl-methacrylate (PMMA) phantoms and arrays of thermoluminescence dosimeters. The Monte Carlo N-Particle transport code was used to simulate standard CTDI measurements on a 16-slice CT scanner. The optimum pencil-chamber length for accurate determination of CTDIw was estimated as the minimum chamber length for which a further increase in length does not alter the value of the CTDI. CTDIw/CTDIair ratios were determined using Monte Carlo simulation and the optimum detector length for all selectable tube-voltage values and for three different values of beam filtration. To verify the Monte Carlo results, measured values of CTDIw/CTDIair ratios using the standard 100 mm pencil ionization chamber were compared with corresponding values calculated with Monte Carlo experiments. The underestimation in the determination of CTDIair using the 100 mm pencil chamber was less than 1% for all beam collimations. The underestimation in CTDIw was 15% and 27% for head and body phantoms, respectively. The optimum detector length for accurate CTDIw measurements was found to be 50 cm for the beam collimations commonly employed in modern multi-detector (MD) CT scanners. The ratio of CTDIw/CTDIair determined using the optimum detector length was found to be independent of beam collimation. Percentage differences between measured and calculated corresponding CTDIw/CTDIair ratios were always less than 8% for head and less than 5% for body PMMA phantoms. In conclusion, the CTDIair of MDCT scanners may be measured accurately with a 100 mm pencil chamber. However, the CTDI100 concept was found to be inadequate for accurate CTDIw determination for the wide beam collimations commonly used in MDCT scanners. Accurate CTDIw determination presupposes the use of a pencil chamber and PMMA phantoms at least 50 cm long.
引用
收藏
页码:6485 / 6495
页数:11
相关论文
共 50 条
  • [1] CT: status on multi-detector row scanners
    Martinez-Souloumiac, L.
    Gosso, F.
    JOURNAL DE RADIOLOGIE, 2006, 87 (7-8): : 927 - 935
  • [2] Multi-detector CT dose index assessment with MOSFET technology
    Yoshizumi, T
    Frush, D
    Samei, E
    Goodman, P
    Simoni, P
    Toncheva, G
    Nguyen, G
    Barnes, L
    Lowry, C
    MEDICAL PHYSICS, 2004, 31 (06) : 1841 - 1841
  • [3] Accuracy of the CT numbers of simulated lung nodules imaged with multi-detector CT scanners
    Goodsitt, Mitchell M.
    Chan, Heang-Ping
    Way, Ted W.
    Larson, Sandra C.
    Christodoulou, Emmanuel G.
    Kim, Jeomsoon
    MEDICAL PHYSICS, 2006, 33 (08) : 3006 - 3017
  • [4] Comparison of methods for assessing geometric efficiency on multi-detector CT scanners
    Berris, Theocharis
    Perisinakis, Kostas
    Papadakis, Antonios E.
    Damilakis, John
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (03): : 312 - 322
  • [5] Multi-detector row CT: Radiation dose characteristics
    Hamberg, LM
    Rhea, JT
    Hunter, GJ
    Thrall, JH
    RADIOLOGY, 2003, 226 (03) : 762 - 772
  • [6] CT Dose Optimization When Changing to CT Multi-Detector Row Technology
    Verdun, Francis R.
    Gutierrez, Daniel
    Schnyder, Pierre
    Aroua, Abbas
    Bochud, Francois
    Gudinchet, Francois
    CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2007, 36 (04) : 176 - 184
  • [7] Sixteen-section multi-detector row CT scanners: This changes everything
    Mezrich, R
    ACADEMIC RADIOLOGY, 2003, 10 (04) : 351 - 352
  • [8] Advances in multi-detector CT
    Saini, S
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2004, 28 : S1 - S1
  • [9] Multi-detector CT of the abdomen
    Hamm, B
    EUROPEAN RADIOLOGY, 2003, 13 : M25 - M25
  • [10] Multi-detector CT of the abdomen
    Bernd Hamm
    European Radiology, 2003, 13 (Suppl 5) : 25 - 25