The Supreme Laryngeal Mask Airway™ (LMA): A new neonatal supraglottic device: Comparison with Classic and ProSeal LMA in a manikin

被引:19
|
作者
Trevisanuto, Daniele [1 ]
Parotto, Matteo [2 ]
Doglioni, Nicoletta [1 ]
Ori, Carlo [2 ]
Zanardo, Vincenzo [1 ]
Micaglio, Massimo [2 ]
机构
[1] Univ Padua, Dept Pediat, Sch Med, Azienda Osped, I-35128 Padua, Italy
[2] Univ Padua, Inst Anesthesia & Intens Care, Sch Med, Azienda Osped,Dept Pharmacol & Anesthesiol, I-35128 Padua, Italy
关键词
Airway; Laryngeal mask; Manikin; Neonatal resuscitation; RESUSCITATION; VENTILATION; MANAGEMENT; INFANTS;
D O I
10.1016/j.resuscitation.2011.07.032
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: The study aims to compare the performances (ease of insertion, time to establish effective ventilation and maximal inflation pressure) of classic (TM) (cLMA), ProSeal (TM) (PLMA) and Supreme (TM) (SLMA) Laryngeal Mask Airway when used in a neonatal airway management manikin by inexperienced delivery room trainees. The quality of the three devices, as perceived by participants, was also evaluated. Methods: Health-care professional trainees were given a brief supervised training with the three devices. Every trainee was then observed positioning each of the three different LMAs in a single occasion. Success rate, time (IT) and maximal inflation pressure (PImax) were recorded by a single unblinded observer. A 4-point scale was used to rate participants' perceived quality. Results: A total of 40 health-care professional trainees participated in the study. There were five, three and one failed insertions at the first attempt with the cLMA, PLMA and SLMA, respectively. No failures to establish an effective airway within three attempts were recorded. The success rate at first attempt was comparable among the three devices. The mean IT was significantly lower with the SLMA as compared with PLMA (p < 0.01), but not to cLMA. The mean PImax was higher with SLMA than with cLMA and PLMA (p < 0.01). The ease of insertion as well as the effectiveness of ventilation were perceived by the participants as superior with SLMA as compared with cLMA and PLMA (p < 0.01). Conclusions: Neonatal SLMA is superior to PLMA in terms of time to establish effective ventilation; furthermore, maximal inflation pressure and quality perceived by the operator are higher with neonatal SLMA than with cLMA and PLMA. These manikin data could provide a useful guide for planning potential future clinical research involving the newly developed supraglottic device in neonates. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:97 / 100
页数:4
相关论文
共 50 条
  • [1] Training for neonatal resuscitation with the laryngeal mask airway: a comparison of the LMA-ProSeal and the LMA-Classic in an airway management manikin
    Micaglio, Massimo
    Doglioni, Nicoletta
    Parotto, Matteo
    Zanardo, Vincenzo
    Ori, Carlo
    Trevisanuto, Daniele
    [J]. PEDIATRIC ANESTHESIA, 2006, 16 (10) : 1028 - 1031
  • [2] A new supraglottic airway device: LMA-Supreme™, comparison with LMA-Proseal™
    Hosten, T.
    Gurkan, Y.
    Ozdamar, D.
    Tekin, M.
    Toker, K.
    Solak, M.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2009, 53 (07) : 852 - 857
  • [3] Comparison of laryngeal mask airway (LMA)-Proseal™ and the LMA-Classic™ in ventilated children receiving neuromuscular blockade
    Lardner D.R.R.
    Cox R.G.
    Ewen A.
    Dickinson D.
    [J]. Canadian Journal of Anesthesia/Journal canadien d’anesthésie, 2008, 55 (1): : 29 - 35
  • [4] A comparison of the laryngeal tube s and the LMA ProSeal laryngeal mask airway
    Cook, T. M.
    [J]. ANAESTHESIA, 2007, 62 (12) : 1297 - 1298
  • [5] Comparison between the LMA Classic™ and new I-gel™ supraglottic airway device: a manikin study
    Kwak, D. I.
    Yoo, J. H.
    No, H. W.
    Sohn, J. D.
    Kim, Y. J.
    Nah, J. H.
    [J]. HONG KONG JOURNAL OF EMERGENCY MEDICINE, 2013, 20 (01) : 25 - 33
  • [6] Comparison of classic, and ProSeal laryngeal mask airways (LMA) in pediatric patients
    Gungor, Gurcan
    Sutas Bozkurt, Pervin
    Yuceyayar, Lale
    [J]. IZMIR DR BEHCET UZ COCUK HASTANESI DERGISI, 2016, 6 (01): : 46 - 52
  • [7] A comparison of the laryngeal tube s and the LMA ProSeal laryngeal mask airway - A reply
    Klaver, N. S.
    Kuisenga, K.
    Ballas, A.
    Fidler, V.
    [J]. ANAESTHESIA, 2007, 62 (12) : 1298 - 1298
  • [8] Clinical study of the laryngeal mask in paediatric. Comparison of the LMA-ProSeal and LMA-classic
    Bordes, M.
    Semjen, F.
    Meymat, Y.
    Zaghet, B.
    Suriray, I.
    Cros, A. -M.
    [J]. ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2006, 25 (08): : 806 - 810
  • [9] Comparison of ProSeal LMA with Supreme LMA in paediatric patients
    Hosten, T.
    Gurkan, Y.
    Kus, A.
    Ozdamar, D.
    Aksu, C.
    Solak, M.
    Toker, K.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2013, 57 (08) : 996 - 1001
  • [10] A randomised comparison of the LMA Supreme™ and LMA ProSeal™ in children
    Jagannathan, N.
    Sohn, L. E.
    Sawardekar, A.
    Gordon, J.
    Langen, K. E.
    Anderson, K.
    [J]. ANAESTHESIA, 2012, 67 (06) : 632 - 639