Comparison of full-field digital mammography and film-screen mammography:: image quality and lesion detection

被引:53
|
作者
Fischmann, A
Siegmann, KC
Wersebe, A
Claussen, CD
Müller-Schimpfle, M
机构
[1] Univ Tubingen Hosp, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Stad Kliniken Frankfurt Main, Cent Inst Radiol, Frankfurt, Germany
来源
BRITISH JOURNAL OF RADIOLOGY | 2005年 / 78卷 / 928期
关键词
D O I
10.1259/bjr/33317317
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objective of this study is to compare image quality and lesion detection for full field digital mammography (FFDM) and film-screen mammography (FSM). In 200 women we performed digital mammography of one breast and film-screen mammography of the other breast. Imaging parameters were set automatically. Image quality, visualization of calcifications and masses were rated by three readers independently. Mean glandular dose was calculated for both systems. We found no significant difference in mean glandular dose. Image quality was rated by reader A/B/C as excellent for FFDM in 153/155/167 cases and for FSM in 139/116/114 cases (p < 0.03/0.001/0.001). Microcalcifications were detected by FFDM in 103/89/98 and by FSM in 76/76/76 cases (p < 0.01/0.06/0.01). Detection of masses did not differ significantly. FFDM provided significantly better visibility of skin and nipple-areola region (p < 0.01). FFDM demonstrated improved image quality compared with film-screen mammography. Microcalcification detection was also significantly better with the digital mammography system for two of the three readers.
引用
收藏
页码:312 / 315
页数:4
相关论文
共 50 条
  • [1] A comparison of the accuracy of film-screen mammography, full-field digital mammography, and digital breast tomosynthesis
    Michell, M. J.
    Iqbal, A.
    Wasan, R. K.
    Evans, D. R.
    Peacock, C.
    Lawinski, C. P.
    Douiri, A.
    Wilson, R.
    Whelehan, P.
    [J]. CLINICAL RADIOLOGY, 2012, 67 (10) : 976 - 981
  • [2] Full-field digital mammography vs film-screen mammography in patients with microcalcifications: Comparison of detection and characterization
    Fischer, U
    Baum, FT
    Luftner-Nagel, S
    Obenauer, S
    Von Heyden, D
    Grabbe, EH
    [J]. RADIOLOGY, 2000, 217 : 201 - 201
  • [3] Detection and characterization of microcalcification with film-screen and full-field digital mammography (FFDM)
    O'Riordan, E
    Bukhanov, K
    Wilson, C
    Sidlofsky, S
    Goldberg, F
    Muradali, D
    [J]. RADIOLOGY, 2000, 217 : 200 - 200
  • [4] Comparison of a new full-field digital mammography system and conventional film-screen mammography: A phantom study
    Bick, U
    Diekmann, F
    Marth, F
    Le Roux, A
    Juran, R
    Friedrich, M
    [J]. RADIOLOGY, 1999, 213P : 152 - 152
  • [5] Contrast-detail-resolution of full-field digital mammography in comparison to conventional film-screen mammography
    Bick, U
    Diekmann, F
    Grebe, S
    Marth, F
    Juran, R
    Friedrich, M
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 627 - 632
  • [6] Visualization of microcalcifications on mammographies obtained by digital full-field mammography in comparison to conventional film-screen mammography
    Diekmann, S
    Bick, U
    von Heyden, H
    Diekmann, F
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2003, 175 (06): : 775 - 779
  • [7] Visualization of microcalcifications on mammographies obtained by digital full-field mammography in comparison to conventional film-screen mammography
    Diekmann, SD
    Von Heyden, H
    Bick, U
    Von Heyden, D
    Luftner-Nagel, S
    Hamm, BK
    [J]. RADIOLOGY, 2002, 225 : 416 - 416
  • [8] Improved detection of low contrast objects with full-field digital mammography compared to film-screen mammography
    Niklason, LT
    OpsahlOng, B
    Landberg, C
    Castleberry, DE
    Venkatakrishnan, V
    Kopans, DB
    [J]. RADIOLOGY, 1997, 205 : 1249 - 1249
  • [9] Comparison of the microcalcifications between full-field digital mammography and screen film mammography
    Tohno, E
    Kujiraoka, Y
    Watanabe, Y
    Itai, Y
    [J]. RADIOLOGY, 2001, 221 : 189 - 189
  • [10] Visualization of ductal carcinoma in situ and early invasive carcinoma: Comparison of film-screen mammography and full-field digital mammography
    Feig, SA
    Eskola-Feig, C
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 451 - 460