摘要
In photon radiotherapy, thermoluminescence dosemeters (TLD) are commonly used for in vivo dosimetry. In proton and neutron therapy, their use raises a particular difficulty since their responses vary significantly with the energy spectrum. With reference to Co-60, the response of CaF2:Tm (TLD-300) and (LiF)-Li-7:Mg,Ti (TLD-700) crystals was investigated in the 85 MeV proton beam and the p(65)+Be neutron beam produced at the cyclotron at Louvain-La-Neuve and used for therapy applications. In the proton beam, TLD dosemeters were irradiated in a phantom, at different positions in the unmodulated and modulated Bragg curve. In the fast neutron beam, TLDs were positioned inside the geometrical limits of the beam at different depths, and outside its limits behind five leaves of the multi-leaf collimator. peak height variation of glow curves was investigated and correlated with the variation of radiation quality.
引用
收藏
页码:299 / 304
页数:6
相关论文
共 50 条
  • [1] Thermoluminescence response of sodalime glass irradiated with proton and neutron beams
    Marrale, M.
    Longo, A.
    Bartolotta, A.
    Basile, S.
    D'Oca, M. C.
    Tomarchio, E.
    Cirrone, G. A. P.
    Di Rosa, F.
    Romano, F.
    Cuttone, G.
    Brai, M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2012, 292 : 55 - 58
  • [2] Thermoluminescence dosimetry in a fast neutron therapy beam
    Bhengu, KJ
    Langen, KM
    Symons, JE
    Jones, DTL
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2004, 71 (3-4) : 981 - 982
  • [3] Fast Neutron Beams for Boron Neutron Capture Therapy?
    Wolber, Gerd
    [J]. Zeitschrift fur Medizinische Physik, 2004, 14 (01): : 55 - 63
  • [4] Fast neutron and proton therapy sources
    Jones, DTL
    [J]. RADIOCHIMICA ACTA, 2001, 89 (4-5) : 265 - 277
  • [5] FAST-NEUTRON BEAMS FOR RADIATION THERAPY
    BRENNAN, JT
    [J]. TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1968, 11 (01): : 398 - &
  • [6] Reference dosimetry for fast neutron and proton therapy
    Jones, DTL
    [J]. RADIOCHIMICA ACTA, 2001, 89 (4-5) : 279 - 287
  • [7] Reference dosimetry for fast neutron and proton therapy
    Jones, DTL
    [J]. INTERNATIONAL CONFERENCE ON NUCLEAR DATA FOR SCIENCE AND TECHNOLOGY, PTS 1 AND 2, 2005, 769 : 1588 - 1591
  • [8] DOSIMETRIC PROPERTIES OF FAST-NEUTRON THERAPY BEAMS AT TAMVEC
    SMITH, AR
    ALMOND, PR
    SMATHERS, JB
    OTTE, VA
    [J]. RADIOLOGY, 1974, 113 (01) : 187 - 193
  • [9] RADIOBIOLOGICAL INTERCOMPARISONS OF FAST-NEUTRON BEAMS USED IN THERAPY
    BEAUDUIN, M
    GUEULETTE, J
    DECOSTER, BM
    GREGOIRE, V
    OCTAVEPRIGNOT, M
    VYNCKIER, S
    WAMBERSIE, A
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 1989, 165 (04): : 263 - 267
  • [10] Thermoluminescence curves of LiF dosemeters irradiated with proton beams
    Bartolotta, A
    Basile, S
    Brai, M
    Bruno, G
    Cuttone, G
    Zabalza, RR
    Teri, G
    [J]. NUCLEAR AND CONDENSED MATTER PHYSICS, 2000, 513 : 35 - 38