Estimation of current density distribution under electrodes for external defibrillation

被引:42
|
作者
Krasteva, Vessela Tz [1 ]
Papazov, Sava P. [1 ]
机构
[1] Bulgarian Acad Sci, Ctr Biomed Engn, Acad G Bonchev Str,Block 105, BU-1113 Sofia, Bulgaria
关键词
Specific Resistivity; Current Density Distribution; Wearable Electrode; Automatic External Defibrillator; Circular Electrode;
D O I
10.1186/1475-925X-1-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Transthoracic defibrillation is the most common life-saving technique for the restoration of the heart rhythm of cardiac arrest victims. The procedure requires adequate application of large electrodes on the patient chest, to ensure low-resistance electrical contact. The current density distribution under the electrodes is non-uniform, leading to muscle contraction and pain, or risks of burning. The recent introduction of automatic external defibrillators and even wearable defibrillators, presents new demanding requirements for the structure of electrodes. Method and Results: Using the pseudo-elliptic differential equation of Laplace type with appropriate boundary conditions and applying finite element method modeling, electrodes of various shapes and structure were studied. The non-uniformity of the current density distribution was shown to be moderately improved by adding a low resistivity layer between the metal and tissue and by a ring around the electrode perimeter. The inclusion of openings in long-term wearable electrodes additionally disturbs the current density profile. However, a number of small-size perforations may result in acceptable current density distribution. Conclusion: The current density distribution non-uniformity of circular electrodes is about 30% less than that of square-shaped electrodes. The use of an interface layer of intermediate resistivity, comparable to that of the underlying tissues, and a high-resistivity perimeter ring, can further improve the distribution. The inclusion of skin aeration openings disturbs the current paths, but an appropriate selection of number and size provides a reasonable compromise.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Estimation of current density distribution under electrodes for external defibrillation
    Vessela Tz Krasteva
    Sava P Papazov
    BioMedical Engineering OnLine, 1
  • [2] Current and energy in external cardiac defibrillation
    Watson, A.B.
    Ellis, M.S.
    Australasian Physical & Engineering Sciences in Medicine, 1994, 17 (04)
  • [3] 3-DIMENSIONAL CURRENT-DENSITY DISTRIBUTION UNDER SURFACE STIMULATION ELECTRODES
    SAGIDOLEV, AM
    PRUTCHI, D
    NATHAN, RH
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1995, 33 (03) : 403 - 408
  • [4] INTRACARDIAC CURRENT DISTRIBUTION IN DEFIBRILLATION
    HOYT, R
    GRAYZEL, J
    KERBER, R
    MEDICAL INSTRUMENTATION, 1980, 14 (01): : 55 - 56
  • [6] EFFECTS OF ANISOTROPIC MYOCARDIAL CONDUCTIVITY IN MODEL OF DEFIBRILLATION CURRENT-DENSITY DISTRIBUTION
    MALIK, M
    SMITS, KFAA
    LINDEMANS, FW
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (04) : S34 - S40
  • [7] UNIFORMITY OF CURRENT-DENSITY UNDER STIMULATING ELECTRODES
    KIM, YM
    ZIEBER, HG
    WANG, FE
    CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1990, 17 (06) : 585 - 619
  • [8] External cardiac defibrillation in pigs with in situ cerebral stimulation electrodes
    Eisner, W
    Kolbitsch, C
    Kleinsasser, A
    Fiegele, T
    Biebl, M
    Löckinger, A
    Lorenz, IH
    Moser, PL
    Bauer, R
    Bale, R
    Twerdy, K
    12TH EUROPEAN CONGRESS OF NEUROSURGERY (EANS), PROCEEDINGS, 2003, : 975 - 978
  • [9] THE CURRENT-DENSITY DISTRIBUTION OF RESISTIVE DISK SURFACE ELECTRODES
    CILLIERS, PJ
    WEED, HR
    PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, : 814 - 815
  • [10] NONUNIFORMITY OF CURRENT-DENSITY IN CARDIAC DEFIBRILLATION
    DOLAN, AM
    HORACEK, BM
    RAUTAHARJU, PM
    CIRCULATION, 1976, 54 (04) : 171 - 171