Method to account for dose fractionation in analysis of IMRT plans: Modified equivalent uniform dose

被引:28
|
作者
Park, CS [1 ]
Kim, Y [1 ]
Lee, N [1 ]
Bucci, KM [1 ]
Quivey, JM [1 ]
Verhey, LJ [1 ]
Xia, P [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, Sch Med, San Francisco, CA 94143 USA
关键词
dose fractionation; intensity modulated radiotherapy; modified equivalent uniform dose; generalized equivalent uniform dose;
D O I
10.1016/j.ijrobp.2004.11.039
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To propose a modified equivalent uniform dose (mEUD) to account for dose fractionation using the biologically effective dose without losing the advantages of the generalized equivalent uniform dose (gEUD) and to report the calculated mEUD and gEUD in clinically used intensity-modulated radiotherapy (IMRT) plans. Methods and Materials: The proposed mEUD replaces the dose to each voxel in the gEUD formulation by a biologically effective dose with a normalization factor. We propose to use the term mEUD(D0/n0). that includes the total dose (Do) and number of fractions (no) and to use the term mEUD(0) that includes the same total dose but a standard fraction size of 2 Gy. A total of 41 IMRT plans for patients with nasopharyngeal cancer treated at our institution between October 1997 and March 2002 were selected for the study. The gEVD and mEUD were calculated for the planning gross tumor volume (pGTV), planning clinical tumor volume (pCTV), parotid glands, and spinal cord. The prescription dose for these patients was 70 Gy to > 95 % of the pGTV and 59.4 Gy to > 95 % of the pCTV in 33 fractions. Results: The calculated average gEUD was 72.2 +/- 2.4 Gy for the pGTV, 54.2 +/- 7.1 Gy for the pCTV, 26.7 +/- 4.2 Gy for the parotid glands, and 34.1 +/- 6.8 Gy for the spinal cord. The calculated average mEUD(D0/n0) using 33 fractions was 71.7 +/- 3.5 Gy for mEUD(70/33) of the pGTV, 49.9 +/- 7.9 Gy for mEUD(59.5/33) of the pCTV, 27.6 +/- 4.8 Gy for mEUD(26/33) of the parotid glands, and 32.7 +/- 7.8 Gy for mEUD(45/33) of the spinal cord. Conclusion: The proposed mEUD, combining the gEUD with the biologically effective dose, preserves all advantages of the gEUD while reflecting the fractionation effects and linear and quadratic survival characteristics. (c) 2005 Elsevier Inc.
引用
收藏
页码:925 / 932
页数:8
相关论文
共 50 条
  • [1] MODIFIED DOSE FRACTIONATION BY BIOLOGICAL EQUIVALENT DOSE VOLUME HISTOGRAM IN IMRT OF NON SMALL CELL LUNG CANCER
    Pedicini, P.
    Caivano, R.
    Califano, G.
    Fiorentino, A.
    Barbieri, V.
    Sanpaolo, P.
    Castaldo, G.
    Fusco, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 98 : S31 - S31
  • [2] A Preliminary Study On Comparison of Dose-Volume Based Optimization and Equivalent Uniform Dose Based Optimization for IMRT Plans
    Subramaniam, B.
    Ding, W.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [3] Target equivalent uniform dose as a quality indicator for head-and-neck IMRT treatment plans
    Constantinescu, C.
    Al-Wassia, R.
    Al-Hebshi, A.
    Bahadur, Y.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S214 - S215
  • [4] Is uniform target dose possible in IMRT plans in the head and neck?
    Vineberg, KA
    Eisbruch, A
    Coselmon, MM
    McShan, DL
    Kessler, ML
    Fraass, BA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (05): : 1159 - 1172
  • [5] IMRT optimization based on the generalized Equivalent Uniform Dose (EUD)
    Wu, QW
    Mohan, R
    Niemierko, A
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 710 - 713
  • [6] IMRT optimization based on the generalized Equivalent Uniform Dose (EUD)
    Wu, QW
    Mohan, R
    Niemierko, A
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 17 - 19
  • [7] Dose Verification of IMRT Plans: Reconsideration of Acceptance Criteria and Analysis Method
    Rezaee, M.
    Wong, J.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E609 - E609
  • [8] COMPARISON OF THE BIOLOGICALLY EFFECTIVE UNIFORM DOSE, EQUIVALENT UNIFORM DOSE, EFFECTIVE DOSE AND GENERALIZED EQUIVALENT UNIFORM DOSE CONCEPTS
    Mavroidis, P.
    Papanikolaou, N.
    Lind, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S239 - S240
  • [9] Dose / fractionation / IMRT / Imaging
    Faivre-Finn, C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S102 - S102
  • [10] Equivalent uniform dose (EUD) based optimization for IMRT treatment planning
    Mihailidis, D.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2295 - 2295