Comparison of the accuracy of digital face scans obtained by two different scanners: An in vivo study

被引:14
|
作者
Pellitteri, Federica [1 ]
Brucculeri, Luca [1 ]
Spedicato, Giorgio Alfredo [2 ]
Siciliani, Giuseppe [3 ]
Lombardo, Luca [1 ]
机构
[1] Univ Ferrara, Postgrad Sch Orthodont, Ferrara, Italy
[2] Catholic Univ Milan, Fac Banking & Insurance, Milan, Italy
[3] Univ Ferrara, Sch Dent, Ferrara, Italy
关键词
Face scanner; Dental Pro; Face Hunter; Orthodontics; Face analysis; SOFT-TISSUE LANDMARKS; FACIAL MORPHOLOGY; REPRODUCIBILITY; POSITION; ADULTS; SEX;
D O I
10.2319/092720-823.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare the degree of accuracy of the Face Hunter facial scanner and the Dental Pro application for facial scanning, with respect to both manual measurements and each other. Materials and Methods: Twenty-five patients were measured manually and scanned using each device. Six reference markers were placed on each subject's face at the cephalometric points Tr, Na', Prn, Pog', and L-R Zyg. Digital measurement software was used to calculate the distances between the cephalometric reference points on each of the scans. Geomagic X Control was used to superimpose the scans, automatically determining the best-fit alignment and calculating the percentage of overlapping surfaces within the tolerance ranges. Results: Individual comparisons of the four distances measured anthropometrically and on the scans yielded an intraclass correlation coefficient index greater than.9. The t-test for matched samples yielded a P value below the significance threshold. Right and left cheeks reached around 60% of the surface, with a margin of error between 0.5 mm and -0.5 mm. The forehead was the only area in which most of the surface fell within the poorly reproducible range, presenting values out of tolerance of more than 20%. Conclusions: Three-dimensional scans of the facial surface provide an excellent analytical tool for clinical evaluation; it does not appear that one or the other of the measuring tools is systematically more accurate, and the cheeks are the area with the highest average percentage of surface in the highly reproducible range.
引用
收藏
页码:641 / 649
页数:9
相关论文
共 50 条
  • [1] The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study
    Amornvit, Pokpong
    Sanohkan, Sasiwimol
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2019, 16 (24)
  • [2] Accuracy, Time, and Comfort of Different Intraoral Scanners: An In Vivo Comparison Study
    Lione, Roberta
    De Razza, Francesca Chiara
    Gazzani, Francesca
    Lugli, Letizia
    Cozza, Paola
    Pavoni, Chiara
    [J]. APPLIED SCIENCES-BASEL, 2024, 14 (17):
  • [3] Accuracy of intraoral scans: An in vivo study of different scanning devices
    Kernen, Florian
    Schlager, Stefan
    Alvarez, Veronica Seidel
    Mehrhof, Juergen
    Vach, Kirstin
    Kohal, Ralf
    Nelson, Katja
    Fluegge, Tabea
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2022, 128 (06): : 1303 - 1309
  • [4] Accuracy Comparison of 3D Face Scans Obtained by Portable Stereophotogrammetry and Smartphone Applications
    Van Lint, Lina
    Christiaens, Lynn
    Stroo, Valerie
    Bila, Michel
    Willaert, Robin
    Sun, Yi
    Van Dessel, Jeroen
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2023, 43 (05) : 550 - 560
  • [5] Accuracy Comparison of 3D Face Scans Obtained by Portable Stereophotogrammetry and Smartphone Applications
    Lina Van Lint
    Lynn Christiaens
    Valerie Stroo
    Michel Bila
    Robin Willaert
    Yi Sun
    Jeroen Van Dessel
    [J]. Journal of Medical and Biological Engineering, 2023, 43 : 550 - 560
  • [6] Accuracy of intraoral scans of edentulous jaws with different generations of intraoral scanners compared to laboratory scans
    Kontis, Panagiotis
    Gueth, Jan-Frederik
    Schubert, Oliver
    Keul, Christine
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2021, 13 (05): : 316 - 326
  • [7] Accuracy of 3D facial scans: a comparison of three different scanning system in an in vivo study
    Federica Pellitteri
    Fabrizio Scisciola
    Francesca Cremonini
    Matilde Baciliero
    Luca Lombardo
    [J]. Progress in Orthodontics, 24
  • [8] Accuracy of 3D facial scans: a comparison of three different scanning system in an in vivo study
    Pellitteri, Federica
    Scisciola, Fabrizio
    Cremonini, Francesca
    Baciliero, Matilde
    Lombardo, Luca
    [J]. PROGRESS IN ORTHODONTICS, 2023, 24 (01)
  • [9] Comparison of the Accuracy of 3D Images Obtained from Different Types of Scanners: A Systematic Review
    Kustrzycka, Dorota
    Marschang, Tim
    Mikulewicz, Marcin
    Grzebieluch, Wojciech
    [J]. JOURNAL OF HEALTHCARE ENGINEERING, 2020, 2020 (2020)
  • [10] Evaluation of the Accuracy of Digital Impressions Obtained from Intraoral and Extraoral Dental Scanners with Different CAD/CAM Scanning Technologies: An In Vitro Study
    Ellakany, Passent
    El Tantawi, Maha
    Mahrous, Amr A.
    Al-Harbi, Fahad
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (04): : 314 - 319