Segmental and dynamic intensity-modulated radiotherapy delivery techniques for micro-multileaf collimator

被引:21
|
作者
Agazaryan, N [1 ]
Solberg, TD [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiat Oncol, Los Angeles, CA 90095 USA
关键词
radiotherapy; IMRT; leaf sequencing; DMLC; SMLC;
D O I
10.1118/1.1578791
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A leaf sequencing algorithm has been implemented to deliver segmental and dynamic multileaf collimated intensity-modulated radiotherapy (SMLC-IMRT and DMLC-IMRT, respectively) using a linear accelerator equipped with a micro-multileaf collimator (mMLC). The implementation extends a previously published algorithm for the SMLC-IMRT to include the dynamic MLC-IMRT method and several dosimetric considerations. The algorithm has been extended to account for the transmitted radiation and minimize the leakage between opposing and neighboring leaves. The underdosage problem associated with the tongue-and-groove design of the MLC is significantly reduced by synchronizing the MLC leaf movements. The workings of the leaf sequencing parameters have been investigated and the results of the planar dosimetric investigations show that the sequencing parameters affect the measured dose distributions as intended. Investigations of clinical cases suggest that SMLC and DMLC delivery methods produce comparable results with leaf sequences obtained by root-mean-square (RMS) errors specification of 1.5% and lower, approximately corresponding to 20 or more segments. For SMLC-IMRT, there is little to be gained by using an RMS error specification smaller than 2%, approximately corresponding to 15 segments; however, more segments directly translate to longer treatment time and more strain on the MLC. The implemented leaf synchronization method does not increase the required monitor units while it reduces the measured TG underdoses from a maximum of 12% to a maximum of 3% observed with single field measurements of representative clinical cases studied. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1758 / 1767
页数:10
相关论文
共 50 条
  • [1] Dosimetric advantage and clinical implication of a micro-multileaf collimator in the treatment of prostate with intensity-modulated radiotherapy
    Wang, L
    Hoban, P
    Paskalev, K
    Yang, J
    Li, JS
    Chen, LL
    Xiong, WJ
    Ma, CCM
    [J]. MEDICAL DOSIMETRY, 2005, 30 (02) : 97 - 103
  • [2] Delivery of intensity-modulated treatments with dynamic multileaf collimator
    Boyer, AL
    Strait, JP
    [J]. PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 13 - 15
  • [3] High leaf-speed effects in delivery of intensity modulated radiotherapy with a micro-multileaf collimator
    Garibaldi, C
    Azzeroni, R
    Cattani, F
    Cambria, R
    Tosi, G
    Orecchia, R
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S97 - S98
  • [4] Skull Base Meningioma - Comparison of Intensity-modulated Radiotherapy Planning Techniques using the Moduleaf Micro-multileaf Collimator and Helical Tomotherapy
    Estall, V.
    Fairfoul, J.
    Jena, R.
    Jefferies, S. J.
    Burton, K. E.
    Burnet, N. G.
    [J]. CLINICAL ONCOLOGY, 2010, 22 (03) : 179 - 184
  • [5] Intensity-modulated stereotactic radiosurgery using dynamic micro-multileaf collimation
    Benedict, SH
    Cardinale, RM
    Wu, QW
    Zwicker, RD
    Broaddus, WC
    Mohan, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (03): : 751 - 758
  • [6] Delivery of intensity-modulated radiation therapy with a conventional multileaf collimator: Comparison of dynamic and segmental methods
    Chui, CS
    Chan, MF
    Yorke, E
    Spirou, S
    Ling, CC
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2441 - 2449
  • [7] Delivery of intensity-modulated radiotherapy with dynamic multileaf collimators
    Mohan, R
    Wu, Y
    Wu, QW
    [J]. GENERAL PRACTICE OF RADIATION ONCOLOGY PHYSICS IN THE 21ST CENTURY, 2000, (26): : 113 - 135
  • [8] A simple plan validation procedure for intensity-modulated radiotherapy with dynamic multileaf collimator
    Selvaraj, R
    King, G
    Chern, SS
    Lalonde, R
    Wu, A
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1268 - 1269
  • [9] FOUR-DIMENSIONAL INTENSITY-MODULATED RADIOTHERAPY PLANNING FOR DYNAMIC MULTILEAF COLLIMATOR TRACKING RADIOTHERAPY
    Liang, Yueqiang
    Xu, Hongbing
    Yao, Jonathan
    Li, Zhihui
    Chen, Wendy
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (01): : 266 - 274
  • [10] A modified method of planning and delivery for dynamic multileaf collimator intensity-modulated radiation therapy
    Dogan, N
    Leybovich, LB
    Sethi, A
    Krasin, M
    Emami, B
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (01): : 241 - 245