Deep brain stimulation in intraoperative MRI environment - Comparison of imaging techniques and electrode fixation methods

被引:36
|
作者
Lee, MWY
De Salles, AAF
Frighetto, L
Torres, R
Behnke, E
Bronstein, JM
机构
[1] Univ Calif Los Angeles, Sch Med, Div Neurosurg, Los Angeles, CA 90024 USA
[2] Univ Calif Los Angeles, Sch Med, Dept Neurol, Los Angeles, CA 90024 USA
关键词
deep brain stimulation; electrode fixation device; intraoperative magnetic resonance imaging; stereotactic targeting;
D O I
10.1055/s-2004-830169
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We performed 118 consecutive DBS cases from November 1999 to June 2002. Intraoperatively there were 10 cases studied with fluoroscopy, 73 with 0.2 Tesla (T) MRI and 35 with 1.5 T MRI. Ten electrodes were secured by Medtronic caps, 25 by methyl methacrylate with titanium miniplates, and 82 by Navigus caps. The 3-dimensional displacement between the planned target and actual electrode position (3DD) was determined by fusing the postoperative MRI with the preoperative imaging. The 3DD for using Medtronic caps, methyl methacrylate with miniplates, and Navigus caps were 4.80 +/- 3.16,2.64 +/- 1.26 and 2.23 +/- 1.15 mm (mean +/- SD), respectively. Navigus caps had statistically significant accuracy (P = 0.03) in holding the electrode when compared with Medtronic caps, and it facilitated electrode revision. The Fixation devices significantly affect the final vertical position of the electrode. The 3DD for fluoroscopy, 0.2 T and 1.5 T MRI cases were 4.80 +/- 3.16, 2.31 +/- 1.21 and 2.34 +/- 1.14 mm (mean SD), respectively. No statistically significant difference (P = 0.91) in 3DD was demonstrated between 0.2 T and 1.5 T MRI cases. The presence of intraoperative 1.5 T MRI allowed near real-time electrode position confirmation and early detection of hemorrhagic complications. Satisfactory microelectrode recording was feasible in low-field 0.2 T and high-field 1.5 T MRI environments. Further studies on performing DBS in real-time intraoperative MRI are warranted.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 50 条
  • [1] Deep brain stimulation and intraoperative MRI
    Elias, W. Jeff
    [J]. JOURNAL OF NEUROSURGERY, 2016, 124 (01) : 59 - 60
  • [2] Intraoperative MRI for deep brain stimulation electrode positioning in Parkinson's disease
    Sidiropoulos, C.
    LeWitt, P. A.
    Taylor, D.
    Kaminski, P.
    Scarpace, L.
    Gorham, J.
    Schwalb, J.
    [J]. MOVEMENT DISORDERS, 2013, 28 : S459 - S459
  • [3] Deep brain stimulation and intraoperative MRI Response
    Cui, Zhiqiang
    Ling, Zhipei
    [J]. JOURNAL OF NEUROSURGERY, 2016, 124 (01) : 60 - 61
  • [4] Intraoperative MRI for optimizing electrode placement for deep brain stimulation of the subthalamic nucleus in Parkinson disease
    Cui, Zhiqiang
    Pan, Longsheng
    Song, Huifang
    Xu, Xin
    Xu, Bainan
    Yu, Xinguang
    Ling, Zhipei
    [J]. JOURNAL OF NEUROSURGERY, 2016, 124 (01) : 62 - 69
  • [5] Comparison of Intraoperative Computed Tomography Scan with Postoperative Magnetic Resonance Imaging for Determining Deep Brain Stimulation Electrode Coordinates
    Girgis, Fady
    Zarabi, Hosniya
    Said, Mena
    Zhang, Lin
    Shahlaie, Kiarash
    Saez, Ignacio
    [J]. WORLD NEUROSURGERY, 2020, 138 : E330 - E335
  • [6] Intraoperative Stereotactic Magnetic Resonance Imaging for Deep Brain Stimulation Electrode Planning in Patients with Movement Disorders
    Jakobs, Martin
    Krasniqi, Erenik
    Kloss, Mania
    Neumann, Jan-Oliver
    Campos, Benito
    Unterberg, Andreas W.
    Kiening, Karl L.
    [J]. WORLD NEUROSURGERY, 2018, 119 : E801 - E808
  • [7] Modified Cement-Based Fixation of the Deep Brain Stimulation Electrode
    Smeets, Anouk Y. J. M.
    Ackermans, Linda
    Oosterloo, Mayke
    Kuijf, Mark L.
    van Overbeeke, Jacobus J.
    Visser-Vandewalle, Veerle
    Temel, Yasin
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2015, 93 (01) : 67 - 67
  • [8] Deep Brain Stimulation in Olomouc - Techniques, Electrode Locations, and Outcomes
    Krahulik, D.
    Nevrly, M.
    Otruba, P.
    Kanovsky, P.
    [J]. CESKA A SLOVENSKA NEUROLOGIE A NEUROCHIRURGIE, 2014, 77 (01) : 54 - 58
  • [9] Accuracy of stereotactic electrode placement in deep brain stimulation by intraoperative computed tomography
    Fiegele, Thomas
    Feuchtner, Gudrun
    Sohm, Florian
    Bauer, Richard
    Anton, Juergen Volker
    Gotwald, Thaddaeus
    Twerdy, Klaus
    Eisner, Wilhelm
    [J]. PARKINSONISM & RELATED DISORDERS, 2008, 14 (08) : 595 - 599
  • [10] Verification of the Deep Brain Stimulation Electrode Position Using Intraoperative Electromagnetic Localization
    Burchiel, Kim J.
    Kinsman, Michael
    Mansfield, Kevin
    Mitchell, Ann
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2020, 98 (01) : 37 - 42