Control of photon beam dose profiles by localized transverse magnetic fields

被引:23
|
作者
Reiffel, L
Li, A
Chu, J
Wheatley, RW
Naqvi, S
Pillsbury, R
Saxena, A
机构
[1] Exelar Corp, Chicago, IL 60614 USA
[2] Rush Presbyterian St Lukes Med Ctr, Dept Med Phys, Chicago, IL 60612 USA
[3] Rush Presbyterian St Lukes Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
[4] Superconduct Associates Inc, Niskayuna, NY 12309 USA
[5] Sherbrooke Consulting Inc, Washington, DC 20006 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2000年 / 45卷 / 12期
关键词
D O I
10.1088/0031-9155/45/12/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Unlike electron beams, scant attention has been paid in the literature to possible magnetic field effects on therapeutic photon beams. Generally, dose profiles are considered to be fully determined by beam shape, photon spectrum and the substances in the beam path. Here we show that small superconducting magnets can exercise potentially useful control over photon dose profiles. The magnet produces a locally strong transverse field with large gradients and is applied to the tissue surface below which the photon beam is passing. For one practical magnet design, our simulations, which use the EGS-4 Monte Carlo code modified to include magnetic field effects, show significant intensification and shielding effects. In water phantoms, the effects extend to 3-4 cm or more beyond the warm face of the cryostat and greater distances are achieved in phantoms simulating lung (density similar to0.3). Advances in applying the concept and in superconducting materials and magnet design hold promise for extending these ranges.
引用
收藏
页码:N177 / N182
页数:6
相关论文
共 50 条
  • [1] Effect of Transverse Magnetic Fields On MV Photon Dose Distributions in Heterogeneous Media
    Gauthier, M.
    Bouchard, H.
    Despres, P.
    MEDICAL PHYSICS, 2009, 36 (06) : 2618 - +
  • [2] Conformal photon-beam therapy with transverse magnetic fields: Monte Carlo study
    Li, XA
    Reiffel, L
    Chu, J
    Naqvi, S
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 3200 - 3202
  • [3] Conformal photon-beam therapy with transverse magnetic fields: A Monte Carlo study
    Li, XA
    Reiffel, L
    Chu, J
    Naqvi, S
    MEDICAL PHYSICS, 2001, 28 (02) : 127 - 133
  • [4] Impact of transverse magnetic fields on dose response of a nanoDot OSLD in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Hoshida, Kento
    Ohno, Takeshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 70 : 153 - 160
  • [5] Influence of a Localized Transverse Magnetic Field on Dose Distributions of MV Photon Beams in Inhomogeneous Medium
    Ahn, Woo Sang
    Choi, Wonsik
    Ka, Young Ran
    Kwak, Jungwon
    Jung, In-Hye
    Ahn, Sohyun
    Shin, Seong Soo
    JOURNAL OF MAGNETICS, 2020, 25 (03) : 409 - 414
  • [6] Impact of transverse magnetic fields on dose response of a radiophotoluminescent glass dosimeter in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Ohno, Takeshi
    MEDICAL PHYSICS, 2020, 47 (04) : 1995 - 2004
  • [7] Electron beam therapy with transverse magnetic fields
    Nardi, E
    Barnea, G
    MEDICAL PHYSICS, 1999, 26 (06) : 967 - 973
  • [8] Experimental measurements of dose modulation induced by a localized magnetic field on 18 MV photon beam
    Hsi, WC
    Chu, JCH
    Reiffel, L
    Moyer, D
    Saxena, AV
    MEDICAL PHYSICS, 2003, 30 (06) : 1509 - 1510
  • [9] Recovery of dose profiles of small photon fields in magnetic field using artificial neural networks
    Looe, H. K.
    Blum, I.
    Schoenfeld, A.
    Tekin, T.
    Delfs, B.
    Poppe, B.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S93 - S94
  • [10] Dose calculation and beam modeling for small photon fields
    Ahnesjo, A.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S124 - S124