Protocol for a prospective, controlled, cross-sectional, diagnostic accuracy study to evaluate the specificity and sensitivity of ambulatory monitoring systems in the prompt detection of hypoxia and during movement

被引:6
|
作者
Areia, Carlos [1 ]
Vollam, Sarah [1 ]
Piper, Philippa [2 ]
King, Elizabeth [2 ]
Ede, Jody [1 ]
Young, Louise [1 ]
Santos, Mauro [3 ]
Pimentel, Marco A. F. [3 ]
Roman, Cristian [3 ]
Harford, Mirae [1 ]
Shah, Akshay [4 ]
Gustafson, Owen [2 ]
Rowland, Matthew [1 ]
Tarassenko, Lionel [3 ]
Watkinson, Peter J. [1 ]
机构
[1] Univ Oxford, Crit Care Res Grp, Nuffield Dept Clin Neurosci, Oxford, England
[2] Oxford Univ Hosp NHS Fdn Trust, Adult Intens Care Unit, Oxford, England
[3] Univ Oxford, Inst Biomed Engn, Dept Engn Sci, Oxford, England
[4] Univ Oxford, Radcliffe Dept Med, Oxford, England
来源
BMJ OPEN | 2020年 / 10卷 / 01期
基金
英国工程与自然科学研究理事会;
关键词
hypoxia; ambulatory monitoring; vital signs; wearable devices;
D O I
10.1136/bmjopen-2019-034404
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Automated continuous ambulatory monitoring may provide an alternative to intermittent manual vital signs monitoring. This has the potential to improve frequency of measurements, timely escalation of care and patient safety. However, a major barrier to the implementation of these wearable devices in the ward environment is their uncertain reliability, efficiency and data fidelity. The purpose of this study is to test performance of selected devices in a simulated clinical setting including during movement and low levels of peripheral oxygen saturation. Methods and analysis This is a single centre, prospective, controlled, cross-sectional, diagnostic accuracy study to determine the specificity and sensitivity of currently available ambulatory vital signs monitoring equipment in the detection of hypoxia and the effect of movement on data acquisition. We will recruit up to 45 healthy volunteers who will attend a single study visit; starting with a movement phase and followed by the hypoxia exposure phase where we will gradually decrease saturation levels down to 80%. We will simultaneously test one chest patch, one wrist worn only and three wrist worn with finger probe devices against 'clinical standard 'and 'gold standard' references. We will measure peripheral oxygen saturations, pulse rate, heart rate and respiratory rate continuously and arterial blood gases intermittently throughout the study. Ethics and dissemination This study has received ethical approval by the East of Scotland Research Ethics Service REC 2 (19/ES/0008). The results will be broadly distributed through conference presentations and peer-reviewed publications.
引用
收藏
页数:8
相关论文
共 8 条
  • [1] A prospective, cross-sectional, multicentre study to evaluate the clinical performance of the ColoSTAT in vitro diagnostic for the detection of biomarkers associated with colorectal cancer
    He, Emily
    Van Hazel, Guy A.
    Sloss, Andrew
    Pianko, Stephen
    Brown, Gregor
    Clingan, Philip R.
    Singh, Rajvinder
    Solterbeck, Ann
    Traficante, Robert
    Lockett, Trevor
    Formby-Miller, Louise
    Macrae, Finlay
    JOURNAL OF CLINICAL ONCOLOGY, 2023, 41 (16)
  • [2] Diagnostic accuracy of simple tools in monitoring patients with chronic hypoventilation treated with non-invasive ventilation; a prospective cross-sectional study
    Aarrestad, Sigurd
    Qvarfort, Magnus
    Kleiven, Anne Louise
    Tollefsen, Elin
    Skjonsberg, Ole Henning
    Janssens, Jean-Paul
    RESPIRATORY MEDICINE, 2018, 144 : 30 - 35
  • [3] Diagnostic accuracy of a novel point-of-care urine lipoarabinomannan assay for the detection of tuberculosis among adult outpatients in Zambia: a prospective cross-sectional study
    Muyoyeta, Monde
    Kerkhoff, Andrew D.
    Chilukutu, Lophina
    Moreau, Emmanuel
    Schumacher, Samuel G.
    Ruhwald, Morten
    EUROPEAN RESPIRATORY JOURNAL, 2021, 58 (05)
  • [4] To determine the sensitivity, specificity, and diagnostic accuracy of diffusion-weighted MRI in localization of non-palpable undescended testes taking laparoscopic findings as the gold standard: A cross-sectional study from Pakistan
    Fazal, Kamran
    Hussain, Shaiq
    Khan, Faheemullah
    Ullah, Irfan
    Tahir, Muhammad Junaid
    Mehmood, Qasim
    Yousaf, Zohaib
    ANNALS OF MEDICINE AND SURGERY, 2022, 73
  • [5] Monitoring the threatened utility of malaria rapid diagnostic tests by novel high-throughput detection of Plasmodium falciparum hrp2 and hrp3 deletions: A cross-sectional, diagnostic accuracy study
    Kreidenweiss, Andrea
    Trauner, Franziska
    Rodi, Miriam
    Koehne, Erik
    Held, Jana
    Wyndorps, Lea
    Manouana, Gedeon Prince
    McCall, Matthew
    Adegnika, Ayola Akim
    Lalremruata, Albert
    Kremsner, Peter G.
    Fendel, Rolf
    Sandri, Thaisa Lucas
    EBIOMEDICINE, 2019, 50 : 14 - 22
  • [6] Cross-sectional area variations of internal jugular veins during supine head rotation in multiple sclerosis patients with chronic cerebrospinal venous insufficiency: a prospective diagnostic controlled study with duplex ultrasound investigation
    Farina, Massimiliano
    Novelli, Eugenio
    Pagani, Raffaello
    BMC NEUROLOGY, 2013, 13
  • [7] Cross-sectional area variations of internal jugular veins during supine head rotation in multiple sclerosis patients with chronic cerebrospinal venous insufficiency: a prospective diagnostic controlled study with duplex ultrasound investigation
    Massimiliano Farina
    Eugenio Novelli
    Raffaello Pagani
    BMC Neurology, 13
  • [8] Development of machine learning support for reading whole body diffusion-weighted MRI (WB-MRI) in myeloma for the detection and quantification of the extent of disease before and after treatment (MALIMAR): protocol for a cross-sectional diagnostic test accuracy study
    Satchwell, Laura
    Wedlake, Linda
    Greenlay, Emily
    Li, Xingfeng
    Messiou, Christina
    Glocker, Ben
    Barwick, Tara
    Barfoot, Theodore
    Doran, Simon
    Leach, Martin O.
    Koh, Dow Mu
    Kaiser, Martin
    Winzeck, Stefan
    Qaiser, Talha
    Aboagye, Eric
    Rockall, Andrea