Dosing accuracy of two insulin pre-filled pens

被引:17
|
作者
Asakura, Toshinari [1 ]
Seino, Hiroaki [3 ]
Kageyama, Miho [2 ]
Yohkoh, Noriaki [1 ]
机构
[1] Niigata Univ Pharm & Appl Life Sci, Dept Clin Pharm, Fac Pharmaceut Sci, Niigata, Japan
[2] Ohta Nishinouchi Hosp, Koriyama, Fukushima, Japan
[3] Seino Clin, Koriyama, Fukushima, Japan
关键词
dose accuracy; FlexPen; insulin pen; SoloStar;
D O I
10.1185/030079908X297394
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: This study was designed to determine the dose accuracy of two commonly available insulin pre-filled pens for use in diabetes - FlexPen* (Novo Nordisk A/S, Bagsvaerd, Denmark) (FP) and SoloStart (Sanofi-Aventis, Paris, France) (SS). Research design and methods: Dosing accuracy was tested at 5 U, 10 U and 30 U doses - three previously unused pens of both pre-filled pens were used for each dose. At the 5 U dose each pen was tested 42 times, at the 10 U dose each pen was tested 25 times and at the 30 U dose each pen was tested 9 times. The pre-filled pens were used strictly according to manufacturers' instructions and measurements were made on a sensitive balance and corrected for the specific density of the insulin formulations used. Specified limits were based on ISO standards (+/- 1 U for the 5 U and 10 U doses and +/- 1.5 U for the 30 U dose). Results: FP was more accurate for injecting 5 U, 10 U and 30 U doses (absolute mean 4.95 +/- 0.19 U, 9.61 +/- 0.27 U and 29.70 +/- 0.34 U, respectively) than SS (absolute mean 4.86 +/- 0.39 U, 9.27 +/- 0.52 U and 28.73 +/- 0.47 U, respectively). No doses were outside specified limits for 5 U and 30 U with FP and 1.3% of doses were outside these limits at the 10 U dose. For SS; 1.6%, 29.3% and 33.3% of doses were below the pre-specified threshold for the 5 U, 10 U and 30 U doses, respectively. Conclusions: This non-blinded comparison indicates that FP is a more accurate insulin pre-filled pen than SS at three different insulin doses. In this single-user study, FP delivered consistent and accurate doses of insulin but SS had a high frequency of under dosing. No assessment was made of user variability in this study.
引用
收藏
页码:1429 / 1434
页数:6
相关论文
共 50 条
  • [1] Comment and reply on: Dosing accuracy of two insulin pre-filled pens
    Vaur, Laurent
    [J]. CURRENT MEDICAL RESEARCH AND OPINION, 2008, 24 (08) : 2373 - 2374
  • [2] Comment and reply on: Dosing accuracy of two insulin pre-filled pens - Reply
    Asakura, Toshinari
    Seino, Hiroaki
    Kageyama, Miho
    Yohkoh, Noriaki
    [J]. CURRENT MEDICAL RESEARCH AND OPINION, 2008, 24 (08) : 2374 - 2375
  • [3] Study of the accuracy of two different types of pre-filled insulin pens
    不详
    [J]. METABOLISMES HORMONES DIABETES ET NUTRITION, 2008, 12 (06): : 204 - 204
  • [5] Dosing accuracy of commonly used disposable insulin pens
    Krzywon, M.
    Abdel-Tawab, M.
    van der Burg, T.
    Fuhr, U.
    Schubert-Zsilavecz, M.
    [J]. CURRENT MEDICAL RESEARCH AND OPINION, 2010, 26 (04) : 901 - 905
  • [6] Study on the Dosing Accuracy of Common Disposable Insulin Pens
    Krzywon, Meike
    Van Der Burg, Thomas
    Fuhr, Uwe
    Schubert-Zsilavecz, Manfred
    Abdel-Tawab, Mona
    [J]. DIABETES, 2010, 59 : A142 - A143
  • [7] Dosing accuracy of different insulin glargine pens available in India
    Bohnet, J.
    Korger, V.
    Schubert-Zsilavecz, M.
    Abdel-Tawab, M.
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 2014, 103 : S61 - S61
  • [8] Study on the Dosing Accuracy of Commonly Used Disposable Insulin Pens
    Krzywon, Meike
    van der Burg, Thomas
    Fuhr, Uwe
    Schubert-Zsilavecz, Manfred
    Abdel-Tawab, Mona
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (09) : 804 - 809
  • [9] Dosing Accuracy of Insulin Pens Versus Conventional Syringes and Vials
    Luijf, Y. M.
    DeVries, J. H.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 : S73 - S77
  • [10] Experience with Pharmacist Prepared Pre-filled SCIT Dosing in an Urban Allergy Clinic
    Nicolaides, Rory
    Williams, Marya
    Wright, Reshma
    Khan, David A.
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2019, 143 (02) : AB306 - AB306