Automation of Monte Carlo-based treatment plan verification for proton therapy

被引:8
|
作者
Kaluarachchi, Maduka [1 ]
Moskvin, Vadim [1 ]
Pirlepesov, Fakhriddin [1 ]
Wilson, Lydia J. [1 ]
Xie, Fang [1 ]
Faught, Austin M. [1 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiat Oncol, 332 N Lauderdale St, Memphis, TN 38105 USA
来源
关键词
automation; Monte Carlo; proton therapy; BRAIN-STEM; BRAGG PEAK; SIMULATION; HETEROGENEITIES; PARAMETERS; TOPAS;
D O I
10.1002/acm2.12923
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Independent calculations of proton therapy plans are an important quality control procedure in treatment planning. When using custom Monte Carlo (MC) models of the beamline, deploying the calculations can be laborious, time consuming, and require in-depth knowledge of the computational environment. We developed an automated framework to remove these barriers and integrate our MC model into the clinical workflow. Materials and Methods The Eclipse Scripting Application Programming Interface was used to initiate the automation process. A series of MATLAB scripts were then used for preprocessing of input data and postprocessing of results. Additional scripts were used to monitor the calculation process and appropriately deploy calculations to an institutional high-performance computing facility. The automated framework and beamline models were validated against 160 patient specific QA measurements from an ionization chamber array and using a +/- 3%/3 mm gamma criteria. Results The automation reduced the human-hours required to initiate and run a calculation to 1-2 min without leaving the treatment planning system environment. Validation comparisons had an average passing rate of 99.4% and were performed at depths ranging from 1 to 15 cm. Conclusion An automated framework for running MC calculations was developed which enables the calculation of dose and linear energy transfer within a clinically relevant workflow and timeline. The models and framework were validated against patient specific QA measurements and exhibited excellent agreement. Before this implementation, execution was prohibitively complex for an untrained individual and its use restricted to a research environment.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 50 条
  • [1] Monte Carlo-based Inverse Treatment Plan Optimization for Intensity Modulated Proton Therapy
    Li, Y.
    Tian, Z.
    Song, T.
    Wu, Z.
    Liu, Y.
    Jiang, S. B.
    Jia, X.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 475 - 478
  • [2] Monte Carlo-Based Inverse Treatment Plan Optimization for Intensity Modulated Proton Therapy
    Li, Y.
    Tian, Z.
    Song, T.
    Wu, Z.
    Liu, Y.
    Jiang, S.
    Jia, X.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3616 - 3616
  • [3] A Monte Carlo-based treatment planning tool for proton therapy
    Mairani, A.
    Bohlen, T. T.
    Schiavi, A.
    Tessonnier, T.
    Molinelli, S.
    Brons, S.
    Battistoni, G.
    Parodi, K.
    Patera, V.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08): : 2471 - 2490
  • [4] Verification of a Monte Carlo-based treatment planning toot
    Chow, J.
    Leung, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S92 - S92
  • [5] A Monte Carlo Based Plan Verification Framework for Double-Scattering Proton Therapy
    Yuan, J.
    Mansur, D.
    Yao, M.
    Biswas, T.
    Jesseph, R.
    Jin, J.
    Machtay, M.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E551 - E551
  • [7] Monte carlo calculations supporting Patient Plan Verification in Proton Therapy
    Lima, Thiago V. M.
    Dosanjh, Manjit
    Ferrari, Alfredo
    Molineli, Silvia
    Ciocca, Mario
    Mairani, Andrea
    [J]. FRONTIERS IN ONCOLOGY, 2016, 6
  • [8] Monte Carlo calculations supporting patient plan verification in proton therapy
    Lima, Thiago V. M.
    Dosanjh, Manjit
    Ferrari, Alfredo
    Molineli, Silvia
    Ciocca, Mario
    Mairani, Andrea
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 863 - 863
  • [9] A Monte Carlo-based tool for proton dose verification on patient CT data
    Szymanowski, H
    Nill, S
    Oelfke, U
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S319 - S319
  • [10] A new Monte Carlo-based treatment plan optimization approach for intensity modulated radiation therapy
    Li, Yongbao
    Tian, Zhen
    Shi, Feng
    Song, Ting
    Wu, Zhaoxia
    Liu, Yaqiang
    Jiang, Steve
    Jia, Xun