The accuracy of the palpation meter (PALM) for measuring pelvic crest height difference and leg length discrepancy

被引:52
|
作者
Petrone, MR
Guinn, J
Reddin, A
Sutlive, TG
Flynn, TW
Garber, MP
机构
[1] Moncrief Army Community Hosp, Ft Jackson, SC USA
[2] Tripler Army Med Ctr, Honolulu, HI 96859 USA
[3] Blanchfield Army Community Hosp, Ft Campbell, KY USA
[4] USA, Baylor Univ, Grad Program Phys Therapy, Ft Sam Houston, TX USA
[5] USA Command & Gen Staff Coll, Ft Leavenworth, KS USA
[6] Brooke Army Med Ctr, USA, Orthoped Phys Therapy Residency, Ft Sam Houston, TX 78234 USA
来源
关键词
leg length inequality; measurement; pelvic obliquity; reliability; validity;
D O I
10.2519/jospt.2003.33.6.319
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Study Design: Test-retest reliability and validity. Objective: To determine the validity and reliability of the Palpation Meter (PALM). Background: Leg length discrepancy (LLD) has been associated with a variety of musculoskeletal disorders. Therefore, the clinical measurement of LLD has become a routine and important part of the physical examination. The PALM is an instrument that was recently developed to indirectly measure LLD, but little is known about its measurement properties. Methods and Measures: Fifteen healthy and 15 symptomatic subjects with suspected LLD participated in this study. Measurements of pelvic crest height difference (PD) were obtained by 2 examiners using the PALM. A standing antero-posterior (AP) radiograph of each subject's pelvis was taken, and PD and LLD (femoral head height difference) were determined from the radiograph for comparison with the PALM values. Intraclass correlation coefficients (ICCs) were calculated to determine the validity and reliability estimates of the PALM. Results: For all subjects, the validity estimates (ICC2,3) of the PALM for PD were excellent (0.90 for rater 1 and 0.92 for rater 2) when compared with the standing AP radiograph of the pelvis. The PALM was less accurate (ICC2,3 of 0.76 and 0.78 for rater 1 and 2, respectively) as an indirect estimate of LLD. Intrarater reliability for each rater was excellent (ICC2,3 = 0.97 and 0.98) and interrater reliability was very good (ICC2,3 = 0.88). Conclusion: The PALM is a reliable and valid instrument for measuring PD. Clinicians should consider this convenient, cost-effective clinical tool as an alternative to radiographic measurement of pelvic crest height difference.
引用
收藏
页码:319 / 325
页数:7
相关论文
共 46 条
  • [1] Measuring leg-length discrepancy by the "iliac crest palpation and book correction" method: Reliability and validity
    Hanada, E
    Kirby, RL
    Mitchell, M
    Swuste, JM
    [J]. ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2001, 82 (07): : 938 - 942
  • [2] Development and evaluation of deep-learning measurement of leg length discrepancy: bilateral iliac crest height difference measurement
    Kim, Min Jong
    Choi, Young Hun
    Lee, Seul Bi
    Cho, Yeon Jin
    Lee, Seung Hyun
    Shin, Chang Ho
    Shin, Su-Mi
    Cheon, Jung-Eun
    [J]. PEDIATRIC RADIOLOGY, 2022, 52 (11) : 2197 - 2205
  • [3] Development and evaluation of deep-learning measurement of leg length discrepancy: bilateral iliac crest height difference measurement
    Min Jong Kim
    Young Hun Choi
    Seul Bi Lee
    Yeon Jin Cho
    Seung Hyun Lee
    Chang Ho Shin
    Su-Mi Shin
    Jung-Eun Cheon
    [J]. Pediatric Radiology, 2022, 52 : 2197 - 2205
  • [4] PELVIC OBLIQUITY AND LEG LENGTH DISCREPANCY
    GRILL, F
    CHOCHOLE, M
    SCHULTZ, A
    [J]. ORTHOPADE, 1990, 19 (05): : 244 - 262
  • [5] CT SCANOGRAMS FOR MEASURING LEG LENGTH DISCREPANCY
    HELMS, CA
    MCCARTHY, S
    [J]. RADIOLOGY, 1984, 151 (03) : 802 - 802
  • [6] Pelvic imbalance and leg length discrepancy in Marfan scoliosis
    Mosconi, Mario
    De Rosa, Federica
    Cuzzocrea, Fabrizio
    Martocchi, Chiara
    Pelillo, Francesco
    Costa, Pietro
    Peschiera, Valeria
    Urtis, Mario
    Arbustini, Eloisa
    [J]. MINERVA ORTHOPEDICS, 2021, 72 (03): : 308 - 312
  • [7] EMGs analysis of lumbar, pelvic and leg muscles in leg length discrepancy adolescents
    Sotelo-Barroso, F
    Márquez-Gamiño, S
    Caudillo-Cisneros, C
    [J]. MEDICAL PHYSICS, 2004, 724 : 203 - 205
  • [8] ASSESSMENT OF THE ACCURACY OF 3 TECHNIQUES TO PREDICT LEG LENGTH DISCREPANCY
    SMITH, DJ
    ZINDRICK, MR
    LAMBERT, RW
    MILLAR, EA
    [J]. ORTHOPEDICS, 1982, 5 (06) : 737 - 738
  • [9] MEASURING THE LEG LENGTH AND LEG LENGTH DIFFERENCE BY MEANS OF REAL-TIME-SONOGRAPHY
    HOLST, A
    THOMAS, W
    [J]. INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1988, 9 (05) : 397 - 397
  • [10] Assessing the accuracy of measuring leg length discrepancy and genu varum/valgum using a markerless motion analysis system
    Hurley, Robert J.
    Davey, Martin S.
    Newell, Micheal
    Devitt, Aiden
    [J]. JOURNAL OF ORTHOPAEDICS, 2021, 26 : 45 - 48