Complications of Cardiac Perforation and Lead Dislodgement with an MRI-Conditional Pacing Lead: A Korean Multi-Center Experience

被引:7
|
作者
Kwon, Chang Hee [1 ]
Choi, Jin Hee [2 ]
Kim, Jun [2 ]
Jo, Uk [2 ]
Lee, Ji Hyun [2 ]
Lee, Woo Seok [3 ]
Kim, Yoo Ri [4 ]
Lee, Soo Yong [5 ]
Whang, Ki Won [5 ]
Yang, Jihyun [6 ]
Kim, Sung-Hwan [6 ]
Oh, Yong-Seog [6 ]
Park, Kyoung-Min [7 ]
Nam, Gi-Byoung [2 ]
Choi, Kee-Joon [2 ]
Kim, You-Ho [2 ]
机构
[1] Konkuk Univ, Sch Med, Div Cardiol, Dept Internal Med,Med Ctr, Seoul, South Korea
[2] Univ Ulsan, Div Cardiol, Dept Internal Med, Asan Med Ctr,Coll Med, 88 Olymp Ro 43 Gil, Seoul 05505, South Korea
[3] Vet Hlth Serv Med Ctr, Div Cardiol, Dept Internal Med, Seoul, South Korea
[4] Catholic Univ Korea, Incheon St Marys Hosp, Div Cardiol, Dept Internal Med, Inchon, South Korea
[5] Pusan Natl Univ, Div Cardiol, Dept Internal Med, Yangsan Hosp, Yangsan, South Korea
[6] Catholic Univ Korea, Seoul St Marys Hosp, Div Cardiol, Dept Internal Med, Seoul, South Korea
[7] Sungkyunkwan Univ, Heart Vasc Stroke Inst, Div Cardiol, Samsung Med Ctr,Dept Internal Med,Sch Med, Seoul, South Korea
关键词
Complications; Magnetic Resonance Imaging; Pacemaker; Cardiac Perforation; RIGHT-VENTRICULAR PERFORATION; PACEMAKER IMPLANTATION; SINGLE-CENTER; DEVICE IMPLANTATION; PREDICTORS; MANAGEMENT; SUBACUTE; THERAPY; SYSTEM; SAFETY;
D O I
10.3346/jkms.2016.31.9.1397
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Medtronic CapSureFix MRI 5086 pacing lead (5086; Medtronic, Inc., Minneapolis, MN, USA) has been reported to be associated with increased cardiac perforation and lead dislodgement. This study aimed to compare the incidence of cardiac perforation and lead dislodgement within 30 days after pacemaker implantation between 5086 MRI lead and previous Medtronic CapSureFix Novus 5076 non-MRI pacing lead. This was a nationwide, multicenter retrospective study in which we compared the incidence of adverse events between 277 patients implanted with 5086 lead and 205 patients implanted with 5076 lead between March 2009 and September 2014. Cardiac perforation within 30 days of pacemaker implantation occurred in 4 patients (1.4%) with the 5086 lead and in no patient with the 5076 lead (P = 0.084). Lead dislodgement occurred in 8 patients (2.9%) with the 5086 lead and in 5 patients (2.4%) with the 5076 lead (P = 0.764). On multivariate logistic regression analysis, age was significantly associated with cardiac perforation. Congestive heart failure and implantation of right atrial (RA) lead at RA free wall or septum were significant factors for the incidence of lead dislodgement and lead revision. The incidence of cardiac perforation and lead dislodgement were not statistically different between the patients with 5086 lead and the patients with 5076 lead. However, careful attention for cardiac perforation may be needed when using the 5086 MRI lead, especially in elderly patients.
引用
收藏
页码:1397 / 1402
页数:6
相关论文
共 13 条
  • [1] Complications with Cardiac Perforation and Lead Dislodgement of the 5086 MRI Pacing Lead
    Choi, Jin Hee
    Kim, Jun
    Jo, Uk
    Kwon, Chang Hee
    Lee, Woo Seok
    Kim, Yoo Ri
    Nam, Gi-Byoung
    Choi, Kee-Joon
    Kim, You-Ho
    [J]. CIRCULATION, 2014, 130
  • [2] Cardiac Perforation and Lead Dislodgement after Implantation of a MR-Conditional Pacing Lead: A Single-Center Experience
    Elmouchi, Darryl A.
    Rosema, Shelly
    Vanoosterhout, Stacie M.
    Khan, Mohsin
    Davis, Alan T.
    Gauri, Andre J.
    Finta, Bohuslav
    Woelfel, Alan K.
    Chalfoun, Nagib T.
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (01): : 4 - 10
  • [3] Increased Perforation Risk with an MRI-Conditional Pacing Lead: A Single-Center Study
    Acha, Moshe Rav
    Keaney, John J.
    Lubitz, Steven A.
    Milan, David J.
    Mansour, Moussa
    Heist, Kevin E.
    Ptaszek, Leon M.
    Singh, Jagmeet P.
    Blendea, Dan
    Mela, Theofanie
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2015, 38 (03): : 334 - 342
  • [4] CLINICAL EXPERIENCE WITH A QUADRIPOLAR TRANSVENOUS LEAD FOR LEFT VENTRICULAR PACING: PRELIMINARY RESULTS OF A PROSPECTIVE MULTI-CENTER STUDY
    Forleo, G.
    Della Rocca, D.
    Panattoni, G.
    Di Molfetta, A.
    Papavasileiou, L.
    Sergi, D.
    Magliano, G.
    Cioe, R.
    Santini, L.
    Romeo, F.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S56 - S56
  • [5] Single-center experience of a quadripolar pacing lead for cardiac resynchronization therapy
    Antonello Vado
    Endrj Menardi
    Guido Rossetti
    Gianpaolo Ballari
    Mauro Feola
    Marco Bobbio
    [J]. Journal of Interventional Cardiac Electrophysiology, 2014, 39 : 161 - 165
  • [6] Single-center experience of a quadripolar pacing lead for cardiac resynchronization therapy
    Vado, Antonello
    Menardi, Endrj
    Rossetti, Guido
    Ballari, Gianpaolo
    Feola, Mauro
    Bobbio, Marco
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2014, 39 (02) : 161 - 165
  • [7] Initial Single-Center Experience of a Quadripolar Pacing Lead for Cardiac Resynchronization Therapy
    Shetty, Anoop K.
    Duckett, Simon G.
    Bostock, Julian
    Roy, Debashis
    Ginks, Matthew
    Hamid, Shoaib
    Rosenthal, Eric
    Razavi, Reza
    Rinaldi, Christopher Aldo
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (04): : 484 - 489
  • [8] MULTI-CENTER CLINICAL-EXPERIENCE WITH AN IMPROVED STEROID-ELUTING PACEMAKER LEAD
    SCHALLHORN, R
    OLESON, K
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (04): : 496 - 496
  • [9] A Retrospective Analysis of a Single-Center Experience With the Tempo Temporary Pacing Lead in Management of Bradyarrhythmia and Other Cardiac Electrophysiology Procedures
    Stelzer, Mitchell
    Fu, Eugene
    Kleman, James
    Chopra, Nagesh
    Padmanabhan, Sandosh
    Kidwell, Gregory A.
    Amin, Anish
    [J]. CIRCULATION, 2021, 144
  • [10] Clinical experience in small area unipolar membrane-covered titanium electrode. Multi-center lead study
    Pieper, R
    Rachor, M
    Bondke, H
    Nilsson, L
    Schneider, W
    Maru, F
    Vesterlund, T
    Arstila, M
    [J]. PACEMAKER LEADS 1997, 1997, : 333 - 336