1D Multi-element CMUT arrays for ultrasound thermal therapy

被引:0
|
作者
N'Djin, William Apoutou [1 ,2 ]
Canney, Michael [1 ,2 ,4 ]
Meynier, Cyril [3 ]
Chavrier, Francoise [1 ,2 ]
Lafon, Cyril [1 ,2 ]
Nguyen-Dinh, An [3 ]
Chapelon, Jean-Yves [1 ,2 ]
Carpentier, Alexandre [5 ]
机构
[1] INSERM, U1032, LabTau, F-69003 Lyon, France
[2] Univ Lyon, F-69003 Lyon, France
[3] Vermon SA, Tours, France
[4] Carthera Brain & Spine Inst, Paris, France
[5] Sorbonne Univ Paris 6, Hop Pitie Salpetriere, AP HP, Neurosurg Dept, F-75013 Paris, France
关键词
CMUT; ultrasound therapy; HICU; thermal ablation; FEASIBILITY;
D O I
10.1063/1.4977621
中图分类号
O59 [应用物理学];
学科分类号
摘要
Interstitial therapeutic ultrasound devices are a promising technology for performing thermal ablation in a wide variety of organs. In this study, the use of Capacitive Micromachined Ultrasound Transducers (CMUTs) for interstitial heating applications was investigated. CMUTs exhibit potential advantages for use in therapeutic ultrasound applications in comparison to standard piezo ultrasound transducer technologies as they have good characteristics in terms of miniaturization (cell size: few dozens of microns), bandwidth (several MHz) and high electro-acoustic efficiency. Two designs of CMUT arrays were studied: (1) a 1D 128-element planar-CMUT array originally dedicated to abdominal ultrasound imaging purposes (5 MHz, element size: 0.3 x 8.0 mm(2)); (2) a 12-element linear-array, 32.4-mm long and 0.8-mm wide, developed specifically for minimally-invasive interstitial therapeutic applications (6 MHz, element size: 2.7 x 0.8 mm(2)). Simulations were performed to evaluate the ability to generate thermal lesions in soft tissues with: (1) 1 single linear array, (2) a combination of multiple linear arrays positioned on a cylindrical catheter. Experimental investigations performed with the CMUT imaging array showed the ability to generate surface acoustic intensities (I-ac) up to 20 W . cm(-2) and to generate intense centimetric thermal lesions in in-vitro turkey breast tissues. At 6 MHz, a single element was able to generate in water a maximum peak pressure of >0.5 MPa. In simulations, the ability to use various power levels and frequencies on independent elements, as well as combinations of multiple linear-arrays offered sufficient flexibility to achieve a wide variety of thermal ablation patterns in 3D. Simulated ablation volumes could be controlled to cover accurately non-symmetrical volumes of brain metastases. In conclusion, CMUT arrays show interesting characteristics, which may open new perspectives of spatial control for conformal interstitial thermal therapy with miniaturized multi-element therapeutic ultrasound catheters.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Large Area 1D CMUT Phased Arrays for Multi-Modality Ultrasound Imaging
    Apte, Nikhil
    Vaithilingam, Srikant
    Sarioglu, Ali Fatih
    Kupnik, Mario
    Khuri-Yakub, Butrus T.
    2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 612 - 615
  • [2] A multi-element interstitial ultrasound applicator for the thermal therapy of brain tumors
    Canney, Michael S.
    Chavrier, Francoise
    Tsysar, Sergey
    Chapelon, Jean-Yves
    Lafon, Cyril
    Carpentier, Alexandre
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2013, 134 (02): : 1647 - 1655
  • [3] CMUT-based single element transducer applied to 1D transient ultrasound elastography
    Certon, Dominique
    Audiere, Stephane
    Colin, Laurent
    Senegond, Nicolas
    Alquier, Daniel
    Gregoire, Jean-Marc
    Sandrin, Laurent
    Miette, Veronique
    2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2018,
  • [4] MULTI-ELEMENT TRANSDUCER ARRAYS
    MCELROY, JT
    MATERIALS EVALUATION, 1977, 35 (03) : S5 - S5
  • [5] Multi-Element Transducers For Conformal Transurethral Thermal Therapy
    Baker, Nicole
    Desouza, Alexander
    Yin, Jianhua
    Boyes, Aaron
    Lukacs, Marc
    Chopra, Rajiv
    Foster, Stuart
    Bronskill, Michael
    THERAPEUTIC ULTRASOUND, 2006, 829 : 415 - +
  • [6] Characterization of Multi-element CZT arrays
    Cirignano, L
    Shah, KS
    Bennett, P
    Li, L
    Lu, F
    Buturlia, J
    Yao, W
    Wright, G
    James, RB
    HARD X-RAY GAMMA-RAY AND NEUTRON DETECTOR PHYSICS II, 2000, 4141 : 23 - 28
  • [7] MR-Guided Interstitial Thermal Therapy for the Treatment of Brain Tumors with a Multi-Element Ultrasound Probe
    Canney, Michael
    Carpentier, Alexandre
    Beccaria, Kevin
    Souchon, Remi
    Chavrier, Francoise
    Lafon, Cyril
    Chapelon, Jean-Yves
    11TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1481 : 32 - 36
  • [8] CUTTING TECHNIQUES FOR MULTI-ELEMENT MINIATURE ARRAYS
    ABBOTT, JG
    BAUER, BJ
    ULTRASONIC IMAGING, 1980, 2 (02) : 189 - 189
  • [9] Advances in InGaAs multi-element linear arrays
    Gasparian, GA
    Schaeffer, PM
    OPTICAL ONLINE INDUSTRIAL PROCESS MONITORING, 1999, 3859 : 29 - 37
  • [10] ASSESSMENT OF DISTORTIONS PRODUCED BY MULTI-ELEMENT ARRAYS
    GORE, JC
    LEEMAN, S
    PRIDIE, RB
    BRITISH JOURNAL OF RADIOLOGY, 1978, 51 (603): : 231 - 231