Neonatal and Adult ICU Ventilators to Provide Ventilation in Neonates, Infants, and Children: A Bench Model Study

被引:24
|
作者
Vignaux, Laurence [1 ,2 ,3 ]
Piquilloud, Lise [4 ]
Tourneux, Pierre [2 ,5 ]
Jolliet, Philippe [4 ]
Rimensberger, Peter C. [6 ]
机构
[1] Hop La Tour, Cardioresp Physiotherapy Unit, CH-1217 Meyrin, Switzerland
[2] Univ Picardie Jules Verne, Inst Natl Environm Ind & Risques, EA 4284, Peritox,Unite Mixte 101, Amiens, France
[3] Univ Hosp Geneva, Adult Intens Care Unit, Geneva, Switzerland
[4] Univ Lausanne Hosp, Adult Intens Care & Burn Unit, Lausanne, Switzerland
[5] Univ Hosp North, Pediat Intens Care Unit, Amiens, France
[6] Univ Hosp Geneva, Neonatal & Pediat Intens Care Unit, Geneva, Switzerland
关键词
mechanical ventilators; ventilatory support; neonatal; pediatrics; intensive care units; equipment safety; respiration; artificial; models; PRESSURE-SUPPORT VENTILATION; ACUTE RESPIRATORY-FAILURE; NONINVASIVE VENTILATION; MANDATORY VENTILATION; TIDAL VOLUME; ASSIST; ASYNCHRONY; PERFORMANCE; NAVA; FLOW;
D O I
10.4187/respcare.02540
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS: PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate triggering, pressurization, and cycling in both the absence and presence of leaks. Delivered tidal volumes were also assessed. Three patients were simulated: a preterm infant (resistance 100 cm H2O/L/s, compliance 2 mL/cm H2O, inspiratory time of the patient [T-1] 400 ms, inspiratory effort 1 and 2 cm H2O), a full-term infant (resistance 50 cm H2O/L/s, compliance 5 mL/cm H2O, T-1 500 ms, inspiratory effort 2 and 4 cm H2O), and a child (resistance 30 cm H2O/L/s, compliance 10 mL/cm H2O, T-1 600 ms, inspiratory effort 5 and 10 cm H2O). Two PS levels were tested (10 and 15 cm H2O) with and without leaks and with and without the leak compensation algorithm activated. RESULTS: Without leaks, only 2 neonatal ventilators and one adult ventilator had trigger delays under a given predefined acceptable limit (1/8 T-1). Pressurization showed high variability between ventilators. Most ventilators showed T-1 in excess high enough to seriously impair patient-ventilator synchronization (> 50% of the T-1 of the subject). In some ventilators, leaks led to autotriggering and impairment of ventilation performance, but the influence of leaks was generally lower in neonatal ventilators. When a noninvasive ventilation algorithm was available, this was partially corrected. In general, tidal volume was calculated too low by the ventilators in the presence of leaks; the noninvasive ventilation algorithm was able to correct this difference in only 2 adult ventilators. CONCLUSIONS: No ventilator performed equally well under all tested conditions for all explored parameters. However, neonatal ventilators tended to perform better in the presence of leaks. These findings emphasize the need to improve algorithms for assisted ventilation modes to better deal with situations of high airway resistance, low pulmonary compliance, and the presence of leaks.
引用
收藏
页码:1463 / 1475
页数:13
相关论文
共 50 条
  • [1] Neonatal and Adult ICU Ventilators to Provide Ventilation in Neonates, Infants, and Children: A Bench Model Study (vol 59, pg 1463, 2014)
    Vignaux, Laurence
    Piquilloud, Lise
    Tourneux, Pierre
    Jolliet, Philippe
    Rimensberger, Peter C.
    RESPIRATORY CARE, 2015, 60 (02) : E43 - E45
  • [2] Adult ICU ventilators to provide neonatal ventilation: a lung simulator study
    Andrew D. Marchese
    Daniel Chipman
    Pedro de la Oliva
    Robert M. Kacmarek
    Intensive Care Medicine, 2009, 35 : 631 - 638
  • [3] Adult ICU ventilators to provide neonatal ventilation: a lung simulator study
    Marchese, Andrew D.
    Chipman, Daniel
    de la Oliva, Pedro
    Kacmarek, Robert M.
    INTENSIVE CARE MEDICINE, 2009, 35 (04) : 631 - 638
  • [4] Comment on “Adult ICU ventilators to provide neonatal ventilation: a lung simulator study”
    Matteo Fontana
    Antoine Payot
    Sylvain Morneau
    Philippe Jouvet
    Intensive Care Medicine, 2009, 35 : 1140 - 1141
  • [5] Comment on "Adult ICU ventilators to provide neonatal ventilation: a lung simulator study"
    Fontana, Matteo
    Payot, Antoine
    Morneau, Sylvain
    Jouvet, Philippe
    INTENSIVE CARE MEDICINE, 2009, 35 (06) : 1140 - 1141
  • [6] Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study
    Laurence Vignaux
    Didier Tassaux
    Philippe Jolliet
    Intensive Care Medicine, 2007, 33 : 1444 - 1451
  • [7] Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study
    Vignaux, Laurence
    Tassaux, Didier
    Jolliet, Philippe
    INTENSIVE CARE MEDICINE, 2007, 33 (08) : 1444 - 1451
  • [8] Performance of ICU ventilators during noninvasive ventilation with large leaks in a total face mask: a bench study
    Viyuki Nakamura, Maria Aparecida
    Vieira Costa, Eduardo Leite
    Ribeiro Carvalho, Carlos Roberto
    Tucci, Vlauro Roberto
    JORNAL BRASILEIRO DE PNEUMOLOGIA, 2014, 40 (03) : 294 - 303
  • [9] Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study
    Ferreira, Juliana C.
    Chipman, Daniel W.
    Kacmarek, Robert M.
    INTENSIVE CARE MEDICINE, 2008, 34 (09) : 1669 - 1675
  • [10] Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study
    Juliana C. Ferreira
    Daniel W. Chipman
    Robert M. Kacmarek
    Intensive Care Medicine, 2008, 34 : 1669 - 1675