Benchmarking daily adaptation using fully automated radiotherapy treatment plan optimization for rectal cancer

被引:6
|
作者
Jagt, Thyrza Z. [1 ]
Janssen, Tomas M. [1 ]
Betgen, Anja [1 ]
Wiersema, Lisa [1 ]
Verhage, Rick [1 ]
Garritsen, Sanne [1 ]
Vijlbrief-Bosman, Tineke [1 ]
de Ruiter, Peter [1 ]
Remeijer, Peter [1 ]
Marijnen, Corrie A. M. [1 ]
Peters, Femke P. [1 ]
Sonke, Jan -Jakob [1 ,2 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
[2] Netherlands Canc Inst, Dept Radiat Oncol, Plesmanlaan 121, NL-1066 CX Amsterdam, Netherlands
关键词
Online plan adaptation; MRI guided radiotherapy; Automated treatment planning; MRI SCANNER; ACCELERATOR; METHODOLOGY; GENERATION;
D O I
10.1016/j.phro.2022.08.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background/purpose: In daily plan adaptation the radiotherapy treatment plan is adjusted just prior to delivery. A simple approach is taking the planning objectives of the reference plan and directly applying these in re -optimization. Here we present a tested method to verify whether daily adaptation without tweaking of the objectives can maintain the plan quality throughout treatment. Materials/methods: For fifteen rectal cancer patients, automated treatment planning was used to generate plans mimicking manual reference plans on the planning scans. For 74 fraction scans (4-5 per patient) an automated plan and a daily adapted plan were generated, where the latter re-optimizes the reference plan objectives without any tweaking. To evaluate the robustness of the daily adaptation, the adapted plans were compared to the autoplanning plans.Results: Median differences between the autoplanning plans on the planning scans and the reference plans were between-1 and 0.2 Gy. The largest interquartile range (1 Gy) was seen for the Lumbar Skin D2%. For the daily scans the PTV D2% and D98% differences between autoplanning and adapted plans were within +/- 0.7 Gy, with mean differences within +/- 0.3 Gy. Positive differences indicate higher values were obtained using autoplanning. For the Bowelarea + Bladder and the Lumbar Skin the D2% and Dmean differences were all within +/- 2.6 Gy, with mean differences between-0.9 and 0.1 Gy.Conclusion: Automated treatment planning can be used to benchmark daily adaptation techniques. The inves-tigated adaptation workflow can robustly perform high quality adaptations without daily adjusting of the patient-specific planning objectives for rectal cancer radiotherapy.
引用
收藏
页码:7 / 13
页数:7
相关论文
共 50 条
  • [1] Fully automated radiotherapy treatment planning: A scan to plan challenge
    Gooding, Mark J.
    Aluwini, Shafak
    Urbano, Teresa Guerrero
    McQuinlan, Yasmin
    Om, Deborah
    Staal, Floor H. E.
    Perennec, Tanguy
    Azzarouali, Sana
    Cardenas, Carlos E.
    Carver, Antony
    Korreman, Stine Sofia
    Bibault, Jean-Emmanuel
    RADIOTHERAPY AND ONCOLOGY, 2024, 200
  • [2] Dosimetric benefits of daily treatment plan adaptation for prostate cancer stereotactic body radiotherapy
    Miriam Eckl
    Gustavo R. Sarria
    Sandra Springer
    Marvin Willam
    Arne M. Ruder
    Volker Steil
    Michael Ehmann
    Frederik Wenz
    Jens Fleckenstein
    Radiation Oncology, 16
  • [3] Dosimetric benefits of daily treatment plan adaptation for prostate cancer stereotactic body radiotherapy
    Eckl, Miriam
    Sarria, Gustavo R.
    Springer, Sandra
    Willam, Marvin
    Ruder, Arne M.
    Steil, Volker
    Ehmann, Michael
    Wenz, Frederik
    Fleckenstein, Jens
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [4] MRI Based, Daily Adaptive Radiotherapy for Rectal Cancer: Contour Adaptation
    Kleijnen, J.
    Burbach, M.
    Verbraeken, T.
    Weggers, R.
    Zoetelief, A.
    Reerink, O.
    Lagendijk, J.
    Raaymakers, B.
    Asselen, B.
    MEDICAL PHYSICS, 2014, 41 (06) : 191 - 191
  • [5] Fully automated multi-criterial VMAT plan optimization for prostate cancer whole-pelvic radiotherapy
    Buschmann, M.
    Sharfo, A. W.
    Penninkhof, J.
    Seppenwoolde, Y.
    Goldner, G.
    Georg, D.
    Breedveld, S.
    Heijmen, B.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1016 - S1016
  • [6] Benchmarking daily plan adaptation on the Unity MR-Linac
    Jagt, T.
    Janssen, T.
    Betgen, A.
    Wiersema, L.
    Verhage, R.
    Garritsen, S.
    Vijlbrief-Bosman, T.
    de Ruiter, P.
    Peters, F.
    Remeijer, P.
    Marijnen, C.
    Sonke, J.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S60 - S62
  • [7] Fully Automated Multicriteria Optimization (MCO) Treatment Plan Generation for Radiation Treatment Planning
    Kirlik, G.
    D'Souza, W. D.
    Zhang, H. H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : S98 - S98
  • [8] OPTIMIZATION OF THE RADIOTHERAPY TREATMENT PLAN
    FYMAT, AL
    GREENFIELD, MA
    LO, SC
    FINDLEY, D
    APPLIED MATHEMATICS AND COMPUTATION, 1988, 27 (02) : 117 - 146
  • [9] Automated Rectal Cancer Radiotherapy Planning
    Huang, K.
    Das, P.
    Olanrewaju, A.
    Cardenas, C.
    Fuentes, D.
    Zhang, L.
    Hancock, D.
    Simonds, H.
    Rhee, D.
    Beddar, S.
    Briere, T.
    Court, L.
    MEDICAL PHYSICS, 2022, 49 (06) : E133 - E134
  • [10] Iterative dataset optimization in automated planning: Implementation for breast and rectal cancer radiotherapy
    Fan, Jiawei
    Wang, Jiazhou
    Zhang, Zhen
    Hu, Weigang